Integrity of the stem-cement interface in THA: Effects of stem surface finish and cement porosity

被引:6
|
作者
Iesaka, Kazuho [1 ]
Jaffe, William L. [2 ]
Kummer, Frederick J. [2 ]
机构
[1] Manaka Mem Clin, Gunma 3700036, Japan
[2] New York Univ Hosp, Joint Dis Orthopaed Inst, Musculoskeletal Res Ctr, New York, NY 10003 USA
关键词
total hip replacement; bone cement; surface finish; cement porosity; stem preheating;
D O I
10.1002/jbm.b.31071
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
No consensus exists for the optimal surface finish on cemented total hip prosthesis stems. The purpose of this study was to determine the effects of stem finish and interfacial cement porosity on the integrity of the stem-cement interface. Simulated stems made of Co-Cr, having polished or matte surfaces, at room temperature or heated to 37 degrees C, were cemented into Sawbones simulated femurs. Push out testing of the stem-cement interface was performed immediately after cement polymerization and after two aging periods in 37 degrees C saline or 37 degrees C air, and the extent of interfacial porosity at the stem-cement interface was determined. Polished stems exhibited an average 60% greater interfacial strength than that of matte stems initially and up to 240% after aging treatments. Cement porosity at the stem-cement interface and incomplete cement interdigitation into surface asperities on matte stems likely allowed saline penetration into the stem-cement interface during wet aging, resulting in a rapid decrease of shear strength. Stem preheating to 37 degrees C virtually eliminated interfacial pores and resulted in greater shear strengths regardless of surface finish. Polished stem surfaces with stem preheating provided the best interfacial shear strength and sealing ability against saline penetration into the stem-cement interface and could result in increased longevity of stem fixation. (c) 2008 Wiley Periodicals, Inc.
引用
收藏
页码:77 / 82
页数:6
相关论文
共 50 条
  • [21] Reproduction of fretting wear at the stem-cement interface in total hip replacement
    Brown, L.
    Zhang, H.
    Blunt, L.
    Barrans, S.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H8) : 963 - 971
  • [22] Investigation of relative micromotion at the stem-cement interface in total hip replacement
    Zhang, H-Y
    Brown, L.
    Barrans, S.
    Blunt, L.
    Jiang, X-Q
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H8) : 955 - 964
  • [23] Fatigue fracture of the stem-cement interface with a clamped cantilever beam test
    Heuer, DA
    Mann, KA
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2000, 122 (06): : 647 - 651
  • [24] Fatigue of bone cement with simulated stem interface porosity
    Tepic, S
    Soltész, U
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1998, 9 (12) : 707 - 709
  • [25] Fatigue of bone cement with simulated stem interface porosity
    S. Tepic
    U. SoltÉsz
    Journal of Materials Science: Materials in Medicine, 1998, 9 : 707 - 709
  • [26] Microdamages: an insight to mechanical performance of stem-cement
    Qi, Gang
    Zhang, Bin
    Mann, Kenneth A.
    PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON MECHANICAL ENGINEERING AND MECHANICS 2007, VOLS 1 AND 2, 2007, : 1182 - 1185
  • [27] Effect of antibiotic loading on the shear strength at the stem-cement interface (Shear strength of antibiotic loaded cement)
    Kilicoglu, Onder
    Koyuncu, L. Ozgur
    Ozden, V. Emre
    Bozdag, Ergun
    Sunbuloglu, Emin
    Yazicioglu, Onder
    INTERNATIONAL ORTHOPAEDICS, 2008, 32 (04) : 437 - 441
  • [28] Influence of proximal stem geometry and stem-cement interface characteristics on bone and cement stresses in femoral hip arthroplasty: finite element analysis
    Massin, P
    Astoin, E
    Lavaste, F
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2003, 89 (02): : 134 - 143
  • [29] Finite element model of the failure of stem-cement interface in total hip arthroplasty
    Amiri, Riad
    Bouiadjra, Bel Abbes Bachir
    Amiri, Ahmed
    UPB Scientific Bulletin, Series D: Mechanical Engineering, 2019, 81 (03): : 201 - 212
  • [30] Investigation of Protein Adsorption Mechanism and Biotribological Properties at Simulated Stem-Cement Interface
    Zhang, Hongyu
    Zhang, Shaohua
    Luo, Jianbin
    Liu, Yuhong
    Qian, Shanhua
    Liang, Fanghui
    Huang, Yongling
    JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2013, 135 (03):