Collimator design for a multipinhole brain SPECT insert for MRI

被引:6
|
作者
Van Audenhaege, Karen [1 ]
Van Holen, Roel [1 ]
Vanhove, Christian [1 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Ghent Univ iMinds Med IT, MEDISIP IBiTech, Dept Elect & Informat Syst, B-9000 Ghent, Belgium
基金
欧盟第七框架计划;
关键词
SPECT; collimator; production; single photon emission computed tomography; optimization; HIGH-RESOLUTION SPECT; SYSTEM; PET; SIMULATION; DETECTOR; MODALITY; COMPACT; IMAGES;
D O I
10.1118/1.4934371
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Brain single photon emission computed tomography (SPECT) imaging is an important clinical tool, with unique tracers for studying neurological diseases. Nowadays, most commercial SPECT systems are combined with x-ray computed tomography (CT) in so-called SPECT/CT systems to obtain an anatomical background for the functional information. However, while CT images have a high spatial resolution, they have a low soft-tissue contrast, which is an important disadvantage for brain imaging. Magnetic resonance imaging (MRI), on the other hand, has a very high soft-tissue contrast and does not involve extra ionizing radiation. Therefore, the authors designed a brain SPECT insert that can operate inside a clinical MRI. Methods: The authors designed and simulated a compact stationary multipinhole SPECT insert based on digital silicon photomultiplier detector modules, which have shown to be MR-compatible and have an excellent intrinsic resolution (0.5 mm) when combined with a monolithic 2 mm thick LYSO crystal. First, the authors optimized the different parameters of the SPECT system to maximize sensitivity for a given target resolution of 7.2 mm in the center of the field-of-view, given the spatial constraints of the MR system. Second, the authors performed noiseless simulations of two multipinhole configurations to evaluate sampling and reconstructed resolution. Finally, the authors performed Monte Carlo simulations and compared the SPECT insert with a clinical system with ultrahigh-resolution (UHR) fan beam collimators, based on contrast-to-noise ratio and a visual comparison of a Hoffman phantom with a 9 mm cold lesion. Results: The optimization resulted in a stationary multipinhole system with a collimator radius of 150.2 mm and a detector radius of 172.67 mm, which corresponds to four rings of 34 diSPM detector modules. This allows the authors to include eight rings of 24 pinholes, which results in a system volume sensitivity of 395 cps/MBq. Noiseless simulations show sufficient axial sampling (in a Defrise phantom) and a reconstructed resolution of 5.0 mm (in a cold-rod phantom). The authors compared the 24-pinhole setup with a 34-pinhole system (with the same detector radius but a collimator radius of 156.63 mm) and found that 34 pinholes result in better uniformity but a worse reconstruction of the cold-rod phantom. The authors also compared the 24-pinhole system with a clinical triple-head UHR fan beam system based on contrast-to-noise ratio and found that the 24-pinhole setup performs better for the 6 mm hot and the 16 mm cold lesions and worse for the 8 and 10 mm hot lesions. Finally, the authors reconstructed noisy projection data of a Hoffman phantom with a 9 mm cold lesion and found that the lesion was slightly better visible on the multipinhole image compared to the fan beam image. Conclusions: The authors have optimized a stationary multipinhole SPECT insert for MRI and showed the feasibility of doing brain SPECT imaging inside a MRI with an image quality similar to the best clinical SPECT systems available. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:6679 / 6689
页数:11
相关论文
共 50 条
  • [21] A PERFORMANCE EVALUATION OF AN ASTIGMATIC COLLIMATOR FOR BRAIN SPECT
    RICHTER, JA
    MARTI, JM
    DOMPER, M
    MALDONADO, A
    GARCIA, MJ
    RUBIO, L
    ARBIZU, J
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1992, 19 (08): : 635 - 635
  • [22] Onboard functional and molecular imaging: A design investigation for robotic multipinhole SPECT
    Bowsher, James
    Yan, Susu
    Roper, Justin
    Giles, William
    Yin, Fang-Fang
    MEDICAL PHYSICS, 2014, 41 (01)
  • [23] Design and evaluation of an adaptive multipinhole collimator for high-performance clinical and preclinical imaging
    Si, Chinhong
    Mok, Greta S. P.
    Chen, Ling
    Tsui, Benjamin M. W.
    NUCLEAR MEDICINE COMMUNICATIONS, 2016, 37 (03) : 313 - 321
  • [24] Acquisition Correction and Reconstruction for a Clinical SPECT/MRI Insert
    Morahan, Ashley J.
    Erlandsson, Kjell
    D'Adda, Ilenia
    Carminati, Marco
    Savi, Annarita
    Moresco, Rosa-Maria
    Nagy, Kalman
    Nyitrai, Zoltan
    Tolgyesi, Botond
    Salvado, Debora
    Van Mullekom, Pieter
    Fiorini, Carlo E.
    Hutton, Brian F.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [25] Design and simulation of a full-ring multi-lofthole collimator for brain SPECT
    Van Audenhaege, Karen
    Vandenberghe, Stefaan
    Deprez, Karel
    Vandeghinste, Bert
    Van Holen, Roel
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (18): : 6317 - 6336
  • [26] Preliminary Brain SPECT Multi-Pinhole Collimator Mechanical Design for DaTscan Imaging
    He, Yulun
    Konik, Arda
    De Beenhouwer, Jan
    Fromme, Timothy J.
    Lindsay, Clifford
    Kalluri, Kesava
    Li, Xin
    Momsen, Neil C.
    Furenlid, Lars R.
    King, Michael A.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [27] Design of a novel slit-slat collimator system for SPECT imaging of the human brain
    Mahmood, S. T.
    Erlandsson, K.
    Cullum, I.
    Hutton, B. F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (11): : 3433 - 3449
  • [28] Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT
    Van Audenhaege, Karen
    Van Holen, Roel
    Deprez, Karel
    Karp, Joel S.
    Metzler, Scott
    Vandenberghe, Stefaan
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4393 - 4397
  • [29] The design and evaluation of optimal multipinhole collimators for a SPECT-MR system
    Rittenbach, Andrew
    Xu, Jingyan
    Liu, Ang
    Hugg, James
    Tsui, Benjamin
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [30] Optimization of multipinhole arrangements for quantitative mouse brain SPECT by Monte Carlo simulation
    Cao, ZX
    Bal, G
    Accorsi, R
    Acton, PD
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2470 - 2474