Validation of a point-of-care glucometer for use in dairy cows

被引:32
|
作者
Wittrock, J. A. M. [1 ]
Duffield, T. F. [1 ]
LeBlanc, S. J. [1 ]
机构
[1] Univ Guelph, Ontario Vet Coll, Dept Populat Med, Guelph, ON N1G 2W1, Canada
关键词
glucometer; insulin resistance; glucose tolerance test; peripartum; INSULIN-RESISTANCE; GLUCOSE; KETOSIS;
D O I
10.3168/jds.2012-6533
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
The purpose of this study was to evaluate the diagnostic performance of a hand-held electronic glucometer (Precision Xtra; Abbott Diabetes Care Inc., Mississauga, ON, Canada) for cow-side use in dairy cattle. This device has been validated for measuring blood concentrations of beta-hydroxybutyrate in dairy cows. This study was designed to assess the accuracy of whole-blood glucose measurements from the glucose meter relative to a reference chemical analyzer in a diagnostic laboratory. Duplicate samples were taken from the same cows at the same time, into blood tubes with either the glycolysis-inhibiting preservative sodium fluoride (NaF) or without preservative. Glucometer readings were taken on whole blood with no preservative, and laboratory measurements were conducted on serum preserved with NaF. Blood samples were collected from cows between 3 wk before and 5 wk after calving, including during a glucose tolerance test conducted 1 wk before expected calving. Passing-Bablok and Bland-Altman data analyses were used to evaluate the performance of the glucometer relative to the laboratory results. A strong correlation was observed in 709 samples from 81 cows between the hand-held meter and serum from samples preserved with NaF (R-2 = 0.95). Overall, 96% of measurements with the glucometer fell within the 95% confidence limits of analysis in the laboratory, although at higher-than-physiologic glucose concentrations (>5.2 mmol/L) the glucometer tended to overestimate. The hand-held glucometer appears suitable for rapid measurement of glucose under field conditions in dairy cattle.
引用
收藏
页码:4514 / 4518
页数:5
相关论文
共 50 条
  • [1] Validation of a point-of-care handheld blood total calcium analyzer in postpartum dairy cows
    Serrenho, Rita Couto
    Bruinje, Tony C.
    Morrison, Emma I.
    Renaud, David L.
    Devries, Trevor J.
    Duffield, Todd F.
    Leblanc, Stephen J.
    JDS COMMUNICATIONS, 2021, 2 (01): : 41 - 45
  • [2] Rapid detection of bacterial meningitis using a point-of-care glucometer
    Rousseau, Geoffroy
    Asmolov, Romain
    Grammatico-Guillon, Leslie
    Auvet, Adrien
    Laribi, Said
    Garot, Denis
    Jouan, Youenn
    Dequin, Pierre-Francois
    Guillon, Antoine
    EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2019, 26 (01) : 41 - 46
  • [3] Reducing Treatment Errors Through Point-of-Care Glucometer Configuration
    Estock, Jamie L.
    Pham, Ivan-Thibault
    Curinga, Holly K.
    Sprague, Benjamin J.
    Boudreaux-Kelly, Monique Y.
    Acevedo, Jeanette
    Jacobs, Katrina
    JOINT COMMISSION JOURNAL ON QUALITY AND PATIENT SAFETY, 2018, 44 (11): : 683 - 694
  • [4] The evaluation of analytical performance of the precision G™ point-of-care glucometer
    Solnica, B
    Naskalski, JW
    Sieradzki, J
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2001, 39 (12) : 1283 - 1286
  • [5] Evaluating the utility of a point-of-care glucometer for the diagnosis of gestational diabetes
    Adam, Sumaiya
    Rheeder, Paul
    INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2018, 141 (01) : 91 - 96
  • [6] POINT-OF-CARE GLUCOSE TESTING - COST SAVINGS AND EASE OF USE WITH THE AMES-GLUCOMETER-ELITE
    INNANEN, VT
    CAMPOS, FB
    CLINICAL CHEMISTRY, 1995, 41 (10) : 1537 - 1538
  • [7] Point-of-Care Glucometer Assessment of Pleural Fluid Glucose: A Promising Approach
    Marchi, G.
    Gabbrielli, L.
    Gherardi, M.
    Morganti, R.
    Pistelli, F.
    Carrozzi, L.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2024, 209
  • [8] Precision and accuracy of a point-of-care glucometer in horses and the effects of sample type
    Rendle, D. I.
    Armstrong, S. K.
    Heller, J.
    Hughes, K. J.
    VETERINARY JOURNAL, 2019, 252
  • [9] Evaluation of a point-of-care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows
    Zakian, A.
    Tehrani-Sharif, M.
    Mokhber-Dezfouli, M. R.
    Nouri, M.
    Constable, P. D.
    AUSTRALIAN VETERINARY JOURNAL, 2017, 95 (04) : 123 - 128
  • [10] Comparison of a continuous indwelling glucometer with a point-of-care device in healthy adult horses
    Cunneen, Alexandra
    Wood, Kelly A.
    Mathison, Kylie
    Herndon, Aaron M.
    Bertin, Francois R.
    VETERINARY RECORD, 2020, 187 (03)