A novel approach to determine primary stability of acetabular press-fit cups

被引:12
|
作者
Weissmann, Volker [1 ,2 ]
Boss, Christian [3 ]
Bader, Rainer [2 ]
Hansmann, Harald [1 ]
机构
[1] Univ Appl Sci Technol Business & Design, Fac Engn, Philipp Muller Str 14, D-23966 Wismar, Germany
[2] Rostock Univ Med, Dept Orthoped, Biomech & Implant Technol Res Lab, Doberaner Str 142, D-18057 Rostock, Germany
[3] Inst Polymer Technol eV, Alter Holzhafen 19, D-23966 Wismar, Germany
关键词
Ti6A14V; Selective laser melting; Press-fit; Primary stability; TOTAL HIP-ARTHROPLASTY; IMPROVED PRIMARY FIXATION; POROUS TITANIUM; MECHANICAL-PROPERTIES; SURFACE ARCHITECTURE; INITIAL STABILITY; DESIGN; STEM; IMPLANTS; LASER;
D O I
10.1016/j.jmbbm.2018.01.015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Today hip cups are used in a large variety of design variants and in increasing numbers of units. Their development is steadily progressing. In addition to conventional manufacturing methods for hip cups, additive methods, in particular, play an increasingly important role as development progresses. The present paper describes a modified cup model developed based on a commercially available press-fit cup (Allofit 54/JJ). The press-fit cup was designed in two variants and manufactured using selective laser melting (SLM). Variant 1 (Ti) was modeled on the Alio& cup using an adapted process technology. Variant 2 (Ti-S) was provided with a porous load bearing structure on its surface. In addition to the typical (complete) geometry, both variants were also manufactured and tested in a reduced shape where only the press-fit area was formed. To assess the primary stability of the press-fit cups in the artificial bone cavity, pull-out and lever-out tests were carried out. Exact fit conditions and two-millimeter press-fit were investigated. The closed-cell PU foam used as an artificial bone cavity was mechanically characterized to exclude any influence on the results of the investigation. The pull-out forces of the Ti-variant (complete-526 N, reduced-468 N) and the Ti-S variant (complete-548 N, reduced-526 N) as well as the lever-out moments of the Ti-variant (complete-10 Nm, reduced-9.8 Nm) and the Ti-S variant (complete-9 Nm, reduced-7.9 N) show no significant differences in the results between complete and reduced cups. The results show that the use of reduced cups in a press-fit design is possible within the scope of development work.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 50 条
  • [31] Press-fit acetabular cup fixation: principles and testing
    Macdonald, W
    Carlsson, LV
    Charnley, GJ
    Jacobsson, CM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H1) : 33 - 39
  • [32] Are additional screws required for press-fit fixation of cementless acetabular cups? A systematic review and meta-analysis
    Shenghui Ni
    Peng Luo
    Lei Guo
    Tianlong Jiang
    Journal of Orthopaedics and Traumatology, 2022, 23
  • [33] Are additional screws required for press-fit fixation of cementless acetabular cups? A systematic review and meta-analysis
    Ni, Shenghui
    Luo, Peng
    Guo, Lei
    Jiang, Tianlong
    JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2022, 23 (01)
  • [34] Risk factors for unsuccessful acetabular press-fit fixation at primary total hip arthroplasty
    Brulc, U.
    Antolic, V.
    Mavcic, B.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2017, 103 (07) : 993 - 997
  • [35] Acetabular Cup Seating Impact Sensing for Press-Fit Acetabular Cup Fixation
    Tang, L.
    Tennant, J. K.
    Forster, A.
    Reddy, M.
    2016 12TH IEEE/ASME INTERNATIONAL CONFERENCE ON MECHATRONIC AND EMBEDDED SYSTEMS AND APPLICATIONS (MESA), 2016,
  • [36] Acetabular revision in total hip replacement with a press-fit cup
    Elke, R
    Berli, B
    Wagner, A
    Morscher, EW
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (08): : 1114 - 1119
  • [37] Press-fit versus threaded acetabular cups in total hip arthroplasty : Functional and radiological results after five years
    Ellenrieder, Martin
    Bader, Rainer
    Bergschmidt, Philipp
    Mittelmeier, Wolfram
    ACTA ORTHOPAEDICA BELGICA, 2016, 82 (01): : 60 - 67
  • [38] Experimental analysis of insertion torques and forces of threaded and press-fit acetabular cups by means of ex vivo and in vivo measurements
    Vogel, Danny
    Rathay, Andreas
    Teufel, Stephanie
    Ellenrieder, Martin
    Zietz, Carmen
    Sander, Manuela
    Bader, Rainer
    ACTA OF BIOENGINEERING AND BIOMECHANICS, 2017, 19 (03) : 155 - 163
  • [39] Load transfer and fixation mode of press-fit acetabular sockets
    Widmer, KH
    Zurfluh, B
    Morscher, EW
    JOURNAL OF ARTHROPLASTY, 2002, 17 (07): : 926 - 935
  • [40] Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems
    Markel, DC
    Hora, N
    Grimm, M
    INTERNATIONAL ORTHOPAEDICS, 2002, 26 (02) : 72 - 75