Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers

被引:1
|
作者
Kim, Sang-Mi [1 ]
Park, Hyung-Doo [1 ]
机构
[1] Sungkyunkwan Univ, Samsung Med Ctr, Dept Lab Med & Genet, Sch Med, 81 Irwon Ro, Seoul 06351, South Korea
关键词
Analytical performance evaluation; Blood gas analyzer; i-SmartCare; 10; Method comparison; Point-of-care; ELECTROLYTE;
D O I
10.3343/alm.2022.42.4.467
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the comparator, according to the CLSI guidelines. Ten analytes (pH, pCO(2), pO(2), Na+, K+, Cl-, iCa(2+), glucose, lactate, and Hct) were tested on six analyzers: i-SmartCare 10, ABL90 FLEX PLUS (Radiometer Medical ApS, Copenhagen, Denmark), i-Stat (Abbott Point of Care Inc., Princeton, NJ, USA), RapidLab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA), Stat Profile pHOx Ultra (Nova Biomedical, Waltham, MA, USA), and Gem Premier 5000 (Instrumentation Laboratory, Bedford, MA, USA). The total imprecision and linearity (r(2)>0.99) were excellent, except for a few analytes that narrowly escaped the preset criteria. Interference was noted for Nat in the presence of a high K+ level and for iCa(2+) in the presence of high K+ and Mg2+ levels. Forty of 48 items demonstrated either a proportional or systematic difference in regression analysis; the relative mean difference (%) of 14/48 items escaped the allowable total error in the difference plot analysis. i-SmartCare 10 shows acceptable performance, and using a single POC blood gas analyzer is recommended for monitoring.
引用
收藏
页码:467 / +
页数:9
相关论文
共 50 条
  • [1] Analytical performance characteristics of the i-SmartCare 10 analyzer for point-of-care testing
    Kim, S.
    Oh, J.
    Park, H.
    [J]. CLINICA CHIMICA ACTA, 2022, 530 : S101 - S101
  • [2] Comparison of point-of-care and central laboratory analyzers for blood gas and lactate measurements
    Indrasari, Nuri Dyah
    Wonohutomo, Jessica Purwanti
    Sukartini, Ninik
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2019, 33 (05)
  • [3] Evaluation of the Element point-of-care blood gas analyzer for use in horses
    Krueger, Clarisa R.
    Hackett, Eileen S.
    Hess, Ann M.
    Mama, Khursheed R.
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (03) : 279 - 285
  • [4] Evaluation of Two Handheld Point-of-Care Blood Gas Analyzers in Healthy Donkeys
    Bonelli, Francesca
    Laus, Fulvio
    Briganti, Angela
    Evangelista, Flavia
    Bazzano, Marilena
    Conte, Giuseppe
    Sgorbini, Micaela
    [J]. JOURNAL OF EQUINE VETERINARY SCIENCE, 2019, 79 : 94 - 99
  • [5] Point-of-care blood gas analysers: a performance evaluation
    St-Louis, P
    [J]. CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 139 - 144
  • [6] COMPARISON OF SIEMENS 405 AND EPOC POINT-OF-CARE BLOOD GAS ANALYZERS WITH SIEMENS 865
    Van Even, Ellen
    Ghys, Tim
    De Puydt, Hendrik
    Louagie, Henk
    [J]. ACTA CLINICA BELGICA, 2012, 67 (01): : 61 - 61
  • [7] Clinical evaluation of a new point-of-care blood gas analyzer, the ''IRMA system''
    Ozaki, M
    Morioka, N
    Matsukawa, T
    Atarashi, K
    Sessler, DI
    Suzuki, H
    [J]. ANESTHESIOLOGY, 1997, 87 (03) : A428 - A428
  • [8] Multicenter study on the analytical performance of a new point-of-care blood gas analyzer
    Goerlach-Graw, A
    Engel, B
    Halwachs-Baumann, G
    Kost, G
    Kraft, M
    Küster, H
    Miedema, K
    Nagel, D
    Normand, H
    Okorodudu, A
    Olander, B
    Rolinski, B
    Wiley, B
    Wu, J
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A171 - A171
  • [9] Method validation of GEM® 4000 blood gas analyzer for point-of-care testing
    Martinez Sanchez, L. M.
    Lafuente Muro, S.
    Castillo Ribelles, L.
    Martinez Casademont, M.
    [J]. CLINICA CHIMICA ACTA, 2019, 493 : S765 - S765
  • [10] Evaluation of the Analytical Performance of the Modified Enterprise Point-of-Care Blood Gas and Electrolyte Analyzer in a Pediatric Hospital
    Agarwal, Shweta
    Shafi, Rabia
    Edwards, Rachel M.
    Collymore, Laura
    Devaraj, Sridevi
    [J]. POINT OF CARE, 2014, 13 (04): : 132 - 136