Analytical Performance Evaluation and Comparability of Patient Results Within a Point-of-Care Blood Gas Network

被引:7
|
作者
Oliver, Paloma [1 ]
Fernandez-Calle, Pilar [1 ]
Rico, Natividad [1 ]
Jose Alcaide, Maria [1 ]
Gomez-Rioja, Ruben [1 ]
Buno, Antonio [1 ]
Manuel Iturzaeta, Jose [1 ]
机构
[1] Hosp Univ La Paz, Dept Clin Pathol, Emergency Lab, Madrid, Spain
来源
POINT OF CARE | 2013年 / 12卷 / 03期
关键词
RapidLab; 1265; ABL90; imprecision; bias; biological variation; method comparison; point of care; blood gas; interchangeability of patient results;
D O I
10.1097/POC.0b013e3182a178c9
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Before an analyzer is replaced by another in a clinical laboratory, the analytical performance of the new device and a comparison with a reference or an already established method must be performed before use on patient samples. In point of care, a large numbers of instruments are used, and it is necessary that all are compared against a laboratory instrument. According to these considerations, the aim of this study was to evaluate 23 blood gas analyzers. Methods: Three types of evaluation were conducted. These included a comparability of RapidLab 1265 (current method) and ABL90 (candidate replacement method), an analytical performance evaluation of ABL90 (precision and bias), and a comparability of the same ABL90 device with the other 22 ABL90. Results: The performance of the candidate method is equivalent to that of the current method. In the analytical performance evaluation of ABL90, estimated bias was higher than allowable bias for only 2 parameters: ionized calcium at 0.8 mmol/L and glucose at 14.49 mmol/L. However, both were less than total allowable error. All the ABL90 devices studied were equivalent to the ABL90 that was compared previously with RapidLab 1265. Just 1 device showed an estimated difference greater than allowable. It was replaced by another analyzer that was also evaluated showing satisfactory results. Conclusions: All analyzers showed results within the acceptance criteria. Studies were carried out in the laboratory before blood gas analyzer installation, before use for patient results. This made it possible to detect technical problems earlier, without altering clinical practice.
引用
收藏
页码:144 / 149
页数:6
相关论文
共 50 条
  • [31] Preanalytical Errors in Point-of-Care Testing Auditing Errors of Patient Identification in the Use of Blood Gas Analyzers
    Smith, Natalie A.
    Housley, David G.
    Freedman, Danielle B.
    [J]. POINT OF CARE, 2011, 10 (04): : 189 - 192
  • [32] Point-of-care testing of electrolytes and calcium using blood gas analysers: it is time we trusted the results
    Mirzazadeh, Mehdi
    Morovat, Alireza
    James, Tim
    Smith, Ian
    Kirby, Justin
    Shine, Brian
    [J]. EMERGENCY MEDICINE JOURNAL, 2016, 33 (03) : 181 - 186
  • [33] A Photochemical Dongle for Point-of-Care Evaluation of Blood Lipid Level
    Huang, Xiwei
    Guo, Jinhong
    [J]. 2017 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2017,
  • [34] Evaluation of a whole blood point-of-care coagulation analyzer in dogs
    Conroy, Elizabeth M.
    Lyons, Bridget M.
    Koenig, Amie
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2024,
  • [35] Performance of a new point-of-care analyser for blood gases and electrolytes
    Aw, T. C.
    Liu, Q.
    Poey, F.
    Pang, R. X.
    Tan, S. P.
    Lau, C. S.
    [J]. CLINICA CHIMICA ACTA, 2022, 530 : S16 - S16
  • [36] Evaluation of a point-of-care blood glucose monitor in healthy goats
    Quandt, Jane E.
    Barletta, Michele
    Cornell, Karen K.
    Giguere, Steeve
    Hofmeister, Erik H.
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2018, 28 (01) : 45 - 53
  • [37] Evaluation of analytical and clinical performance of the AFIAS Tn-I plus assay - a new point-of-care
    Pittie, Guillaume
    Lukas, Pierre
    Massart, Madysone
    Cavalier, Etienne
    Le Goff, Caroline
    [J]. ACTA CARDIOLOGICA, 2024, 79 (03) : 351 - 357
  • [38] Comparability of Results between Point-of-Care and Automated Instruments to Measure B-type Natriuretic Peptide
    Shah, Kevin
    Terracciano, Garrett J.
    Jiang, Kevin
    Maisel, Alan S.
    Fitzgerald, Robert L.
    [J]. WESTERN JOURNAL OF EMERGENCY MEDICINE, 2010, 11 (01) : 44 - 48
  • [39] Clinical Performance and Utility of Point-of-Care Lactate Technology in Patient Care Pathways
    Malic, Andrei
    Ntrivalas, Evangelos
    DuBois, Jeffrey
    [J]. POINT OF CARE, 2019, 18 (02): : 62 - 65
  • [40] Guideline will help to verify comparability of patient results within one health care system
    Lehman, Christopher M.
    Bennett, Sterling T.
    [J]. LABMEDICINE, 2008, 39 (10): : 595 - 596