A complication of an axillary intra-aortic balloon pump

被引:0
|
作者
Kodack, Eileen [1 ]
Patel, Riya [2 ]
Bonner, Carain [1 ]
Pantin, Enrique [1 ]
机构
[1] Univ Med & Dent New Jersey, Dept Anesthesiol & Perioperat Med, New Brunswick, NJ 08901 USA
[2] Rutgers State Univ, Robert Wood Johnson Med Sch, New Brunswick, NJ 08901 USA
来源
JOURNAL OF SURGICAL CASE REPORTS | 2024年 / 2024卷 / 03期
关键词
intra-aortic balloon pump (IABP); intra-aortic balloon pump complication; intra-aortic balloon pump malposition; ascending aorta mass; descending aorta; transesophageal echo; BRIDGE;
D O I
10.1093/jscr/rjae092
中图分类号
R61 [外科手术学];
学科分类号
摘要
Intra-aortic balloon pumps (IABPs) are used to mechanically temporize a failing heart by decreasing afterload while increasing coronary perfusion pressure of the heart while more definitive treatment is sought. We report a case of a 65-year-old male with nonischemic cardiomyopathy, atrial fibrillation, thyroiditis, and non-Hodgkin lymphoma who presented with worsening heart failure. He underwent a percutaneous placement of a left axillary IABP with seemingly no complications. Approximately 3 weeks post-placement, the patient was taken for a heart transplant when an intraoperative transesophageal echo showed that the IABP was in the aortic arch and ascending aorta, instead of its proper placement in the descending aorta. The patient's arterial line showed waveforms appropriate for an IABP patient, and the patient showed no signs indicative of improper placement. This erroneous placement carried the potential to affect the aortic valve function, injure the aortic intima and/or occlude the aortic arch vessels. .
引用
收藏
页数:3
相关论文
共 50 条
  • [1] A rare complication of intra-aortic balloon pump placement
    Varadarajan, B
    Karski, J
    Vegas, A
    Heinrich, L
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2005, 19 (02) : 259 - 260
  • [2] Malpositioned Left Axillary Approach Intra-aortic Balloon Pump
    Drexler, Ian R.
    Martinez, Felipe
    Kirsch, Jacobo
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2020, 2 (06):
  • [3] The Intra-Aortic Balloon Pump
    Gajanan, Ganesh
    Brilakis, Emmanouil S.
    Siller-Matula, Jolanta M.
    Zolty, Ronald L.
    Velagapudi, Poonam
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2021, (168): : 1 - 12
  • [4] THE INTRA-AORTIC BALLOON PUMP
    不详
    BMJ-BRITISH MEDICAL JOURNAL, 1980, 281 (6243): : 764 - 764
  • [5] Significant vascular complications in percutaneous axillary intra-aortic balloon pump
    Nishida, Hidefumi
    Song, Tae
    Onsager, David
    Nguyen, Ann
    Grinstein, Jonathan
    Chung, Bow
    Smith, Bryan
    Kalantari, Sara
    Sarswat, Nitasha
    Kim, Gene
    Pinney, Sean
    Jeevanandam, Valluvan
    Milner, Ross
    Ota, Takeyoshi
    ANNALS OF VASCULAR SURGERY, 2022, 83 : 42 - 52
  • [6] VASCULAR COMPLICATIONS OF PERCUTANEOUS AXILLARY INTRA-AORTIC BALLOON PUMP THERAPY
    Nishida, Hidefumi
    Song, Tae
    Onsager, David
    Ann Nguyen
    Grinstein, Jonathan
    Chung, Bow
    Smith, Bryan
    Kalantari, Sara
    Sarswat, Nitasha
    Kim, Gene
    Pinney, Sean
    Jeevanandam, Valluvan
    Milner, Ross
    Ota, Takeyoshi
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (18) : 642 - 642
  • [7] Pneumatosis Intestinalis, a Dreaded Complication of Intra-Aortic Balloon Pump Use
    Rajpal, Saurabh
    Akkus, Nuri I.
    JOURNAL OF INVASIVE CARDIOLOGY, 2014, 26 (07): : E104 - E105
  • [8] Intra-aortic Balloon Pump Securement
    Holcomb, Brandi
    CRITICAL CARE NURSE, 2022, 42 (06) : 79 - 81
  • [9] Location of an intra-aortic balloon pump
    Fletcher, SN
    Riedel, BJCJ
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2001, 15 (02) : 267 - 268
  • [10] Is the intra-aortic balloon pump leaking?
    Sjauw, Krischan D.
    Piek, Jan J.
    LANCET, 2013, 382 (9905): : 1616 - 1617