Comparison of formula accuracy for intraocular lens power calculation based on measurements by a swept-source optical coherence tomography optical biometer

被引:69
|
作者
Savini, Giacomo [1 ]
Hoffer, Kenneth J. [4 ,5 ]
Balducci, Nicole [1 ]
Barboni, Piero [2 ,3 ]
Schiano-Lomoriello, Domenico [1 ]
机构
[1] IRCCS GB Bietti Fdn, Via Livenza 3, Rome, Italy
[2] Studio Oculist Azeglio, Bologna, Italy
[3] Osped San Raffaele, Milan, Italy
[4] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[5] St Marys Eye Ctr, Santa Monica, CA USA
来源
关键词
SYSTEM;
D O I
10.1016/j.jcrs.2019.08.044
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To analyze the results of intraocular lens (IOL) power calculation using measurements by a swept-source optical coherence tomography (SS-OCT) optical biometer. Setting: IRCCS G.B. Bietti Foundation, Rome, Italy. Design: Evaluation of a diagnostic test instrument. Methods: Preoperative measurements by the OA-2000 (Tomey Inc.) were taken in a consecutive series of patients undergoing cataract surgery with one IOL model (AcrySof SN60WF; Alcon Laboratories, Inc.). Measurements were entered into the following formulas: Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hoffer Q, Holladay 1, Holladay 2, Holladay 2 with axial length adjustment, Kane, Olsen, Panacea, SRIVT, T2, and VRF. When refraction was measured at 1 month postoperatively, the mean arithmetic prediction error, the median absolute error (MedAE), and the percentage of eyes with a error of +/- 0.50 D or less were calculated after constant optimization. Results: We enrolled 150 eyes. All formulas yielded excellent outcomes. The MedAE ranged between 0.200 D and 0.259 D, with a statistically significant difference among formulas (P = .0004). The lowest MedAE values were obtained with the Barrett, EVO, Kane, Olsen(standalone), Radial Basis Function (RBF), and T2 formulas. The percentage of eyes with a prediction error of +/- 0.50 D or less ranged between 80.00% and 90.67%, with a statistically significant difference (P < .0001). The Barrett, EVO, Holladay 2 with axial length adjustment, Kane, RBF, and T2 achieved the highest percentages (>= 88%). Conclusions: Measurements provided by the SS-OCT optical biometer enable accurate IOL power calculation because all formulas yielded a prediction error within 0.50 D in at least 80% of eyes. Copyright (C) 2019 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:27 / 33
页数:7
相关论文
共 50 条
  • [41] Analysis of Crystalline Lens Tilt and Decentration Using a Fully Automated Swept-Source Optical Coherence Tomography Biometer
    Montes-Mico, Robert
    Nilsson, Maria
    Brautaset, Rune
    Venkataraman, Abinaya Priya
    Dominguez-Vicent, Alberto
    JOURNAL OF REFRACTIVE SURGERY, 2025, 41 (02) : e155 - e163
  • [42] Stability of Intraocular Lens With Different Haptic Design: A Swept-Source Optical Coherence Tomography Study
    Xiao, Zixuan
    Wang, Geng
    Zhen, Miaoru
    Zhao, Zifeng
    FRONTIERS IN MEDICINE, 2021, 8
  • [43] Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
    Ribeiro, Bruno Barbosa
    Monteiro, Silvia
    Marques, Joao Heitor
    Abreu, Ana Carolina
    Pinto, Maria do Ceu
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (08) : e562 - e568
  • [44] Comparison of anterior segment measurements obtained using a swept-source optical coherence tomography biometer and a Scheimpflug-Placido tomographer
    Lu, Weicong
    Li, Yue
    Savini, Giacomo
    Song, Benhao
    Hu, Qingjie
    Wang, Qinmei
    Bao, Fangjun
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (03): : 298 - 304
  • [45] Repeatability and reproducibility of a new fully automatic measurement optical low coherence reflectometry biometer and agreement with swept-source optical coherence tomography-based biometer
    Yu, Jinjin
    Zhao, Guoli
    Lei, Chak Seng
    Wan, Ting
    Ning, Rui
    Xing, Wenqian
    Ma, Xindi
    Pan, Hongxian
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Huang, Jinhai
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2024, 108 (05) : 673 - 678
  • [46] Comparison of Precision, Agreement, and Accuracy of Two Swept-Source Optical Coherence Tomography Biometers
    Guarro, Merce
    Vazquez, Meritxell
    Diaz, Juan Carlos
    Ruiz, Sergi
    Gimeno, Maties
    Rodriguez, Lara
    Lopez, Elena
    Sararols, Laura
    Biarnes, Marc
    DIAGNOSTICS, 2024, 14 (21)
  • [47] Advances in swept-source optical coherence tomography and optical coherence tomography angiography
    Zheng, Fang
    Deng, Xiaofeng
    Zhang, Qi
    He, Jingliang
    Ye, Panpan
    Liu, Shan
    Li, Peng
    Zhou, Jian
    Fang, Xiaoyun
    ADVANCES IN OPHTHALMOLOGY PRACTICE AND RESEARCH, 2023, 3 (02): : 67 - 79
  • [48] Changes in the anterior segment after cycloplegia with a biometer using swept-source optical coherence tomography
    Higashiyama, Tomoaki
    Iwasa, Maki
    Ohji, Masahito
    PLOS ONE, 2017, 12 (08):
  • [49] Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer
    Bao Shu
    Fangjun Bao
    Giacomo Savini
    Weicong Lu
    Ruixue Tu
    Haisi Chen
    Benhao Song
    Qinmei Wang
    Jinhai Huang
    Eye and Vision, 7
  • [50] Lower refractive prediction accuracy of total keratometry using intraocular lens formulas loaded onto a swept-source optical biometer
    Yukitaka Danjo
    Reina Ohji
    Sayo Maeno
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2023, 261 : 137 - 146