Average Glandular Dose in Digital Mammography and Breast Tomosynthesis

被引:48
|
作者
Olgar, T. [1 ,2 ]
Kahn, T. [1 ]
Gosch, D. [1 ]
机构
[1] Univ Klinikum Leipzig AoR, Klin & Poliklin Diagnost & Interventionelle Radio, Leipzig, Germany
[2] Ankara Univ, Dept Engn Phys, Fac Engn, TR-06100 Ankara, Turkey
关键词
breast tomosynthesis; average glandular dose; compressed breast thickness; digital mammography; SCREEN-FILM MAMMOGRAPHY; POPULATION; CANCER; UK; COMBINATIONS; MOLYBDENUM; RHODIUM; FILTER; WOMEN;
D O I
10.1055/s-0032-1312877
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine the average glandular dose (AGD) in digital full-field mammography (2 D imaging mode) and in breast tomosynthesis (3 D imaging mode). Materials and Methods: Using the method described by Boone, the AGD was calculated from the exposure parameters of 2247 conventional 2 D mammograms and 984 mammograms in 3 D imaging mode of 641 patients examined with the digital mammographic system Hologic Selenia Dimensions. The breast glandular tissue content was estimated by the Hologic R2 Quantra automated volumetric breast density measurement tool for each patient from right craniocaudal (RCC) and left craniocaudal (LCC) images in 2 D imaging mode. Results: The mean compressed breast thickness (CBT) was 52.7 mm for craniocaudal (CC) and 56.0 mm for mediolateral oblique (MLO) views. The mean percentage of breast glandular tissue content was 18.0% and 17.4% for RCC and LCC projections, respectively. The mean AGD values in 2 D imaging mode per exposure for the standard breast were 1.57 mGy and 1.66 mGy, while the mean AGD values after correction for real breast composition were 1.82 mGy and 1.94 mGy for CC and MLO views, respectively. The mean AGD values in 3D imaging mode per exposure for the standard breast were 2.19 mGy and 2.29 mGy, while the mean AGD values after correction for the real breast composition were 2.53 mGy and 2.63 mGy for CC and MLO views, respectively. No significant relationship was found between the AGD and CBT in 2 D imaging mode and a good correlation coefficient of 0.98 in 3 D imaging mode. Conclusion: In this study the mean calculated AGD per exposure in 3 D imaging mode was on average 34% higher than for 2 D imaging mode for patients examined with the same CBT.
引用
收藏
页码:911 / 918
页数:8
相关论文
共 50 条
  • [1] Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data
    Bouwman, R. W.
    van Engen, R. E.
    Young, K. C.
    den Heeten, G. J.
    Broeders, M. J. M.
    Schopphoven, S.
    Jeukens, C. R. L. P. N.
    Veldkamp, W. J. H.
    Dance, D. R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (20): : 7893 - 7907
  • [2] Digital breast tomosynthesis and digital mammography: A comparison of figures of merit for various average glandular doses
    Kim, Ye-seul
    Park, Hye-Suk
    Park, SuJin
    Kim, Hee-Joung
    Choi, Jae-Gu
    Choi, Young-Wook
    Park, Jun-Ho
    Lee, Jae-Jun
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 62 (09) : 1346 - 1353
  • [3] Digital breast tomosynthesis and digital mammography: A comparison of figures of merit for various average glandular doses
    Ye-seul Kim
    Hye-Suk Park
    SuJin Park
    Hee-Joung Kim
    Jae-Gu Choi
    Young-Wook Choi
    Jun-Ho Park
    Jae-Jun Lee
    [J]. Journal of the Korean Physical Society, 2013, 62 : 1346 - 1353
  • [4] Normalized glandular dose coefficients in mammography, digital breast tomosynthesis and dedicated breast CT
    Sarno, A.
    Mettivier, G.
    Di Lillo, F.
    Tucciariello, R. M.
    Bliznakova, K.
    Russo, P.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 55 : 142 - 148
  • [5] Impact of age and breast thickness on mean glandular dose of standard digital mammography and digital breast tomosynthesis
    Norsuddin, Norhashimah Mohd
    Sin, Justine Go Mei
    Ravintaran, Rathieswari
    Arasaratnam, Shantini
    Karim, Muhammad Khalis Abdul
    [J]. APPLIED RADIATION AND ISOTOPES, 2023, 192
  • [6] Evaluation of average glandular dose and investigation of the relationship with compressed breast thickness in dual energy contrast enhanced digital mammography and digital breast tomosynthesis
    Fusco, Roberta
    Raiano, Nicola
    Raiano, Concetta
    Maio, Francesca
    Vallone, Paolo
    Raso, Mauro Mattace
    Setola, Sergio Venanzio
    Granata, Vincenza
    Rubulotta, Maria Rosaria
    Barretta, Maria Lusia
    Petrosino, Teresa
    Petrillo, Antonella
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2020, 126
  • [7] Evaluation of the Displayed Average Glandular Dose of Hologic Digital Breast Tomosynthesis (DBT) Systems
    Osunbayo, O.
    Feng, J.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [8] Comparison of Mean Glandular Dose between Full-Field Digital Mammography and Digital Breast Tomosynthesis
    Teoh, Kar Choon
    Manan, Hanani Abdul
    Norsuddin, Norhashimah Mohd
    Rizuana, Iqbal Hussain
    [J]. HEALTHCARE, 2021, 9 (12)
  • [9] In vivo average glandular dose evaluation: one-to-one comparison between digital breast tomosynthesis and full-field digital mammography
    Cavagnetto, Francesca
    Taccini, Gianni
    Rosasco, Raffaella
    Bampi, Rossana
    Calabrese, Massimo
    Tagliafico, Alberto
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 157 (01) : 53 - 61
  • [10] Computation of the glandular radiation dose in digital tomosynthesis of the breast
    Sechopoulos, Ioannis
    Suryanarayanan, Sankarararnan
    Vedantham, Srinivasan
    D'Orsi, Carl
    Karellas, Andrew
    [J]. MEDICAL PHYSICS, 2007, 34 (01) : 221 - 232