In situ cryo-FIB/SEM Specimen Preparation Using the Waffle Method

被引:5
|
作者
Klykov, Oleg [1 ]
Bobe, Daija [1 ]
Paraan, Mohammadreza [1 ]
Johnston, Jake D. [1 ,2 ]
Potter, Clinton S. [1 ,3 ]
Carragher, Bridget [1 ,3 ]
Kopylov, Mykhailo [1 ]
Noble, Alex J. [1 ]
机构
[1] New York Struct Biol Ctr, Simons Electron Microscopy Ctr, Natl Ctr Insitu Tomog Ultramicroscopy, New York, NY 10027 USA
[2] Columbia Univ, Dept Physiol & Cellular Biophys, New York, NY USA
[3] Columbia Univ, Dept Biochem & Mol Biophys, New York, NY USA
来源
BIO-PROTOCOL | 2022年 / 12卷 / 21期
关键词
FIB; -SEM; HPF; Waffle Method; cryo-ET; Sample preparation; In situ; AutoTEM Cryo; ELECTRON-MICROSCOPY; EM;
D O I
10.21769/BioProtoc.4544
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Cryo-focused ion beam (FIB) milling of vitrified specimens is emerging as a powerful method for in situ specimen preparation. It allows for the preservation of native and near-native conditions in cells, and can reveal the molecular structure of protein complexes when combined with cryo-electron tomography (cryo-ET) and sub-tomogram averaging. Cryo-FIB milling is often performed on plunge-frozen specimens of limited thickness. However, this approach may have several disadvantages, including low throughput for cells that are small, or at low concentration, or poorly distributed across accessible areas of the grid, as well as for samples that may adopt a preferred orientation. Here, we present a detailed description of the "Waffle Method" protocol for vitrifying thick specimens followed by a semi-automated milling procedure using the Thermo Fisher Scientific (TFS) Aquilos 2 cryo-FIB/scanning electron microscope (SEM) instrument and AutoTEM Cryo software to produce cryo-lamellae. With this protocol, cryolamellae may be generated from specimens, such as microsporidia spores, yeast, bacteria, and mammalian cells, as well as purified proteins and protein complexes. An experienced lab can perform the entire protocol presented here within an 8-hour working day, resulting in two to three cryo-lamellae with target thicknesses of 100-200 nm and dimensions of approximately 12 mu m width and 15-20 mu m length. For cryo-FIB/SEMs with particularly lowcontamination chambers, the protocol can be extended to overnight milling, resulting in up to 16 cryo-lamellae in
引用
收藏
页数:16
相关论文
共 50 条
  • [21] Specimen Preparation in Atom Probe by FIB-Based Method
    Hanaoka, Yuya
    Yamamoto, Takuya
    Mayama, Norihito
    Kaito, Takashi
    Owari, Masanori
    [J]. E-JOURNAL OF SURFACE SCIENCE AND NANOTECHNOLOGY, 2011, 9 : 348 - 351
  • [22] Tuneable in-situ nanoCT workflow using FIB/SEM
    Barabasova, Paulina
    Kovacova, Veronika
    Stejskal, Pavel
    Uncovsky, Marek
    Valterova, Eva
    Vystavel, Tomas
    [J]. ULTRAMICROSCOPY, 2021, 225
  • [23] Cryo-FIB-Preparation and Cryo-SEM Investigation of Gold Nanolayers Used as Absorber for Laser Welding of Polymer Foils
    Neuber, M.
    Cismak, A.
    Heilmann, A.
    [J]. LAYERED NANOSTRUCTURES - POLYMERS WITH IMPROVED PROPERTIES, 2010, 294-I : 11 - 18
  • [24] Preparation and characterization of silicon nanowires using SEM/FIB and TEM
    Banerjee, Sayanti
    Muehle, Uwe
    Loeffler, Markus
    Heinzig, Andre
    Trommer, Jens
    Zschech, Ehrenfried
    [J]. INTERNATIONAL JOURNAL OF MATERIALS RESEARCH, 2015, 106 (07) : 697 - 702
  • [25] A NEW AND SIMPLE METHOD FOR CRYO-EM SPECIMEN PREPARATION
    Zhou, Jingjing
    Xu, Hongyi
    Carroni, Marta
    Lebrette, Hugo
    Wallden, Karin
    Hogbom, Martin
    Zou, Xiaodong
    [J]. ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2019, 75 : E82 - E82
  • [26] Cryogenic vitrification and 3D serial sectioning using high resolution cryo-FIB SEM technology for brine-filled grain boundaries in halite: first results
    Desbois, G.
    Urai, J. L.
    Burkhardt, C.
    Drury, M. R.
    Hayles, M.
    Humbel, B.
    [J]. GEOFLUIDS, 2008, 8 (01) : 60 - 72
  • [27] In-situ TEM straining experiments of Al films on polyimide using a novel FIB design for specimen preparation
    Dehm, G.
    Legros, M.
    Heiland, B.
    [J]. JOURNAL OF MATERIALS SCIENCE, 2006, 41 (14) : 4484 - 4489
  • [28] In situ lift-out using a FIB-SEM system
    Langford, RM
    Clinton, C
    [J]. MICRON, 2004, 35 (07) : 607 - 611
  • [29] In-situ TEM straining experiments of Al films on polyimide using a novel FIB design for specimen preparation
    G. Dehm
    M. Legros
    B. Heiland
    [J]. Journal of Materials Science, 2006, 41 : 4484 - 4489
  • [30] Sample preparation and image registration for correlative cryo-FM and cryo-FIB-SEM of plunge-frozen mammalian cells
    Scher, Nadav
    Rechav, Katya
    Paul-Gilloteaux, Perrine
    Avinoam, Ori
    [J]. STAR PROTOCOLS, 2022, 3 (01):