Evaluation of a point-of-care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows

被引:3
|
作者
Zakian, A. [1 ]
Tehrani-Sharif, M. [2 ]
Mokhber-Dezfouli, M. R. [3 ]
Nouri, M. [1 ]
Constable, P. D. [4 ]
机构
[1] Shahid Chamran Univ, Fac Vet Med, Dept Clin Sci, Ahvaz, Iran
[2] Islamic Azad Univ, Fac Vet Med, Dept Clin Sci, Garmsar Branch, Tehran, Iran
[3] Univ Tehran, Fac Vet Med, Dept Internal Med, Tehran, Iran
[4] Univ Illinois, Coll Vet Med, Dept Vet Clin Med, Urbana, IL 61801 USA
关键词
dairy cattle; hyperketonaemia; hypoglycaemia; ketosis; method comparison; HAND-HELD METER; BETA-HYDROXYBUTYRATE; BLOOD; MILK; GLUCOSE; PLASMA; VALIDATION; ACCURACY; TESTS;
D O I
10.1111/avj.12568
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objectives To evaluate and validate a hand-held electrochemical meter (Precision Xtra (R)) as a screening test for subclinical ketosis and hypoglycaemia in lactating dairy cattle. Design Method comparison study using a convenience sample. Procedure Blood samples were collected into plain tubes from the coccygeal vessels of 181 Holstein cows at 2-4 weeks of lactation during summer in Iran. Blood beta-hydroxybutyrate concentration (BHB) and glucose concentration were immediately measured by the electrochemical meter after applying 20 mu L of blood to the reagent strip. Passing-Bablok regression and Bland-Altman plots were used to determine the accuracy of the meter against laboratory reference methods (BHB dehydrogenase and glucose oxidase). Results Serum BHB ranged from 0.1 to 7.3 mmol/L and serum glucose ranged from 0.9 to 5.1 mmol/L. Passing-Bablok regression analysis indicated that the electrochemical meter and reference methods were linearly related for BHB and glucose, with a slope estimate that was not significantly different from 1.00. Clinically minor, but statistically significant, differences were present for the intercept value for Passing-Bablok regression analysis for BHB and glucose, and bias estimates in the Bland-Altman plots for BHB and glucose. Conclusion The electrochemical meter provided a clinically useful method to detect subclinical ketosis and hypoglycaemia in lactating dairy cows. Compared with other method validation studies using the meter, we attributed the improved performance of the electrochemical meter to application of a fixed volume of blood (20 mu L) to the reagent strip, use of the meter in hot ambient conditions and use of glucose oxidase as the reference method for glucose analysis.
引用
收藏
页码:123 / 128
页数:6
相关论文
共 50 条
  • [1] Evaluation of an electronic cowside test to detect subclinical ketosis in dairy cows
    Iwersen, M.
    Falkenberg, U.
    Voigtsberger, R.
    Forderung, D.
    Heuwieser, W.
    [J]. JOURNAL OF DAIRY SCIENCE, 2009, 92 (06) : 2618 - 2624
  • [2] Efficacy of monensin for the prevention of subclinical ketosis in lactating dairy cows
    Duffield, TF
    Sandals, D
    Leslie, KE
    Lissemore, K
    McBride, BW
    Lumsden, JH
    Dick, P
    Bagg, R
    [J]. JOURNAL OF DAIRY SCIENCE, 1998, 81 (11) : 2866 - 2873
  • [3] Evaluation of five cowside tests for use with milk to detect subclinical ketosis in dairy cows
    Geishauser, T
    Leslie, K
    Kelton, D
    Duffield, T
    [J]. JOURNAL OF DAIRY SCIENCE, 1998, 81 (02) : 438 - 443
  • [4] Use of a hand-held meter for detecting subclinical ketosis in dairy cows
    Voyvoda, Huseyin
    Erdogan, Hasan
    [J]. RESEARCH IN VETERINARY SCIENCE, 2010, 89 (03) : 344 - 351
  • [5] Evaluation of an Electrochemical Point-of-Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle
    Megahed, A. A.
    Hiew, M. W. H.
    Townsend, J. R.
    Messick, J. B.
    Constable, P. D.
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2015, 29 (06) : 1718 - 1727
  • [6] Evaluation of eight ketontests in milk for detection of subclinical ketosis in dairy cows
    Geishauser, T
    Leslie, K
    Tenhag, J
    [J]. PRAKTISCHE TIERARZT, 1999, 80 (12): : 1085 - +
  • [7] An evaluation of β-hydroxybutyrate in milk and blood for prediction of subclinical ketosis in dairy cows
    Samiei, A.
    Liang, J. B.
    Ghorbani, G. R.
    Hirooka, H.
    Yaakub, H.
    Tabatabaei, M.
    [J]. POLISH JOURNAL OF VETERINARY SCIENCES, 2010, 13 (02): : 349 - 356
  • [8] Evaluation of the Change of Serum Copper and Zinc Concentrations of Dairy Cows with Subclinical Ketosis
    Zhang, Zhigang
    Liu, Guowen
    Li, Xiaobing
    Gao, Li
    Guo, Changming
    Wang, Hongbin
    Wang, Zhe
    [J]. BIOLOGICAL TRACE ELEMENT RESEARCH, 2010, 138 (1-3) : 8 - 12
  • [9] EVALUATION OF 2 COWSIDE TESTS FOR THE DETECTION OF SUBCLINICAL KETOSIS IN DAIRY-COWS
    NIELEN, M
    AARTS, MGA
    JONKERS, AGM
    WENSING, T
    SCHUKKEN, YH
    [J]. CANADIAN VETERINARY JOURNAL-REVUE VETERINAIRE CANADIENNE, 1994, 35 (04): : 229 - 232
  • [10] Evaluation of the Change of Serum Copper and Zinc Concentrations of Dairy Cows with Subclinical Ketosis
    Zhigang Zhang
    Guowen Liu
    Xiaobing Li
    Li Gao
    Changming Guo
    Hongbin Wang
    Zhe Wang
    [J]. Biological Trace Element Research, 2010, 138 : 8 - 12