Comparison of two portable clinical analyzers to one stationary analyzer for the determination of blood gas partial pressures and blood electrolyte concentrations in horses

被引:14
|
作者
Kirsch, Katharina [1 ]
Detilleux, Johann [2 ]
Serteyn, Didier [1 ]
Sandersen, Charlotte [1 ]
机构
[1] Univ Liege, Fac Vet Med, Dept Compan Anim & Equids, Equine Clin, Liege, Belgium
[2] Univ Liege, Fac Vet Med, Dept Quantitat Genet, Liege, Belgium
来源
PLOS ONE | 2019年 / 14卷 / 02期
关键词
ACID-BASE; STATISTICAL-METHODS; HEMOGLOBIN; AGREEMENT; INTERFERENCE; POTASSIUM; PRECISION;
D O I
10.1371/journal.pone.0211104
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Portable blood gas analyzers are used to facilitate diagnosis and treatment of disorders related to disturbances of acid-base and electrolyte balance in the ambulatory care of equine patients. The aim of this study was to determine whether 2 portable analyzers produce results in agreement with a stationary analyzer. Blood samples from 23 horses hospitalized for various medical reasons were included in this prospective study. Blood gas analysis and electrolyte concentrations measured by the portable analyzers VetStat and epoc were compared to those produced by the cobas b 123 analyzer via concordance analysis, Passing-Bablok regression and Bland-Altman analysis. Limits of agreement indicated relevant bias between the VetStat and cobas b 123 for partial pressure of oxygen (pO(2); 27.5-33.8 mmHg), sodium ([Na+]; 4.3-21.6 mmol/L) and chloride concentration ([CI-]; 0.3-7.9 mmol/L) and between the epoc and cobas b 123 for pH (0.070-0.022), partial pressure of carbon dioxide (pCO(2); 3.6-7.3 mmHg), pO(2) (36.2-32.7 mmHg) and [Na+] (0.38.1 mmol/L). The VetStat analyzer yielded results that were in agreement with the cobas b 123 analyzer for determination of pH, pCO(2), bicarbonate ([HCO3-]) and potassium concentration [K+], while the epoc analyzer achieved acceptable agreement for [HCO3] and [K+]. The VetStat analyzer may be useful in performing blood gas analysis in equine samples but analysis of [Na+], [Cl-] and pO(2) should be interpreted with caution. The epoc delivered reliable results for [HCO3] and [K+], while results for pH, pCO(2), pO(2) and [Na+] should be interpreted with caution.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] A comparison of two blood gas analyzers and a proposal for an internal quality control protocol.
    Coletto, SL
    Colossi, R
    Camargo, JL
    [J]. CLINICAL CHEMISTRY, 2001, 47 (06) : A83 - A83
  • [22] Comparison of the AVL OPTI 1 (Point of care) blood gas analyzer with traditional laboratory-based analyzers
    Franklin, ML
    Peruzzi, WT
    Shapiro, BA
    Lam, N
    [J]. ANESTHESIOLOGY, 1996, 85 (3A) : A482 - A482
  • [23] Comparison of Enterprise Point-of-Care and Nova Biomedical Critical Care Xpress analyzers for determination of arterial pH, blood gas, and electrolyte values in canine and equine blood
    Elmeshreghi, Taher N.
    Grubb, Tamara L.
    Greene, Stephen A.
    Ragle, Claude A.
    Wardrop, Jane A.
    [J]. VETERINARY CLINICAL PATHOLOGY, 2018, 47 (03) : 415 - 424
  • [24] Analytical Performance Evaluation of the New GEM® Premier™ 5000 in Comparison to the Epoc® Blood Gas Analyzer in Horses
    Sandersen, Charlotte
    Dmitrovic, Petra
    Dupont, Julien
    Cesarini, Carla
    Guyot, Hugues
    Serteyn, Didier
    Kirsch, Katharina
    [J]. VETERINARY SCIENCES, 2023, 10 (02)
  • [25] EVALUATION OF BLOOD-GAS AND PH METHODS ON THE I-STAT(TM) PORTABLE CLINICAL ANALYZER
    BOTTJEN, PL
    HAVEN, MC
    [J]. CLINICAL CHEMISTRY, 1995, 41 (06) : S196 - S196
  • [26] Comparison between blood gas analyzer and central laboratory analyzer for the determination of electrolytes in patients with acute respiratory acidosis
    Garcia-Pachon, Eduardo
    Soler-Sempere, Maria J.
    Garcia-Padilla, Eduardo
    Zamora-Molina, Lucia
    Grau-Delgado, Justo
    Padilla-Navas, Isabel
    Sanchez-Hernandez, Jose F.
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (05) : E125 - E127
  • [27] Three-year monitoring and comparison of results from two identical blood gas analyzers
    Huang, Yun
    Dean, Robert
    Dubbelman, Yvonne
    Vincent, Anne
    Chan, Ying-pui Michael
    [J]. PRACTICAL LABORATORY MEDICINE, 2022, 31
  • [28] Three-year monitoring and comparison of results from two identical blood gas analyzers
    Huang, Yun
    Dean, Robert
    Dubbelman, Yvonne
    Vincent, Anne
    Chan, Ying-pui Michael
    [J]. PRACTICAL LABORATORY MEDICINE, 2022, 31
  • [29] Blood Gas, Oxygen Saturation, pH, and Lactate Values in Elasmobranch Blood Measured with a Commercially Available Portable Clinical Analyzer and Standard Laboratory Instruments
    Gallagher, Austin J.
    Frick, Lorenz H.
    Bushnell, Peter G.
    Brill, Richard W.
    Mandelman, John W.
    [J]. JOURNAL OF AQUATIC ANIMAL HEALTH, 2010, 22 (04) : 229 - 234
  • [30] Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
    Kim, Sang-Mi
    Park, Hyung-Doo
    [J]. ANNALS OF LABORATORY MEDICINE, 2022, 42 (04) : 467 - +