共 50 条
Comparison of two portable clinical analyzers to one stationary analyzer for the determination of blood gas partial pressures and blood electrolyte concentrations in horses
被引:14
|作者:
Kirsch, Katharina
[1
]
Detilleux, Johann
[2
]
Serteyn, Didier
[1
]
Sandersen, Charlotte
[1
]
机构:
[1] Univ Liege, Fac Vet Med, Dept Compan Anim & Equids, Equine Clin, Liege, Belgium
[2] Univ Liege, Fac Vet Med, Dept Quantitat Genet, Liege, Belgium
来源:
关键词:
ACID-BASE;
STATISTICAL-METHODS;
HEMOGLOBIN;
AGREEMENT;
INTERFERENCE;
POTASSIUM;
PRECISION;
D O I:
10.1371/journal.pone.0211104
中图分类号:
O [数理科学和化学];
P [天文学、地球科学];
Q [生物科学];
N [自然科学总论];
学科分类号:
07 ;
0710 ;
09 ;
摘要:
Portable blood gas analyzers are used to facilitate diagnosis and treatment of disorders related to disturbances of acid-base and electrolyte balance in the ambulatory care of equine patients. The aim of this study was to determine whether 2 portable analyzers produce results in agreement with a stationary analyzer. Blood samples from 23 horses hospitalized for various medical reasons were included in this prospective study. Blood gas analysis and electrolyte concentrations measured by the portable analyzers VetStat and epoc were compared to those produced by the cobas b 123 analyzer via concordance analysis, Passing-Bablok regression and Bland-Altman analysis. Limits of agreement indicated relevant bias between the VetStat and cobas b 123 for partial pressure of oxygen (pO(2); 27.5-33.8 mmHg), sodium ([Na+]; 4.3-21.6 mmol/L) and chloride concentration ([CI-]; 0.3-7.9 mmol/L) and between the epoc and cobas b 123 for pH (0.070-0.022), partial pressure of carbon dioxide (pCO(2); 3.6-7.3 mmHg), pO(2) (36.2-32.7 mmHg) and [Na+] (0.38.1 mmol/L). The VetStat analyzer yielded results that were in agreement with the cobas b 123 analyzer for determination of pH, pCO(2), bicarbonate ([HCO3-]) and potassium concentration [K+], while the epoc analyzer achieved acceptable agreement for [HCO3] and [K+]. The VetStat analyzer may be useful in performing blood gas analysis in equine samples but analysis of [Na+], [Cl-] and pO(2) should be interpreted with caution. The epoc delivered reliable results for [HCO3] and [K+], while results for pH, pCO(2), pO(2) and [Na+] should be interpreted with caution.
引用
收藏
页数:14
相关论文
相似文献