Evaluation of the Analytical Performance of the Modified Enterprise Point-of-Care Blood Gas and Electrolyte Analyzer in a Pediatric Hospital

被引:2
|
作者
Agarwal, Shweta [1 ]
Shafi, Rabia [1 ]
Edwards, Rachel M. [2 ]
Collymore, Laura [2 ]
Devaraj, Sridevi [1 ,2 ]
机构
[1] Baylor Coll Med, Houston, TX 77030 USA
[2] Texas Childrens Hosp, 6621 Fannin St, Houston, TX 77030 USA
来源
POINT OF CARE | 2014年 / 13卷 / 04期
关键词
EPOC; blood gas; point of care; electrolyte; evaluation;
D O I
10.1097/POC.0000000000000036
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Arterial blood gas testing is an essential component that directs the evaluation of a patient's general condition, especially in the emergent setting. We evaluated the analytical performance of the modified Enterprise point-of-care (EPOC) blood gas and electrolyte analyzer system with respect to precision and accuracy and compared our results with those obtained using the point-of-care i-STAT method and the criterion standard ABL 835 Radiometer blood gas analyzer in a pediatric hospital. Evaluation studies were performed as per the Clinical Laboratory Standards Institute guidelines and included intra-assay and interassay precision and accuracy, linearity, and concordance analyses with i-STAT and ABL 835 Radiometer. Intra-assay and interassay coefficient of variation for each analyte was calculated and was less than 6% for all analytes in the ranges tested including pH, PO2, PCO2, Na+, K+, Ca2+, glucose, lactate, and hematocrit, Cl-, and creatinine. All of the analytes were linear across the reportable range. For the methods comparison of EPOC system with the i-STAT method, using 40 samples, the correlation coefficients were r > 0.94 for all analytes with the exception of Ca2+ (r = 0.61). For the methods comparison of EPOC system with the criterion standard ABL 835 Radiometer on 40 samples, the correlation coefficients were higher than 0.97 for all analytes with the exception of hematocrit and hemoglobin (r = 0.8). Thus, in our pediatric hospital setting, the EPOC system showed excellent precision and accuracy and compared favorably with the i-STAT and ABL 835 Radiometer assay results. These findings, together with the low cost, room temperature storage of test cards and availability of metabolites such as lactate, glucose, and creatinine along with the wireless connectivity of the EPOC system, provide an operational advantage over other point-of-care blood gas analyzers currently available.
引用
收藏
页码:132 / 136
页数:5
相关论文
共 50 条
  • [1] Evaluation of the Enterprise Point-of-Care (EPOC) System for Blood Gas and Electrolyte Analysis
    Nichols, James H.
    Rajadhyaksha, Aparna
    Rodriguez, Mirian
    [J]. POINT OF CARE, 2008, 7 (01): : 7 - 11
  • [2] Multicenter study on the analytical performance of a new point-of-care blood gas analyzer
    Goerlach-Graw, A
    Engel, B
    Halwachs-Baumann, G
    Kost, G
    Kraft, M
    Küster, H
    Miedema, K
    Nagel, D
    Normand, H
    Okorodudu, A
    Olander, B
    Rolinski, B
    Wiley, B
    Wu, J
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A171 - A171
  • [3] Analytical evaluation of SENDX 100TM blood gas/electrolyte analyzer for point-of-care testing.
    Liu, X
    Mohammad, AA
    Petersen, JR
    Okorodudu, AO
    [J]. CLINICAL CHEMISTRY, 1997, 43 : 277 - 277
  • [4] Multiple site analytical evaluation of a portable blood gas/electrolyte analyzer for point of care testing
    Chance, JJ
    Li, DJ
    Sokoll, LJ
    Silberman, MA
    Engelstad, ME
    Nichols, JH
    Liu, XH
    Mohammad, AA
    Petersen, JR
    Okorodudu, AO
    [J]. CRITICAL CARE MEDICINE, 2000, 28 (06) : 2081 - 2085
  • [5] Evaluation of the Element point-of-care blood gas analyzer for use in horses
    Krueger, Clarisa R.
    Hackett, Eileen S.
    Hess, Ann M.
    Mama, Khursheed R.
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (03) : 279 - 285
  • [6] Point-of-care blood gas analysers: a performance evaluation
    St-Louis, P
    [J]. CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 139 - 144
  • [7] Clinical evaluation of a new point-of-care blood gas analyzer, the ''IRMA system''
    Ozaki, M
    Morioka, N
    Matsukawa, T
    Atarashi, K
    Sessler, DI
    Suzuki, H
    [J]. ANESTHESIOLOGY, 1997, 87 (03) : A428 - A428
  • [8] Analytical Performance Evaluation and Comparability of Patient Results Within a Point-of-Care Blood Gas Network
    Oliver, Paloma
    Fernandez-Calle, Pilar
    Rico, Natividad
    Jose Alcaide, Maria
    Gomez-Rioja, Ruben
    Buno, Antonio
    Manuel Iturzaeta, Jose
    [J]. POINT OF CARE, 2013, 12 (03): : 144 - 149
  • [9] Analytical and pre-analytical performance characteristics of a novel cartridge-type blood gas analyzer for point-of-care and laboratory testing
    Oyaert, Matthijs
    Van Maerken, Tom
    Bridts, Silke
    Van Loon, Silvi
    Laverge, Heleen
    Stove, Veronique
    [J]. CLINICAL BIOCHEMISTRY, 2018, 53 : 116 - 126
  • [10] Innotrac Aio! a point-of-care or a routine analyzer? Analytical performance and plasma/whole blood comparison
    Altinier, Sara
    Zaninotto, Martina
    Mion, Monica Maria
    Plebani, Mario
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (10) : 1278 - 1282