Comparing conventional and computer-assisted surgery baseplate and screw placement in reverse shoulder arthroplasty

被引:19
|
作者
Venne, Gabriel [1 ]
Rasquinha, Brian J. [2 ]
Pichora, David [2 ,3 ,4 ]
Ellis, Randy E. [1 ,2 ,3 ,4 ,5 ]
Bicknell, Ryan [3 ,4 ]
机构
[1] Queens Univ, Dept Biomed & Mol Sci, Kingston, ON, Canada
[2] Queens Univ, Dept Mech & Mat Engn, Kingston, ON, Canada
[3] Queens Univ, Dept Surg, Kingston, ON, Canada
[4] Kingston Gen Hosp, Kingston, ON K7L 2V7, Canada
[5] Queens Univ, Sch Comp, Kingston, ON, Canada
关键词
Computer-assisted surgery; reverse shoulder arthroplasty; GLENOID COMPONENT FIXATION; ACCURACY; NAVIGATION;
D O I
10.1016/j.jse.2014.10.012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Preoperative planning and intraoperative navigation technologies have each been shown separately to be beneficial for optimizing screw and baseplate positioning in reverse shoulder arthroplasty (RSA) but to date have not been combined. This study describes development of a system for performing computer-assisted RSA glenoid baseplate and screw placement, including preoperative planning, intraoperative navigation, and postoperative evaluation, and compares this system with a conventional approach. Materials and methods: We used a custom-designed system allowing computed tomography (CT)-based preoperative planning, intraoperative navigation, and postoperative evaluation. Five orthopedic surgeons defined common preoperative plans on 3-dimensional CT reconstructed cadaveric shoulders. Each surgeon performed 3 computer-assisted and 3 conventional simulated procedures. The 3-dimensional CT reconstructed postoperative units were digitally matched to the preoperative model for evaluation of entry points, end points, and angulations of screws and baseplate. Values were used to find accuracy and precision of the 2 groups with respect to the defined placement. Statistical analysis was performed by t tests (alpha = .05). Results: Comparison of the groups revealed no difference in accuracy or precision of screws or baseplate entry points (P >.05). Accuracy and precision were improved with use of navigation for end points and angulations of 3 screws (P <.05). Accuracy of the inferior screw showed a trend of improvement with navigation (P >. 05). Navigated baseplate end point precision was improved (P <. 05), with a trend toward improved accuracy (P >.05). Conclusion: We conclude that CT-based preoperative planning and intraoperative navigation allow improved accuracy and precision for screw placement and precision for baseplate positioning with respect to a predefined placement compared with conventional techniques in RSA. Level of evidence: Basic Science, Surgical Technique. (C) 2015 Journal of Shoulder and Elbow Surgery Board of Trustees.
引用
收藏
页码:1112 / 1119
页数:8
相关论文
共 50 条
  • [31] Does fluoroscopy improve baseplate position compared to conventional technique in reverse shoulder arthroplasty? A preliminary study
    Galan-Olleros, Maria
    Lopiz, Yaiza
    Ciller, Gabriel
    Alcobia-Diaz, Borja
    Garcia-Fernandez, Carlos
    Marco, Fernando
    SHOULDER & ELBOW, 2023, 15 (01) : 15 - 26
  • [32] Rib fracture as a complication of erroneous baseplate and inferior screw placement in reverse total shoulder replacement: a case report
    Whitham, Robert D. J.
    Clifford, Zoe J.
    Tasker, Andrew J. B.
    Woods, David A.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2020, 29 (03) : E87 - E92
  • [33] An umbrella review comparing computer-assisted and conventional total joint arthroplasty: quality assessment and summary of evidence
    Hasan, Mohamed Mosaad
    Zhang, Manrui
    Beal, Matthew
    Ghomrawi, Hassan M. K.
    BMJ SURGERY INTERVENTIONS & HEALTH TECHNOLOGIES, 2020, 2 (01)
  • [34] COMPUTER-ASSISTED PEDICLE SCREW PLACEMENT PLANNING: TOWARDS CLINICAL PRACTICE
    Knez, Dejan
    Nahle, Imad S.
    Vrtovec, Tomaz
    Parent, Stefan
    Kadoury, Samuel
    2018 IEEE 15TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2018), 2018, : 249 - 252
  • [35] Computer-assisted thoracic pedicle screw placement -: An in vitro feasibility study
    Kim, KD
    Johnson, JP
    Bloch, O
    Masciopinto, JE
    SPINE, 2001, 26 (04) : 360 - 364
  • [36] Differences in component and limb alignment between computer-assisted and conventional surgery total knee arthroplasty
    Tsan-Wen Huang
    Kuo-Ti Peng
    Kuo-Chin Huang
    Mel S. Lee
    Robert Wen-Wei Hsu
    Knee Surgery, Sports Traumatology, Arthroscopy, 2014, 22 : 2954 - 2961
  • [37] Differences in component and limb alignment between computer-assisted and conventional surgery total knee arthroplasty
    Huang, Tsan-Wen
    Peng, Kuo-Ti
    Huang, Kuo-Chin
    Lee, Mel S.
    Hsu, Robert Wen-Wei
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2014, 22 (12) : 2954 - 2961
  • [38] Computer-assisted periacetabular screw placement: Comparison of different fluoroscopy-based navigation procedures with conventional technique
    Ochs, Bjoern Gunnar
    Gonser, Christoph
    Shiozawa, Thomas
    Badke, Andreas
    Weise, Kuno
    Rolauffs, Bernd
    Stuby, Fabian Maria
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2010, 41 (12): : 1297 - 1305
  • [39] Computer-Assisted Surgery and Intraoperative Three-Dimensional Imaging for Screw Placement in Different Pelvic Regions
    Grossterlinden, Lars
    Nuechtern, Jakob
    Begemann, Philipp G. C.
    Fuhrhop, Ina
    Petersen, Jan P.
    Ruecker, Andreas
    Rupprecht, Martin
    Lehmann, Wolfgang
    Schumacher, Udo
    Rueger, Johannes M.
    Briem, Daniel
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 71 (04): : 926 - 932
  • [40] Three-dimensional analysis of baseplate screw penetration in reverse total shoulder arthroplasty: risk of iatrogenic suprascapular neuropathy by screw violation
    Jang, Young Hoon
    Oh, Seung Yeol
    Kim, Sae Hoon
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2022, 31 (05) : 940 - 947