Risk of electromagnetic interference induced by dental equipment on cardiac implantable electrical devices

被引:5
|
作者
Miranda-Rius, Jaume [1 ]
Lahor-Soler, Eduard [1 ]
Brunet-Llobet, Lluis [2 ]
Sabate de la Cruz, Xavier [3 ]
机构
[1] Univ Barcelona, Fac Med & Ciencies Salut, Dept Odontostomatol, Feixa Llarga S-N, Barcelona 08907, Spain
[2] Univ Barcelona, Hosp St Joan de Deu, Serv Odontol, Barcelona, Spain
[3] Univ Barcelona, Fac Med & Ciencies Salut, Bellvitge Hosp, Serv Cardiol,Unitat Aritmies,Dept Ciencies Clin, Barcelona, Spain
关键词
dental equipment; electromagnetic interference; implantable cardioverter defibrillator; pacemaker; risk; CARDIOVERTER-DEFIBRILLATORS; PULP TESTERS; PART II; PACEMAKERS; CAPACITY;
D O I
10.1111/eos.12313
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Patients with cardiac implantable electrical devices should take special precautions when exposed to electromagnetic fields. Proximity to equipment used in clinical dentistry may cause interference. This study evaluated in vitro the risks associated with different types/makes of cardiac devices and types of dental equipment. Six electronic dental tools were tested on three implantable cardioverter defibrillators and three pacemakers made by different manufacturers. Overall, the risk of interference with the pacemakers was 37% lower than with the implantable cardioverter defibrillators. Regarding the types/makes of cardiac devices analysed, that from Boston Scientific had a five-fold greater risk of interference than did that from Biotronik [prevalence ratio (PR) = 5.58]; there was no difference between that from Biotronik and that from Medtronic. Among the dental equipment, the electric pulp tester had the greatest risk of inducing interference and therefore this device was used as the benchmark. The electronic apex locator (PR = 0.29), Periotest M (PR = 0.47), and the ultrasonic dental scaler (PR = 0.59) were less likely to induce interference than the electric pulp tester. The risk was lowest with the electronic apex locator. Pacemakers presented a lower risk of light to moderate interference (PR = 0.63). However, the risk of severe electromagnetic interference was 3.5 times higher with pacemakers than with implantable cardioverter defibrillators (PR = 3.47).
引用
收藏
页码:559 / 565
页数:7
相关论文
共 50 条
  • [31] Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review
    Gordon, Jonathan S.
    Maynes, Elizabeth J.
    O'Malley, Thomas J.
    Pavri, Behzad B.
    Tchantchaleishvili, Vakhtang
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 61 (01) : 1 - 10
  • [32] Electric Cars and Electromagnetic Interference With Cardiac Implantable Electronic Devices: A Cross-sectional Evaluation
    Lennerz, Carsten
    O'Connor, Matthew
    Horlbeck, Lorenz
    Michel, Jonathan
    Weigand, Severin
    Grebmer, Christian
    Blazek, Patrick
    Brkic, Amir
    Semmler, Verena
    Haller, Bernhard
    Reents, Tilko
    Hessling, Gabriele
    Deisenhofer, Isabel
    Whittaker, Peter
    Lienkamp, Markus
    Kolb, Christof
    ANNALS OF INTERNAL MEDICINE, 2018, 169 (05) : 350 - +
  • [33] Comparative effectiveness of cardiac implantable electrical devices
    Kramer, Daniel B.
    Hatfield, Laura A.
    Normand, Sharon-Lise T.
    HEART, 2015, 101 (22) : 1773 - 1775
  • [34] Future of implantable electrical cardiac devices DISCUSSION
    Vacheron, Andre
    Lecomte, Dominique
    BULLETIN DE L ACADEMIE NATIONALE DE MEDECINE, 2014, 198 (03): : 487 - 488
  • [35] MRI for Patients with Cardiac Implantable Electrical Devices
    Chow, Grant V.
    Nazarian, Saman
    CARDIOLOGY CLINICS, 2014, 32 (02) : 299 - +
  • [36] Electromagnetic interference with implantable cardiac pacemakers by video capsule
    Dubner, S
    Dubner, Y
    Gallino, S
    Spallone, L
    Zagalsky, D
    Rubio, H
    Zimmerman, J
    Goldin, E
    GASTROINTESTINAL ENDOSCOPY, 2005, 61 (02) : 250 - 254
  • [37] New aspects of implantable electrical cardiac devices
    Israel, Carsten W.
    Kuck, Karl-Heinz
    HERZ, 2021, 46 (06) : 497 - 498
  • [38] Investigation of Hyfrecators and Their In Vitro Interference with Implantable Cardiac Devices
    Weyer, Christopher
    Siegle, Ronald J.
    Girard, Guillaume
    DERMATOLOGIC SURGERY, 2012, 38 (11) : 1843 - 1848
  • [39] Electromagnetic Interference in Cardiac Contractility Modulation Devices
    Napp, Andreas
    Zink, Matthias
    Kraus, Thomas
    Marx, Nikolaus
    Stunder, Dominik
    CIRCULATION, 2017, 136
  • [40] Electromagnetic Interference in Cardiac Implantable Electronic Devices Due to Dynamic Wireless Power Systems for Electric Vehicles
    Campi, Tommaso
    Cruciani, Silvano
    Maradei, Francesca
    Feliziani, Mauro
    ENERGIES, 2023, 16 (09)