Investigation of Protein Adsorption Mechanism and Biotribological Properties at Simulated Stem-Cement Interface

被引:39
|
作者
Zhang, Hongyu [1 ]
Zhang, Shaohua [1 ]
Luo, Jianbin [1 ]
Liu, Yuhong [1 ]
Qian, Shanhua [1 ]
Liang, Fanghui [2 ]
Huang, Yongling [2 ]
机构
[1] Tsinghua Univ, State Key Lab Tribol, Dept Precis Instruments & Mechanol, Beijing 100084, Peoples R China
[2] Beijing Inst Aeronaut Mat, Jinghang Biomed Engn Div, Beijing 100095, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
stem-cement interface; biotribology; surface characterization; protein adsorption; STATIC SHEAR-STRENGTH; FEMORAL STEM; FRETTING WEAR; LUBRICATION CONDITIONS; SURFACE-ROUGHNESS; HIP; FRICTION; BEHAVIOR; COMPONENTS; INITIATION;
D O I
10.1115/1.4023802
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Debonding of the stem-cement interface occurs inevitably for almost all stem designs under physiological loading, and the wear debris generated at this interface is showing an increasing significance in contributing to the mechanical failure of cemented total hip replacements. However, the influence of protein adsorption onto the femoral stem and the bone cement surfaces has not been well taken into consideration across previous in vitro wear simulations. In the present study, the protein adsorption mechanism and biotribological properties at the stem-cement interface were investigated through a series of frictional tests using bone cements and femoral stems with two kinds of surface finishes, lubricated by calf serum at body temperature. The friction coefficient was dependent on the surface finish of the samples, with an initial much lower value obtained for the polished contacting pairs followed by a sudden increase in the friction coefficient with regard to the tests performed at higher frequencies. The friction coefficient did not change much during the tests for the glass-bead blasted contacting pairs. In addition, proteins from the calf serum were found to adsorb onto both the femoral stem and the bone cement surfaces, and the thickness of the physically adsorbed proteins on the polished metallic samples was more than 10 mu m, which was measured using an optical interferometer and validated through a vertical scanning methodology based on Raman spectroscopy. An initial protein adsorption mechanism and biotribological properties at the stem-cement interface were examined in this study, and it suggested that wear at the stem-cement interface may be postponed or reduced by tailoring physicochemical properties of the femoral components to promote protein adsorption.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] The influence of protein concentration on the biotribological properties of the stem-cement interface
    Zhang, Hong-Yu
    Zhou, Ming
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 173 - 179
  • [2] Biotribological properties at the stem-cement interface lubricated with different media
    Zhang, H. Y.
    Luo, J. B.
    Zhou, M.
    Zhang, Y.
    Huang, Y. L.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2013, 20 : 209 - 216
  • [3] MECHANICAL CHARACTERISTICS OF THE STEM-CEMENT INTERFACE
    MANN, KA
    BARTEL, DL
    WRIGHT, TM
    INGRAFFEA, AR
    JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (06) : 798 - 808
  • [4] Investigation of relative micromotion at the stem-cement interface in total hip replacement
    Zhang, H-Y
    Brown, L.
    Barrans, S.
    Blunt, L.
    Jiang, X-Q
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H8) : 955 - 964
  • [5] Results of in vitro studies about the mechanism of wear in the stem-cement interface of THR
    Wirz, D
    Zurfluh, B
    Göpfert, B
    Li, F
    Frick, W
    Morscher, EW
    STAINLESS STEELS FOR MEDICAL AND SURGICAL APPLICATIONS, 2003, 1438 : 222 - 234
  • [6] Effects of preheating of hip prostheses on the stem-cement interface
    Thienpont, E
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (11): : 2247 - 2248
  • [7] Experiment Investigation on Fretting Wear and Wear Debris Performance for the Stem-Cement Interface
    Zhang, Lanfeng
    Ge, Shirong
    Liu, Hongtao
    JOURNAL OF NANOMATERIALS, 2015, 2015
  • [8] Effects of the initial temperature of acrylic bone cement liquid monomer on the properties of the stem-cement interface and cement polymerization
    Iesaka, K
    Jaffe, WL
    Kummer, FJ
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2004, 68B (02) : 186 - 190
  • [9] Residual stresses at the stem-cement interface of an idealized cemented hip stem
    Nuño, N
    Avanzolini, G
    JOURNAL OF BIOMECHANICS, 2002, 35 (06) : 849 - 852
  • [10] A better understanding of the surface topography at the stem-cement interface
    Zhang, H.
    Brown, L.
    Blunt, L.
    PROCEEDINGS OF THE ASME/STLE INTERNATIONAL JOINT TRIBOLOGY CONFERENCE, PTS A AND B, 2008, : 513 - 515