Analytical Performance Evaluation and Comparability of Patient Results Within a Point-of-Care Blood Gas Network

被引:7
|
作者
Oliver, Paloma [1 ]
Fernandez-Calle, Pilar [1 ]
Rico, Natividad [1 ]
Jose Alcaide, Maria [1 ]
Gomez-Rioja, Ruben [1 ]
Buno, Antonio [1 ]
Manuel Iturzaeta, Jose [1 ]
机构
[1] Hosp Univ La Paz, Dept Clin Pathol, Emergency Lab, Madrid, Spain
来源
POINT OF CARE | 2013年 / 12卷 / 03期
关键词
RapidLab; 1265; ABL90; imprecision; bias; biological variation; method comparison; point of care; blood gas; interchangeability of patient results;
D O I
10.1097/POC.0b013e3182a178c9
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Before an analyzer is replaced by another in a clinical laboratory, the analytical performance of the new device and a comparison with a reference or an already established method must be performed before use on patient samples. In point of care, a large numbers of instruments are used, and it is necessary that all are compared against a laboratory instrument. According to these considerations, the aim of this study was to evaluate 23 blood gas analyzers. Methods: Three types of evaluation were conducted. These included a comparability of RapidLab 1265 (current method) and ABL90 (candidate replacement method), an analytical performance evaluation of ABL90 (precision and bias), and a comparability of the same ABL90 device with the other 22 ABL90. Results: The performance of the candidate method is equivalent to that of the current method. In the analytical performance evaluation of ABL90, estimated bias was higher than allowable bias for only 2 parameters: ionized calcium at 0.8 mmol/L and glucose at 14.49 mmol/L. However, both were less than total allowable error. All the ABL90 devices studied were equivalent to the ABL90 that was compared previously with RapidLab 1265. Just 1 device showed an estimated difference greater than allowable. It was replaced by another analyzer that was also evaluated showing satisfactory results. Conclusions: All analyzers showed results within the acceptance criteria. Studies were carried out in the laboratory before blood gas analyzer installation, before use for patient results. This made it possible to detect technical problems earlier, without altering clinical practice.
引用
收藏
页码:144 / 149
页数:6
相关论文
共 50 条
  • [1] Point-of-care blood gas analysers: a performance evaluation
    St-Louis, P
    [J]. CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 139 - 144
  • [2] Multicenter study on the analytical performance of a new point-of-care blood gas analyzer
    Goerlach-Graw, A
    Engel, B
    Halwachs-Baumann, G
    Kost, G
    Kraft, M
    Küster, H
    Miedema, K
    Nagel, D
    Normand, H
    Okorodudu, A
    Olander, B
    Rolinski, B
    Wiley, B
    Wu, J
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A171 - A171
  • [3] Evaluation of the Analytical Performance of the Modified Enterprise Point-of-Care Blood Gas and Electrolyte Analyzer in a Pediatric Hospital
    Agarwal, Shweta
    Shafi, Rabia
    Edwards, Rachel M.
    Collymore, Laura
    Devaraj, Sridevi
    [J]. POINT OF CARE, 2014, 13 (04): : 132 - 136
  • [4] Analytical performance evaluation of bioactive adrenomedullin on point-of-care platform
    Joshi, Netra
    Loeffler, Julia
    Szczesna, Karolina
    Do, Tyler
    Vu, Minh
    Cobb, Jared
    Ngo, Huy
    Nguyen, John
    Carbone, Vincenzo
    Bergmann, Deborah
    Shin, Elizabeth
    Hartmann, Tobias
    Gruson, Damien
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (01) : E13 - E16
  • [5] The evaluation of analytical performance of the precision G™ point-of-care glucometer
    Solnica, B
    Naskalski, JW
    Sieradzki, J
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2001, 39 (12) : 1283 - 1286
  • [6] Performance evaluation of the epoc® point-of-care blood analysis system
    Chen, Yu
    Gorman, Monique
    O'Reilly, Bill
    [J]. CLINICAL BIOCHEMISTRY, 2015, 48 (15) : 1013 - 1013
  • [7] Evaluation of the Element point-of-care blood gas analyzer for use in horses
    Krueger, Clarisa R.
    Hackett, Eileen S.
    Hess, Ann M.
    Mama, Khursheed R.
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (03) : 279 - 285
  • [8] Evaluation of a novel in-line point-of-care blood gas analyser
    Miles, L. F.
    Giraud, K.
    Ferris, R.
    Klein, A. A.
    Martinez, G. C.
    Jenkins, D. P.
    Saulankey, K.
    [J]. ANAESTHESIA, 2016, 71 (09) : 1044 - 1052
  • [9] Analytical evaluation of SENDX 100TM blood gas/electrolyte analyzer for point-of-care testing.
    Liu, X
    Mohammad, AA
    Petersen, JR
    Okorodudu, AO
    [J]. CLINICAL CHEMISTRY, 1997, 43 : 277 - 277
  • [10] Evaluation of the analytical and clinical performance of the Micropoint qLabs INR point-of-care system
    van den Heuvel, J. K.
    Kena, N.
    van Hattum, T.
    van Dolder, D.
    van Wijnen, M.
    [J]. THROMBOSIS RESEARCH, 2019, 176 : 104 - 107