Phantom Measurements and Simulations of Cardiac and Brain Studies Using a Multipinhole Collimator with 20 Apertures

被引:0
|
作者
Bowen, Jason D. [1 ]
Huang, Qiu [2 ]
Gullberg, Grant T. [3 ]
Seo, Youngho [1 ]
机构
[1] Univ Calif San Francisco, San Francisco, CA 94143 USA
[2] Shanghai Jiao Tong Univ, Shanghai 200030, Peoples R China
[3] Lawrence Berkeley Natl Lab, Berkeley, CA 94720 USA
关键词
Single photon emission computed tomography (SPECT); multipinhole; myocardial perfusion imaging (MPI); dual-isotope imaging;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
SPECT, despite its success, has an uncertain future This is partly due to the continued use of parallel hole collimators. These collimators possess poor geometric sensitivity which leads to increased study times and higher dose requirements. In recent years there has been activity investigating dedicated SPECT cameras which attempt to build on the strengths of SPECT while addressing its weaknesses. Here we present measurements of a cardiac and Hoffman brain phantom (Data Spectrum, Hillsborough, NC) performed with a pair of focusing 20-pinhole collimators with 7.5 mm diameter tungsten apertures mounted on an Infinia Hawkeye 4 dual-head SPECT/CT system (GE Uealthcare, Chalfont St. Giles, UK). Comparisons are made with measurements taken with LEUR parallelhole collimators. We also present an assessment of point source sensitivity and image resolution. The cardiac and brain phantoms contained 3.8 mCi and 1.5 mCi of 99mTc-pertechnetate (140 keV), respectively. The cardiac phantom was imaged with six 300 s views over 180 degrees and the brain phantom was imaged with eight 300 s views over 360 degrees. The gantry orbit was circular with a radius of rotation between 22.5 cm and 25.0 cm. Shortened study times were simulated by scaling the projection data and adding Poisson noise. The multipinhole projection data were then reconstructed with 300 iterations of MLEM. We find that despite the relatively large pinhole sizes we obtain image quality comparable to that obtained with LEHR collimators at a fraction the acquisition time. For the cardiac phantom good quality images are obtained in 10 minutes, while for the brain phantom comparable image quality is attained in 10 minutes. The measured point source sensitivity near the central field of view is 7.4 x 10(-4), and optimal resolution is between 9 mm and 13 mm. These results indicate that imaging performance approaching that capable with dedicated SPECT cameras may be possible with these collimators while leveraging the existing SPECT install base.
引用
收藏
页码:3417 / 3421
页数:5
相关论文
共 50 条
  • [1] Multipinhole collimator with 20 apertures for a brain SPECT application
    Lee, Tzu-Cheng
    Ellin, Justin R.
    Huang, Qiu
    Shrestha, Uttam
    Gullberg, Grant T.
    Seo, Youngho
    MEDICAL PHYSICS, 2014, 41 (11)
  • [2] IMPROVED ACCURACY OF EJECTION FRACTION MEASUREMENTS USING A COMMERCIAL CARDIAC PHANTOM
    TONGE, CM
    FRENCH, AP
    PEAKE, JCF
    HORTON, PW
    NUCLEAR MEDICINE COMMUNICATIONS, 1986, 7 (01) : 65 - 69
  • [3] Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom
    Garcia-Garduno, Olivia A.
    Larraga-Gutierrez, Jose M.
    Rodriguez-Villafuerte, Mercedes
    Martinez-Davalos, Arnulfo
    Celis, Miguel A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 (02) : 250 - 253
  • [4] RELIABILITY OF MEASUREMENTS FROM TWO-DIMENSIONAL ECHOCARDIOGRAPHIC IMAGES USING A CARDIAC PHANTOM
    ZOTZ, R
    MEYER, CR
    DELP, EJ
    BOOKSTEIN, FL
    BUDA, AJ
    CLINICAL RESEARCH, 1985, 33 (02): : A240 - A240
  • [5] COLLIMATOR AND FILTER SELECTION - SIMULATING TC-99M CEREBRAL PERFUSION STUDIES WITH A 3-DIMENSIONAL HOFFMAN BRAIN PHANTOM
    KIM, HJ
    KARP, JS
    KUNG, HF
    MOZLEY, PD
    ALAVI, A
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05) : P195 - P195
  • [6] Studies on the crosstalk rate in simultaneous myocardial imaging with dual radionuclides using cardiac focusing collimator
    Kunimoto, K.
    Abe, S.
    Fujita, N.
    Honda, M.
    Matsuzawa, S.
    Mukumoto, R.
    Odagawa, T.
    Kato, K.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S868 - S868
  • [7] Optimization Of Technical Aspects For Using The Innovative Stepbrain Brain Phantom In Nuclear Medicine Imaging Simulations
    Ciotola, D.
    Pecchia, G.
    Alfano, B.
    Del Vasto, C.
    Di Giorgio, E.
    Imbimbo, S.
    Lanotte, C.
    Mansi, L.
    Restaino, A.
    Piccolo, S.
    Pirozzi, M. A.
    Quarantelli, M.
    Silva, C.
    Spadafora, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S941 - S942
  • [8] Molecular imaging using the theranostic agent 197(m)Hg: phantom measurements and Monte Carlo simulations
    Robert Freudenberg
    Rudi Apolle
    Martin Walther
    Holger Hartmann
    Jörg Kotzerke
    EJNMMI Physics, 5
  • [9] Molecular imaging using the theranostic agent 197(m)Hg: phantom measurements and Monte Carlo simulations
    Freudenberg, Robert
    Apolle, Rudi
    Walther, Martin
    Hartmann, Holger
    Kotzerke, Joerg
    EJNMMI PHYSICS, 2018, 5
  • [10] Image reconstruction for diagnosis and prognosis of breast cancer using fluorescence measurements: Phantom studies
    Roy, R
    Godavarty, A
    Thompson, AB
    Sevick-Muraca, EM
    Optical Tomography and Spectroscopy of Tissue VI, 2005, 5693 : 203 - 209