Innotrac Aio! a point-of-care or a routine analyzer? Analytical performance and plasma/whole blood comparison

被引:2
|
作者
Altinier, Sara [1 ]
Zaninotto, Martina [1 ]
Mion, Monica Maria [1 ]
Plebani, Mario [1 ]
机构
[1] Univ Hosp Padova, Dept Lab Med, Padua, Italy
关键词
cardiac markers; creatine kinase-MB; myoglobin; point-of-care testing; troponin I;
D O I
10.1515/CCLM.2006.223
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background and methods: The aim of the present study was to evaluate the analytical performance of the Innotrac Aio! analyzer in measuring the cardiac markers troponin I (Tnl), myoglobin (Myo) and creatine kinase-MB (CK-MB), both in lithium heparin plasma and in whole blood. Results: TnI analytical sensitivity was 0.012 mu g/L and the concentration corresponding to CV=10% was 0.036 mu g/L. In healthy subjects, the 99th percentile TnI values were 0.023 and 0.016 mu g/L in whole blood and in plasma, respectively. One hundred samples were tested both in whole blood and in plasma: Tnl-whole-bloods=1.16-plasmaq+0.011, bias=0.18 (95% CI from -0.22 to +0.59); Myo-whole-blood=1.04-plasma -1.93, bias=+1.08 (95% CI from -6.17 to +8.32); CK-MB-whole-blood=1.11-plasma -0.09, bias=+0.96 (95% CI from -0.53 to +2.45). The method comparison with the RxL Dimension analyzer for TnI and Myo gave the following results: TnI AioI =0.30 TnI RxL + 0.00; bias=-5.80 (95% CI from -7.81 to -3.78), Myo Aio! =1.33 Myo RxL + 0.42, bias=+53.09 (95% CI from +38.44 to +67.74). Conclusions: The analytical performance of the Innotrac Aio! analyzer was satisfactory for all three cardiac markers evaluated and, in particular, the TnI method provided sensitive and accurate results. The most important finding in this study is the possibility to perform the tests as either routine or point-of-care analysis, thus overcoming the variability of results obtained employing different methods.
引用
收藏
页码:1278 / 1282
页数:5
相关论文
共 50 条
  • [1] Analytical performance of Immunotrac AIO analyzer cardiac markers on plasma and whole blood samples
    Altinier, S
    Zaninotto, M
    Mion, M
    Plebani, M
    [J]. CLINICA CHIMICA ACTA, 2005, 355 : S395 - S395
  • [2] Analytical performance of the Abaxis Piccolo Xpress® point of care analyzer in whole blood, serum, and plasma
    Murata, Kazunori
    Glaser, Laurel
    Nardiello, Mary
    Richardson, Shaun
    Ramanathan, Lakshmi V.
    Carlow, Dean C.
    [J]. CLINICAL BIOCHEMISTRY, 2015, 48 (18) : 1344 - 1346
  • [3] Multicenter study on the analytical performance of a new point-of-care blood gas analyzer
    Goerlach-Graw, A
    Engel, B
    Halwachs-Baumann, G
    Kost, G
    Kraft, M
    Küster, H
    Miedema, K
    Nagel, D
    Normand, H
    Okorodudu, A
    Olander, B
    Rolinski, B
    Wiley, B
    Wu, J
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A171 - A171
  • [4] Evaluation of a whole blood point-of-care coagulation analyzer in dogs
    Conroy, Elizabeth M.
    Lyons, Bridget M.
    Koenig, Amie
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2024,
  • [5] Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma
    Murata, Kazunori
    Glaser, Laurel
    Nardiello, Mary
    Ramanathan, Lakshmi V.
    Carlow, Dean C.
    [J]. DATA IN BRIEF, 2018, 16 : 81 - 89
  • [6] A rapid and quantitative D-Dimer assay in whole blood and plasma on the point-of-care PATHFAST analyzer
    Fukuda, Teruko
    Kasai, Hidetoshi
    Kusano, Takeo
    Shimazu, Chisato
    Kawasugi, Kazuo
    Miyazawa, Yukihisa
    [J]. THROMBOSIS RESEARCH, 2007, 120 (05) : 695 - 701
  • [7] Evaluation of the Analytical Performance of the Modified Enterprise Point-of-Care Blood Gas and Electrolyte Analyzer in a Pediatric Hospital
    Agarwal, Shweta
    Shafi, Rabia
    Edwards, Rachel M.
    Collymore, Laura
    Devaraj, Sridevi
    [J]. POINT OF CARE, 2014, 13 (04): : 132 - 136
  • [8] Comparison of a point-of-care glucometer and an automated biochemical analyzer to measure glucose concentrations in whole blood, serum, and plasma from porcine blood samples.
    Gohlke, M. K.
    Nunnelley, W. Z.
    Brandebourg, T. D.
    [J]. JOURNAL OF ANIMAL SCIENCE, 2017, 95 : 157 - 158
  • [9] Investigating Interferences of a Whole-Blood Point-of-Care Creatinine Analyzer: Comparison to Plasma Enzymatic and Definitive Creatinine Methods in an Acute-Care Setting
    Straseski, Joely A.
    Lyon, Martha E.
    Clarke, William
    DuBois, Jeffrey A.
    Phelan, Lois A.
    Lyon, Andrew W.
    [J]. CLINICAL CHEMISTRY, 2011, 57 (11) : 1566 - 1573
  • [10] Evaluation of Troponin I, CK-MB and Myoglobin assays on the new Innotrac Aio! point-of-care immunoanalyzer.
    Wahl, HG
    Eckert, E
    Nikolay, T
    Trabandt, N
    Müller, C
    Bohner, J
    [J]. CLINICAL CHEMISTRY, 2002, 48 (06) : A189 - A189