Integrity of the stem-cement interface in THA: Effects of stem surface finish and cement porosity

被引:6
|
作者
Iesaka, Kazuho [1 ]
Jaffe, William L. [2 ]
Kummer, Frederick J. [2 ]
机构
[1] Manaka Mem Clin, Gunma 3700036, Japan
[2] New York Univ Hosp, Joint Dis Orthopaed Inst, Musculoskeletal Res Ctr, New York, NY 10003 USA
关键词
total hip replacement; bone cement; surface finish; cement porosity; stem preheating;
D O I
10.1002/jbm.b.31071
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
No consensus exists for the optimal surface finish on cemented total hip prosthesis stems. The purpose of this study was to determine the effects of stem finish and interfacial cement porosity on the integrity of the stem-cement interface. Simulated stems made of Co-Cr, having polished or matte surfaces, at room temperature or heated to 37 degrees C, were cemented into Sawbones simulated femurs. Push out testing of the stem-cement interface was performed immediately after cement polymerization and after two aging periods in 37 degrees C saline or 37 degrees C air, and the extent of interfacial porosity at the stem-cement interface was determined. Polished stems exhibited an average 60% greater interfacial strength than that of matte stems initially and up to 240% after aging treatments. Cement porosity at the stem-cement interface and incomplete cement interdigitation into surface asperities on matte stems likely allowed saline penetration into the stem-cement interface during wet aging, resulting in a rapid decrease of shear strength. Stem preheating to 37 degrees C virtually eliminated interfacial pores and resulted in greater shear strengths regardless of surface finish. Polished stem surfaces with stem preheating provided the best interfacial shear strength and sealing ability against saline penetration into the stem-cement interface and could result in increased longevity of stem fixation. (c) 2008 Wiley Periodicals, Inc.
引用
收藏
页码:77 / 82
页数:6
相关论文
共 50 条
  • [41] The influence of the stem-cement interface in total hip replacement - A comparison of experimental and finite element approaches
    Wheeler, JPG
    Miles, AW
    Clift, SE
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (02) : 181 - 186
  • [42] Effects of temperature on the evolution of stresses at the stem cement interface
    Kaci, Djafar Ait
    Moulgada, Abdelmadjid
    Achache, Habib
    Bounoua, Noureddine
    ADVANCES IN COMPUTATIONAL DESIGN, 2019, 4 (03): : 239 - 250
  • [43] Effect of metal surface topography on mechanical bonding at simulated total hip stem-cement interfaces
    Chen, CQL
    Scott, W
    Barker, TM
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1999, 48 (04): : 440 - 446
  • [44] The significance of stem-cement loosening of grit-blasted femoral components
    Schmalzried, TP
    Zahiri, CA
    Woolson, ST
    ORTHOPEDICS, 2000, 23 (11) : 1157 - 1164
  • [45] Analysis of the debonding of the stem-cement interface in intramedullary fixation using a non-linear fracture mechanics approach
    Pérez, MA
    García, JM
    Doblaré, GM
    ENGINEERING FRACTURE MECHANICS, 2005, 72 (08) : 1125 - 1147
  • [46] Effect of bone porosity on the mechanical integrity of the bone-cement interface
    Graham, J
    Ries, M
    Pruitt, L
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (10): : 1901 - 1908
  • [47] The effect of surface finish and interstitial fluid on the cement-in-cement interface in revision surgery of the hip
    Rudol, G.
    Wilcox, R.
    Jin, Z.
    Tsiridis, E.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (02): : 188 - 193
  • [48] Mechanical effects of stem cement interface characteristics in total hip replacement
    Verdonschot, N
    Huiskes, R
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (329) : 326 - 336
  • [49] Early cement damage around a femoral stem is concentrated at the cement/bone interface
    Race, A
    Miller, MA
    Ayers, DC
    Mann, KA
    JOURNAL OF BIOMECHANICS, 2003, 36 (04) : 489 - 496
  • [50] Stem-cement porosity may explain early loosening of cemented femoral hip components: Experimental-computational in vitro study
    Mann, Kenneth A.
    Damron, Leatha A.
    Miller, Mark A.
    Race, Amos
    Clarke, Michael T.
    Cleary, Richard J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2007, 25 (03) : 340 - 350