Erroneous Pressure Reading from an Unpowered Intra-aortic Balloon Pump: Case Report

被引:0
|
作者
Belway, Dean [1 ]
Corrigan, Andrea [1 ]
Manchuk, Derek [2 ]
Bittira, Bindu [3 ,4 ]
机构
[1] Perfus Serv, Sudbury, ON P3E 5J1, Canada
[2] Northern Ontario Sch Med, Dept Anesthesia, Sudbury, ON P3E 5J1, Canada
[3] Northern Ontario Sch Med, Div Cardiac Surg, Sudbury, ON P3E 5J1, Canada
[4] Hlth Sci North, 41 Ramsey Lake Rd, Sudbury, ON P3E 5J1, Canada
关键词
case report; equipment malfunctions; intra-aortic balloon pump; CARDIAC-SURGERY; COUNTERPULSATION;
D O I
10.1097/MAT.0000000000001695
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intra-aortic balloon pump (IABP) technology is an established treatment modality for patients with acute cardiac failure. Advances in IABP technology have simplified use and reliability such that electromechanical malfunctions are relatively rare. We present a case of an unanticipated pressure output signal from a powered off Cardiosave IABP console resulting in an erroneous mean arterial blood pressure waveform being displayed during cardiopulmonary bypass. The patient suffered no ill effects and made a full recovery. As a result of this incident, our policies have been revised to ensure that all patients with an IABP brought to the operating room for surgery have at least two arterial lines, one of which is not off the IABP, to prevent future occurrences.
引用
收藏
页码:E211 / E214
页数:4
相关论文
共 50 条
  • [1] Ruptured intra-aortic balloon pump - A case report
    Kirksey, L
    Woody, DJ
    Plazk, L
    JOURNAL OF CARDIOVASCULAR SURGERY, 2002, 43 (04): : 461 - 464
  • [2] The Intra-Aortic Balloon Pump
    Gajanan, Ganesh
    Brilakis, Emmanouil S.
    Siller-Matula, Jolanta M.
    Zolty, Ronald L.
    Velagapudi, Poonam
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2021, (168): : 1 - 12
  • [3] THE INTRA-AORTIC BALLOON PUMP
    不详
    BMJ-BRITISH MEDICAL JOURNAL, 1980, 281 (6243): : 764 - 764
  • [4] Case report of visceral ischemia: The ''tail'' of an intra-aortic balloon pump
    Rivard, Justin
    Vergis, Ashley
    Kassum, Diamond
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2008, 135 (05): : 1167 - 1168
  • [5] Management of Intra-Aortic Balloon Pump Entrapment A Case Report and Review of the Literature
    Fitzmaurice, Gerard J.
    Collins, Anton
    Parissis, Haralambos
    TEXAS HEART INSTITUTE JOURNAL, 2012, 39 (05) : 621 - 626
  • [6] The Effects of Exercise on Cardiogenic Shock with an Intra-Aortic Balloon Pump: A Case Report
    Ferreira, Vanessa M.
    Rodrigues, Dayane Nunes
    Contreras, Carlos Alberto Mendez
    Rossi, Joao M.
    Ramos, Rui Fernando
    Oliveira, Gustavo
    Oliveira, Mayron F.
    ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2024, 121 (01)
  • [7] Intra-aortic balloon pump for hemodynamic support in hepatobiliary surgery: report of a case
    Resnick, Shelby
    Matsuoka, Lea
    Cesario, David
    Darwish, Ahmed
    Selby, R. Rick
    Sher, Linda
    SURGERY TODAY, 2012, 42 (08) : 793 - 796
  • [8] Intra-aortic balloon pump for hemodynamic support in hepatobiliary surgery: report of a case
    Shelby Resnick
    Lea Matsuoka
    David Cesario
    Ahmed Darwish
    R. Rick Selby
    Linda Sher
    Surgery Today, 2012, 42 : 793 - 796
  • [9] Location of an intra-aortic balloon pump
    Fletcher, SN
    Riedel, BJCJ
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2001, 15 (02) : 267 - 268
  • [10] Intra-aortic Balloon Pump Securement