Monte Carlo dose verification for intensity-modulated arc therapy

被引:32
|
作者
Li, XA [1 ]
Ma, LJ [1 ]
Naqvi, S [1 ]
Shih, RP [1 ]
Yu, C [1 ]
机构
[1] Univ Maryland, Dept Radiat Oncol, Baltimore, MD 21201 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2001年 / 46卷 / 09期
关键词
D O I
10.1088/0031-9155/46/9/301
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intensity-modulated are therapy (IMAT), a technique which combines beam rotation and dynamic multileaf collimation, has been implemented in our clinic. Dosimetric errors can be created by the inability of the planning system to accurately account for the effects of tissue inhomogeneities and physical characteristics of the multileaf collimator (MLC). The objective of this study is to explore the use of Monte Carlo (MC) simulation for IMAT dose verification. The BEAM/DOSXYZ Monte Carlo system was implemented to perform dose verification for the IMAT treatment. The implementation includes the simulation of the linac head/MLC (Elekta SL20), the conversion of patient CT images and beam arrangement for 3D dose calculation, the calculation of gantry rotation and leaf motion by a series of static beams and the development of software to automate the entire MC process. The MC calculations were verified by measurements for conventional beam settings. The agreement was within 2%. The IMAT dose distributions generated by a commercial forward planning system (RenderPlan, Elekta) were compared with those calculated by the MC package. For the cases studied, discrepancies of over 10% were found between the MC and the RenderPlan dose calculations. These discrepancies were due in part to the inaccurate dose calculation of the RenderPlan system. The computation time for the IMAT MC calculation was in the range of 20-80 min on 15 Pentium-III computers. The MC method was also useful in verifying the beam apertures used in the IMAT treatments.
引用
收藏
页码:2269 / 2282
页数:14
相关论文
共 50 条
  • [21] Clinical Implementation of Intensity-Modulated Arc Therapy
    Shepard, David M.
    Cao, Daliang
    IMRT IGRT SBRT- ADVANCES IN THE TREATMENT PLANNING AND DELIVERY OF RADIOTHERAPY, 2011, 43 : 80 - 98
  • [22] Dose integration of intensity-modulated arc therapy and interstitial brachytherapy of cervix cancer
    Frohlich, G.
    Vizkeleti, J.
    Nguyen, N. Anhhong
    Major, T.
    Polgar, C.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1178 - S1179
  • [23] Gel-layer dosimetry for dose verification in intensity-modulated radiation therapy
    Tomatis, S.
    Carrara, M.
    Gambarini, G.
    Marchesini, R.
    Valente, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 506 - 509
  • [24] Verification of an optimizer algorithm by the beam delivery evaluation of intensity-modulated arc therapy plans
    Pocza, Tamas
    Szegedi, Domonkos
    Major, Tibor
    Pesznyak, Csilla
    RADIOLOGY AND ONCOLOGY, 2021, 55 (04) : 508 - 515
  • [25] Dependences of mucosal dose on photon beams in head-and-neck intensity-modulated radiation therapy: a Monte Carlo study
    Chow, James C. L.
    Owrangi, Amir M.
    MEDICAL DOSIMETRY, 2012, 37 (02) : 195 - 200
  • [26] Comparison of Monte Carlo and analytical dose computations for intensity modulated proton therapy
    Yepes, Pablo
    Adair, Antony
    Grosshans, David
    Mirkovic, Dragan
    Poenisch, Falk
    Titt, Uwe
    Wang, Qianxia
    Mohan, Radhe
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (04):
  • [27] Verification system for intensity-modulated radiation therapy with scintillator
    Ando, Yasuharu
    Miki, Kentaro
    Araki, Jun
    Tsuneda, Masato
    Kiriu, Hiroshi
    Nishio, Teiji
    Nagata, Yasushi
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2021, 44 (01) : 9 - 21
  • [28] Intensity-modulated radiation therapy dose verification using fluence and portal imaging device
    Sumida, Iori
    Yamaguchi, Hajime
    Das, Indra J.
    Kizaki, Hisao
    Aboshi, Keiko
    Tsujii, Mari
    Yamada, Yuji
    Suzuki, Osamu
    Seo, Yuji
    Isohashi, Fumiaki
    Ogawa, Kazuhiko
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 259 - 271
  • [29] In vivo dosimetric verification of intensity-modulated radiation therapy
    Chytyk-Praznik, K.
    Van Uytven, E.
    Van Beek, T.
    McCurdy, B. M. C.
    MEDICAL PHYSICS, 2012, 39 (07) : 4629 - 4629
  • [30] Dose integration method of intensity-modulated arc therapy and HDR brachytherapy of prostate cancer
    Frohlich, G.
    Agoston, P.
    Jorgo, K.
    Polgar, C.
    Major, T.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S804 - S804