Tuneable in-situ nanoCT workflow using FIB/SEM

被引:0
|
作者
Barabasova, Paulina [1 ]
Kovacova, Veronika [1 ,2 ]
Stejskal, Pavel [1 ]
Uncovsky, Marek [1 ]
Valterova, Eva [1 ,3 ]
Vystavel, Tomas [1 ]
机构
[1] Thermo Fisher Sci, Vlastimila Pecha 12, Brno 62700, Czech Republic
[2] Masaryk Univ, Fac Sci, Dept Anthropol, Lab Morphol & Forens Anthropol, Kotlarska 2, Brno 61137, Czech Republic
[3] Brno Univ Technol, Fac Elect Engn & Commun, Dept Biomed Engn, Tech 12, Brno 61600, Czech Republic
关键词
Computed tomography; electron microscope; X-ray absorption; Timepix detector; radiograph; 3D reconstruction; RAY COMPUTED-TOMOGRAPHY;
D O I
10.1016/j.ultramic.2021.113283
中图分类号
TH742 [显微镜];
学科分类号
摘要
Inspired by the standard computed tomography, a new method of 3D X-ray imaging embedded in FIB-SEM microscope is proposed. The unique combination of TEM-like specimen stage enabling in lens STEM detection (referred to as CompuStage), nanomanipulator (referred to as EasyLift) facilitating in-situ sample transfer from bulk sample to TEM-like stage and pixelated in-situ Timepix X-ray detector in Helios G4 FX FIB-SEM system offers an unprecedented workflow. Motivated by common circular CT scan known from microCT world, the object under study is placed on CompuStage rod which enables two possible rotation (in TEM/SEM terminology called tilt) movements - alpha-tilt - rotation of the CompuStage rod around its axis, and beta-tilt - rotation around axis perpendicular to CompuStage rod. beta-tilt rotation enables a circular movement of the sample while alpha-tilt sets the correct position of sample with respect to target and detector. Thin metal lamella of suitable material welded to EasyLift manipulator needle is used as an X-ray target. The final target-sample geometry - position, distance - can be fine-tuned using position control of CompuStage and EasyLift and in-situ monitored by SEM. Both sample and target can also be easily prepared in-situ. Radiographs are recorded by Timepix detector with inherent noise-free operation and energy filtration. For the 3D reconstruction standard microCT reconstruction algorithm is used with the procedure adjusted for the format and quality of nanoCT images. The experiments were carried out on Helios G4 FX DualBeam using titanium and tungsten targets and various semiconductor samples. The ultimate resolution of the proposed method in orders of tens of nanometers was achieved both by the possibility of close target to sample positioning and of adjustment of primary beam energy down to low energies reducing the interaction volume in the target. Since the lower energy radiation is well suited for life-science, the method was also tested on several bio-samples using silver target. The silver target, thanks to its massive low energy L-alpha line, allowed to distinguish subtle structures in the resin embedded stained mouse brain and also to observe and reconstruct canaliculi in the mouse bone (earlier reported by Dierolf et al. 2010, Nature 467 436).
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Overview on established and novel FIB based miniaturized mechanical testing using in-situ SEM
    Kiener, Daniel
    Motz, Christian
    Dehm, Gerhard
    Pippan, Reinhard
    [J]. INTERNATIONAL JOURNAL OF MATERIALS RESEARCH, 2009, 100 (08) : 1074 - 1087
  • [2] Nanomanipulation and Electrical Behaviour of a Single Gold Nanowire Using In-situ SEM-FIB-Nanomanipulators
    Peng, Y.
    Luxmoore, I.
    Forster, M. D.
    Cullis, A. G.
    Inkson, B. J.
    [J]. EMAG: ELECTRON MICROSCOPY AND ANALYSIS GROUP CONFERENCE 2007, 2008, 126
  • [3] Detailed in-situ characterization of uraninite from North American deposits using a combined FIB-SEM approach
    Lewis, Stefanie
    Simonetti, Antonio
    Corcoran, Loretta
    Spano, Tyler
    Burns, Peter
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 255
  • [4] A novel surface patterning using FIB and in-situ etching
    Asaoka, Y
    Arai, T
    Sano, N
    Kaneko, T
    [J]. COMPOUND SEMICONDUCTORS 2001, 2002, (170): : 331 - 334
  • [5] In situ lift-out using a FIB-SEM system
    Langford, RM
    Clinton, C
    [J]. MICRON, 2004, 35 (07) : 607 - 611
  • [6] In-situ nanoindentation in the SEM
    Nowak, Julia D.
    Rzepiejewska-Malyska, Karolina A.
    Major, Ryan C.
    Warren, Oden L.
    Michler, Johann
    [J]. MATERIALS TODAY, 2010, 12 : 44 - 45
  • [7] In situ cryo-FIB/SEM Specimen Preparation Using the Waffle Method
    Klykov, Oleg
    Bobe, Daija
    Paraan, Mohammadreza
    Johnston, Jake D.
    Potter, Clinton S.
    Carragher, Bridget
    Kopylov, Mykhailo
    Noble, Alex J.
    [J]. BIO-PROTOCOL, 2022, 12 (21):
  • [8] In situ FIB-SEM characterization and manipulation methods
    Antoniou, Nicholas
    Rykaczewski, Konrad
    Uchic, Michael D.
    [J]. MRS BULLETIN, 2014, 39 (04) : 347 - 352
  • [9] In situ FIB-SEM characterization and manipulation methods
    Nicholas Antoniou
    Konrad Rykaczewski
    Michael D. Uchic
    [J]. MRS Bulletin, 2014, 39 : 347 - 352
  • [10] Protruding microgripper with force amplification and parallel jaw motion for in-situ sample manipulation in SEM and FIB-machines
    Krijnen, G
    Haanstra, R
    Potters, E
    Berenschot, JW
    von Harrach, S
    Elwenspoek, M
    [J]. BOSTON TRANSDUCERS'03: DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2003, : 268 - 271