Collimator design for a multipinhole brain SPECT insert for MRI

被引:6
|
作者
Van Audenhaege, Karen [1 ]
Van Holen, Roel [1 ]
Vanhove, Christian [1 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Ghent Univ iMinds Med IT, MEDISIP IBiTech, Dept Elect & Informat Syst, B-9000 Ghent, Belgium
基金
欧盟第七框架计划;
关键词
SPECT; collimator; production; single photon emission computed tomography; optimization; HIGH-RESOLUTION SPECT; SYSTEM; PET; SIMULATION; DETECTOR; MODALITY; COMPACT; IMAGES;
D O I
10.1118/1.4934371
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Brain single photon emission computed tomography (SPECT) imaging is an important clinical tool, with unique tracers for studying neurological diseases. Nowadays, most commercial SPECT systems are combined with x-ray computed tomography (CT) in so-called SPECT/CT systems to obtain an anatomical background for the functional information. However, while CT images have a high spatial resolution, they have a low soft-tissue contrast, which is an important disadvantage for brain imaging. Magnetic resonance imaging (MRI), on the other hand, has a very high soft-tissue contrast and does not involve extra ionizing radiation. Therefore, the authors designed a brain SPECT insert that can operate inside a clinical MRI. Methods: The authors designed and simulated a compact stationary multipinhole SPECT insert based on digital silicon photomultiplier detector modules, which have shown to be MR-compatible and have an excellent intrinsic resolution (0.5 mm) when combined with a monolithic 2 mm thick LYSO crystal. First, the authors optimized the different parameters of the SPECT system to maximize sensitivity for a given target resolution of 7.2 mm in the center of the field-of-view, given the spatial constraints of the MR system. Second, the authors performed noiseless simulations of two multipinhole configurations to evaluate sampling and reconstructed resolution. Finally, the authors performed Monte Carlo simulations and compared the SPECT insert with a clinical system with ultrahigh-resolution (UHR) fan beam collimators, based on contrast-to-noise ratio and a visual comparison of a Hoffman phantom with a 9 mm cold lesion. Results: The optimization resulted in a stationary multipinhole system with a collimator radius of 150.2 mm and a detector radius of 172.67 mm, which corresponds to four rings of 34 diSPM detector modules. This allows the authors to include eight rings of 24 pinholes, which results in a system volume sensitivity of 395 cps/MBq. Noiseless simulations show sufficient axial sampling (in a Defrise phantom) and a reconstructed resolution of 5.0 mm (in a cold-rod phantom). The authors compared the 24-pinhole setup with a 34-pinhole system (with the same detector radius but a collimator radius of 156.63 mm) and found that 34 pinholes result in better uniformity but a worse reconstruction of the cold-rod phantom. The authors also compared the 24-pinhole system with a clinical triple-head UHR fan beam system based on contrast-to-noise ratio and found that the 24-pinhole setup performs better for the 6 mm hot and the 16 mm cold lesions and worse for the 8 and 10 mm hot lesions. Finally, the authors reconstructed noisy projection data of a Hoffman phantom with a 9 mm cold lesion and found that the lesion was slightly better visible on the multipinhole image compared to the fan beam image. Conclusions: The authors have optimized a stationary multipinhole SPECT insert for MRI and showed the feasibility of doing brain SPECT imaging inside a MRI with an image quality similar to the best clinical SPECT systems available. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:6679 / 6689
页数:11
相关论文
共 50 条
  • [11] SPECT/MRI INSERT Compatibility: Assessment, Solutions, and Design Guidelines
    Carminati, Marco
    Montagnani, Giovanni L.
    Occhipinti, Michele
    Kuehne, Andre
    Niendorf, Thoralf
    Nagy, Kalman
    Nagy, Andras
    Czeller, Miklos
    Fiorini, Carlo
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (04) : 369 - 379
  • [12] Evaluation of a partial ring design for the INSERT SPECT/MRI system
    Kell Erlandsson
    Debora Slavado
    Alexandre Bousse
    Brian Hutton
    EJNMMI Physics, 2 (Suppl 1)
  • [13] INSERT: A Novel Clinical Scanner for Simultaneous SPECT/MRI Brain Studies
    Hutton, B. F.
    Erlandsson, K.
    Salvado, D.
    Occhipinti, M.
    Papp, Z.
    Tolgyesi, B.
    Willems, J.
    Piemonte, C.
    Carminati, M.
    Bukki, T.
    Legrady, G.
    Kuehne, A.
    Nyitrai, Z.
    Niendorf, T.
    Van Mullekom, P.
    Waiczies, H.
    Nagy, K.
    De Francesco, I.
    Mathe, D.
    Ottobrini, L.
    Short, S. C.
    Fiorini, C.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [14] Design of an Optimized Multi-pinhole Collimator for Brain SPECT
    Chen, L.
    Mok, G. S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S212 - S212
  • [15] Theoretical Bounds and System Design for Multipinhole SPECT
    Nillius, Peter
    Danielsson, Mats
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (07) : 1390 - 1400
  • [16] Collimator Design in SPECT, an Optimisation Tool
    Fuin, Niccolo
    Bousse, Alexandre
    Pedemonte, Stefano
    Arridge, Simon
    Ourselin, Sebastien
    Hutton, Brian
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3149 - 3154
  • [17] The design and manufacture of an annular variable-focusing collimator for highsensitivity brain SPECT
    Ouyang, Jinsong
    El Fakhri, Georges
    Xia, Weishi
    Kijewski, Marie Foley
    Genna, Sebastian
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) : 2613 - 2618
  • [18] Multi-pinhole collimator design in different numbers of projections for brain SPECT
    Huang, Wenbo
    Mok, Greta S. P.
    FRONTIERS IN MEDICINE, 2023, 10
  • [19] Phantom Measurements and Simulations of Cardiac and Brain Studies Using a Multipinhole Collimator with 20 Apertures
    Bowen, Jason D.
    Huang, Qiu
    Gullberg, Grant T.
    Seo, Youngho
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3417 - 3421
  • [20] The Design of Optimal Multipinhole Collimators for a Seamless SPECT Detector Ring
    Rittenbach, Andrew J.
    Xu, Jingyan
    Hugg, James
    Tsui, Benjamin M. W.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3402 - 3405