Risk of electromagnetic interference induced by dental equipment on cardiac implantable electrical devices

被引:5
|
作者
Miranda-Rius, Jaume [1 ]
Lahor-Soler, Eduard [1 ]
Brunet-Llobet, Lluis [2 ]
Sabate de la Cruz, Xavier [3 ]
机构
[1] Univ Barcelona, Fac Med & Ciencies Salut, Dept Odontostomatol, Feixa Llarga S-N, Barcelona 08907, Spain
[2] Univ Barcelona, Hosp St Joan de Deu, Serv Odontol, Barcelona, Spain
[3] Univ Barcelona, Fac Med & Ciencies Salut, Bellvitge Hosp, Serv Cardiol,Unitat Aritmies,Dept Ciencies Clin, Barcelona, Spain
关键词
dental equipment; electromagnetic interference; implantable cardioverter defibrillator; pacemaker; risk; CARDIOVERTER-DEFIBRILLATORS; PULP TESTERS; PART II; PACEMAKERS; CAPACITY;
D O I
10.1111/eos.12313
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Patients with cardiac implantable electrical devices should take special precautions when exposed to electromagnetic fields. Proximity to equipment used in clinical dentistry may cause interference. This study evaluated in vitro the risks associated with different types/makes of cardiac devices and types of dental equipment. Six electronic dental tools were tested on three implantable cardioverter defibrillators and three pacemakers made by different manufacturers. Overall, the risk of interference with the pacemakers was 37% lower than with the implantable cardioverter defibrillators. Regarding the types/makes of cardiac devices analysed, that from Boston Scientific had a five-fold greater risk of interference than did that from Biotronik [prevalence ratio (PR) = 5.58]; there was no difference between that from Biotronik and that from Medtronic. Among the dental equipment, the electric pulp tester had the greatest risk of inducing interference and therefore this device was used as the benchmark. The electronic apex locator (PR = 0.29), Periotest M (PR = 0.47), and the ultrasonic dental scaler (PR = 0.59) were less likely to induce interference than the electric pulp tester. The risk was lowest with the electronic apex locator. Pacemakers presented a lower risk of light to moderate interference (PR = 0.63). However, the risk of severe electromagnetic interference was 3.5 times higher with pacemakers than with implantable cardioverter defibrillators (PR = 3.47).
引用
收藏
页码:559 / 565
页数:7
相关论文
共 50 条
  • [11] Electromagnetic interference of implantable cardiac devices from a shoulder massage machine
    Yoshida, Saeko
    Fujiwara, Kousaku
    Kohira, Satoshi
    Hirose, Minoru
    JOURNAL OF ARTIFICIAL ORGANS, 2014, 17 (03) : 243 - 249
  • [12] Electromagnetic interference of implantable cardiac devices from a shoulder massage machine
    Saeko Yoshida
    Kousaku Fujiwara
    Satoshi Kohira
    Minoru Hirose
    Journal of Artificial Organs, 2014, 17 : 243 - 249
  • [13] Electromagnetic Interference in Cardiac Implantable Electronic Devices Is the Use of Smartphones Safe?
    Lennerz, Carsten
    Pavaci, Herribert
    Grebmer, Christian
    Semmler, Verena
    Bourier, Felix
    Haller, Bernhard
    Reents, Tilko
    Hessling, Gabriele
    Deisenhofer, Isabel
    Kolb, Christof
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (01) : 108 - 110
  • [14] Modern Security Screening and Electromagnetic Interference With Cardiac Implantable Electronic Devices
    Lennerz, Carsten
    O'Connor, Matthew J.
    Blazek, Patrick
    Weigand, Severin
    Grebmer, Christian
    Reif, Sebastian
    Friedrich, Lena
    Reents, Tilko
    Bourier, Felix
    Schunkert, Heribert
    Deisenhofer, Isabel
    Kolb, Christof
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (10) : 1238 - 1239
  • [15] Risk of cardioversion for patients with cardiac implantable electrical devices
    Yoneda, Zachary T.
    Crossley, George H.
    HEART RHYTHM, 2023, 20 (09) : 1236 - 1237
  • [16] Electromagnetic interference in current implantable devices
    Irnich, W
    CARDIAC ARRHYTHMIAS, PACING & ELECTROPHYSIOLOGY: THE EXPERT VIEW, 1998, 201 : 427 - 436
  • [17] Implantable rhythm devices and electromagnetic interference
    Lamaison, D.
    ARCHIVES DES MALADIES PROFESSIONNELLES ET DE L ENVIRONNEMENT, 2012, 73 (03) : 332 - 333
  • [18] Cardiac implantable electronic devices: Electromagnetic interference from electrocauterization, lithotripsy and physiotherapy
    Czermak T.
    Fichtner S.
    Herzschrittmachertherapie + Elektrophysiologie, 2019, 30 (2) : 168 - 176
  • [19] Electromagnetic interference: Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter–defibrillator
    Nowak B.
    Przibille O.
    Napp A.
    Herzschrittmachertherapie + Elektrophysiologie, 2022, 33 (3) : 297 - 304
  • [20] Hyfrecation and Interference With Implantable Cardiac Devices
    Amin, Sima D.
    Homan, Katie B.
    Assar, Manish
    Lee, Mary
    Housewright, Chad D.
    DERMATOLOGIC SURGERY, 2020, 46 (05) : 612 - 615