Comparison of formula accuracy for intraocular lens power calculation based on measurements by a swept-source optical coherence tomography optical biometer
被引:69
|作者:
Savini, Giacomo
[1
]
Hoffer, Kenneth J.
[4
,5
]
Balducci, Nicole
[1
]
Barboni, Piero
[2
,3
]
Schiano-Lomoriello, Domenico
[1
]
机构:
[1] IRCCS GB Bietti Fdn, Via Livenza 3, Rome, Italy
[2] Studio Oculist Azeglio, Bologna, Italy
[3] Osped San Raffaele, Milan, Italy
[4] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[5] St Marys Eye Ctr, Santa Monica, CA USA
来源:
关键词:
SYSTEM;
D O I:
10.1016/j.jcrs.2019.08.044
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose: To analyze the results of intraocular lens (IOL) power calculation using measurements by a swept-source optical coherence tomography (SS-OCT) optical biometer. Setting: IRCCS G.B. Bietti Foundation, Rome, Italy. Design: Evaluation of a diagnostic test instrument. Methods: Preoperative measurements by the OA-2000 (Tomey Inc.) were taken in a consecutive series of patients undergoing cataract surgery with one IOL model (AcrySof SN60WF; Alcon Laboratories, Inc.). Measurements were entered into the following formulas: Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hoffer Q, Holladay 1, Holladay 2, Holladay 2 with axial length adjustment, Kane, Olsen, Panacea, SRIVT, T2, and VRF. When refraction was measured at 1 month postoperatively, the mean arithmetic prediction error, the median absolute error (MedAE), and the percentage of eyes with a error of +/- 0.50 D or less were calculated after constant optimization. Results: We enrolled 150 eyes. All formulas yielded excellent outcomes. The MedAE ranged between 0.200 D and 0.259 D, with a statistically significant difference among formulas (P = .0004). The lowest MedAE values were obtained with the Barrett, EVO, Kane, Olsen(standalone), Radial Basis Function (RBF), and T2 formulas. The percentage of eyes with a prediction error of +/- 0.50 D or less ranged between 80.00% and 90.67%, with a statistically significant difference (P < .0001). The Barrett, EVO, Holladay 2 with axial length adjustment, Kane, RBF, and T2 achieved the highest percentages (>= 88%). Conclusions: Measurements provided by the SS-OCT optical biometer enable accurate IOL power calculation because all formulas yielded a prediction error within 0.50 D in at least 80% of eyes. Copyright (C) 2019 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:27 / 33
页数:7
相关论文