Traceable calibration of a broadband seismometer down to 5 mHz

被引:4
|
作者
Shimoda, Tomofumi [1 ]
Kokuyama, Wataru [1 ]
Nozato, Hideaki [1 ]
机构
[1] Natl Metrol Inst Japan, Natl Inst Adv Ind Sci & Technol, 1-1-1 Umezono, Tsukuba, Ibaraki 3058563, Japan
关键词
seismometer; calibration; traceability; low-frequency vibration; tilt coupling;
D O I
10.1088/1361-6501/ac9077
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Broadband seismometers have been widely employed in geophysical observation and disaster prevention. To ensure the reliability of the measurement, traceable calibration of their sensitivities following ISO 16063-11 down to 8 mHz is required. However, the calibration accuracy generally degrades below 0.1 Hz due to tilt-coupling error. In this paper, two tilt-coupling correction methods are investigated: one is a dynamic tilt measurement method using an autocollimator, and the other is a comparison method using a reference accelerometer. These methods were introduced to the primary calibration system to calibrate a seismometer between 5 mHz and 200 Hz. As a result, the tilt-coupling error was corrected with an estimated uncertainty of less than 2% at 5 mHz, and the bandpass response of the seismometer sensitivity was confirmed with the traceable method.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Traceable calibration of a phase noise standard
    Bernier, Laurent-Guy
    Stalder, Daniel
    Morel, Jacques
    Kucera, Jakub
    Dahinden, Stefan
    PROCEEDINGS OF THE 48TH ANNUAL PRECISE TIME AND TIME INTERVAL SYSTEMS AND APPLICATIONS MEETING, 2017, : 335 - 342
  • [22] Traceable Calibration of pH Measurement Devices
    Spitzer, Petra
    Adel, Beatrice
    Werner, Barbara
    Jehnert, Diana
    TM-TECHNISCHES MESSEN, 2010, 77 (03) : 156 - 161
  • [23] Field Test of Broadband Fiber Optic Interferometric Seismometer
    Zhang, Wentao
    Huang, Wenzhu
    Wang, Yibo
    Yao, Yi
    Zhang, Jianxiang
    Lv, Bing
    Li, Fang
    2020 ASIA COMMUNICATIONS AND PHOTONICS CONFERENCE (ACP) AND INTERNATIONAL CONFERENCE ON INFORMATION PHOTONICS AND OPTICAL COMMUNICATIONS (IPOC), 2020,
  • [24] Development of a Compact Broadband Ocean-Bottom Seismometer
    Shinohara, Masanao
    Yamada, Tomoaki
    Shiobara, Hajime
    Yamashita, Yusuke
    SEISMOLOGICAL RESEARCH LETTERS, 2021, 92 (06) : 3610 - 3625
  • [25] Traceable High Impedance Calibration Standards
    Haase, Martin
    Hoffmann, Karel
    Hudlicka, Martin
    2016 87TH ARFTG MICROWAVE MEASUREMENT CONFERENCE (ARFTG), 2016,
  • [26] Uncertainty estimates in broadband seismometer sensitivities using microseisms
    Ringler, A. T.
    Storm, T.
    Gee, L. S.
    Hutt, C. R.
    Wilson, D.
    JOURNAL OF SEISMOLOGY, 2015, 19 (02) : 317 - 327
  • [27] Uncertainty estimates in broadband seismometer sensitivities using microseisms
    A. T. Ringler
    T. Storm
    L. S. Gee
    C. R. Hutt
    D. Wilson
    Journal of Seismology, 2015, 19 : 317 - 327
  • [28] AN ELECTROMAGNETIC CALIBRATION COIL FOR A WILLMORE MK 2 SEISMOMETER
    STEWART, I
    SUTTON, DJ
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1967, 44 (12): : 1035 - &
  • [29] EFFECT OF INTERPOSED SKIN AT DOPPLER FLOW ESTIMATION AT 5 MHZ AND 10 MHZ - WITH A DESCRIPTION OF A CALIBRATION DEVICE
    FORSBERG, L
    OLIN, T
    ACTA RADIOLOGICA-DIAGNOSIS, 1980, 21 (2A): : 203 - 208
  • [30] Traceable measurements and calibration: a primer on uncertainty analysis
    Csavina, Janae
    Roberti, Joshua A.
    Taylor, Jeffrey R.
    Loescher, Henry W.
    ECOSPHERE, 2017, 8 (02):