Risk of electromagnetic interference induced by dental equipment on cardiac implantable electrical devices

被引:5
|
作者
Miranda-Rius, Jaume [1 ]
Lahor-Soler, Eduard [1 ]
Brunet-Llobet, Lluis [2 ]
Sabate de la Cruz, Xavier [3 ]
机构
[1] Univ Barcelona, Fac Med & Ciencies Salut, Dept Odontostomatol, Feixa Llarga S-N, Barcelona 08907, Spain
[2] Univ Barcelona, Hosp St Joan de Deu, Serv Odontol, Barcelona, Spain
[3] Univ Barcelona, Fac Med & Ciencies Salut, Bellvitge Hosp, Serv Cardiol,Unitat Aritmies,Dept Ciencies Clin, Barcelona, Spain
关键词
dental equipment; electromagnetic interference; implantable cardioverter defibrillator; pacemaker; risk; CARDIOVERTER-DEFIBRILLATORS; PULP TESTERS; PART II; PACEMAKERS; CAPACITY;
D O I
10.1111/eos.12313
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Patients with cardiac implantable electrical devices should take special precautions when exposed to electromagnetic fields. Proximity to equipment used in clinical dentistry may cause interference. This study evaluated in vitro the risks associated with different types/makes of cardiac devices and types of dental equipment. Six electronic dental tools were tested on three implantable cardioverter defibrillators and three pacemakers made by different manufacturers. Overall, the risk of interference with the pacemakers was 37% lower than with the implantable cardioverter defibrillators. Regarding the types/makes of cardiac devices analysed, that from Boston Scientific had a five-fold greater risk of interference than did that from Biotronik [prevalence ratio (PR) = 5.58]; there was no difference between that from Biotronik and that from Medtronic. Among the dental equipment, the electric pulp tester had the greatest risk of inducing interference and therefore this device was used as the benchmark. The electronic apex locator (PR = 0.29), Periotest M (PR = 0.47), and the ultrasonic dental scaler (PR = 0.59) were less likely to induce interference than the electric pulp tester. The risk was lowest with the electronic apex locator. Pacemakers presented a lower risk of light to moderate interference (PR = 0.63). However, the risk of severe electromagnetic interference was 3.5 times higher with pacemakers than with implantable cardioverter defibrillators (PR = 3.47).
引用
收藏
页码:559 / 565
页数:7
相关论文
共 50 条
  • [41] Effects of electromagnetic interference on implanted cardiac devices
    Blitzer, ML
    Schoenfeld, MH
    CARDIAC ARRHYTHMIAS 2003, 2004, : 911 - 918
  • [42] Freedom from electromagnetic interference between cardiac implantable electronic devices and the FMwand ferromagnetic surgical system
    Weiss, J. Peter
    Manwaring, Preston
    Bateman, Orlon
    JOURNAL OF CLINICAL ANESTHESIA, 2013, 25 (08) : 681 - 684
  • [43] Does ultrasonic dental equipment affect cardiovascular implantable electronic devices?
    Stoopler, Eric T.
    Sia, Ying Wai
    Kuperstein, Arthur S.
    JOURNAL OF THE CANADIAN DENTAL ASSOCIATION, 2011, 77
  • [44] Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven
    Hirose, M
    Hida, M
    Sato, E
    Kokubo, K
    Nie, M
    Kobayashi, H
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (06): : 540 - 548
  • [45] Remote interrogation and reprogramming for cardiac implantable electrical devices
    Clarke, John-Ross D.
    Kramer, Daniel B.
    HEART RHYTHM, 2023, 20 (04) : 552 - 553
  • [46] Reproducibility of CMR in Patients With Cardiac Implantable Electrical Devices
    Ostovaneh, Mohammad R.
    Ward, Caroline
    Ambale-Venkatesh, Bharath
    Chamera, Elzbieta
    Kato, Yoko
    Bolli, Roberto
    Mitrani, Raul
    Perin, Emerson C.
    Henry, Timothy D.
    Hare, Joshua M.
    Moye, Lem
    Nazarian, Saman
    Lima, Joao A. C.
    JACC-CARDIOVASCULAR IMAGING, 2022, 15 (05) : 952 - 954
  • [47] SMALL BOWEL CAPSULE ENDOSCOPY AND CARDIAC IMPLANTABLE ELECTRONIC DEVICES: WHAT IS THE REAL RISK OF INTERFERENCE?
    Condon, Ashwinee S.
    Lenhart, Adrienne
    Grewal, Jasleen
    Kumar, Puneet
    Marya, Neil
    Kim, Stephen
    GASTROINTESTINAL ENDOSCOPY, 2020, 91 (06) : AB610 - AB610
  • [48] Selective interference with pacemaker activity by electrical dental devices
    Miller, CS
    Leonelli, FM
    Latham, E
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1998, 85 (01): : 33 - 36
  • [49] Potential for Personal Digital Assistant interference with implantable cardiac devices
    Tri, JL
    Trusty, JM
    Hayes, DL
    MAYO CLINIC PROCEEDINGS, 2004, 79 (12) : 1527 - 1530
  • [50] Electromagnetic interference of an implantable loop recorder by commonly encountered electronic devices
    de Cock, CC
    Spruijt, HJ
    van Campen, LMC
    Plu, AW
    Visser, CA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (10): : 1516 - 1518