Comparison of detector performance in small 6 MV and 6 MV FFF beams using a Versa HD accelerator

被引:12
|
作者
Denia, Paula Monasor [1 ]
Castellet Garcia, Maria del Carmen [1 ]
Manjon Garcia, Carla [1 ]
Quiros Higueras, Juan David [1 ]
Blancas, Noelia de Marco [1 ]
Alandi, Jorge Bonaque [1 ]
Juan Senabre, Xavier Jordi [1 ]
Serra, Agustin Santos [1 ]
Lopez-Tarjuelo, Juan [1 ]
机构
[1] Consorcio Hosp Prov Castellon, Serv Radiofis & Protecc Radiol, Castellon De La Plana, Spain
来源
PLOS ONE | 2019年 / 14卷 / 03期
关键词
UNFLATTENED PHOTON BEAMS; OUTPUT FACTOR MEASUREMENTS; FILTER FREE BEAMS; MONTE-CARLO; DOSIMETRIC PROPERTIES; SMALL FIELDS; CLINICAL-APPLICATION; FORMALISM; DENSITY;
D O I
10.1371/journal.pone.0213253
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
1. Background & purpose Investigate the applicability of a series of detectors in small field dosimetry and the possible differences between their responses to FF and FFF beams. This work extends upon the series of detectors used by other authors to also include metal-oxide-semiconductor field-effect transistors (MOSFETs) detectors and radiochromic film. We also included a later correction of output factors (OFs) recommended by the recently published IAEA's code of practice TRS 483 on dosimetry of small static fields used in external beam radiotherapy. 2. Materials & methods The OFs, profiles, and PDDs of 6 MV and 6 MV FFF beams were measured with 11 different detectors using field sizes between 0.6 x 0.6 cm(2) and 10 x 10 cm(2). 3. Results The OFs of the FFF beams were lower than those of the FF beams for field sizes larger than 3 x 3 cm(2) but higher for field sizes smaller than 3 x 3 cm(2). After applying the IAEA's TRS 483 corrections, the final OFs were compatible with our initial results when considering uncertainties involved. Small-volume detectors are preferable for measuring the penumbra of these small fields where this attribute is higher in the crossline direction than in the inline direction. The R-100 of equivalent-quality FFF beams was higher compared to the corresponding flattened beams. 4. Conclusions We observed no difference for the dose responses between 6 MV and 6 MV FFF beams for any of the detectors. OF results, profiles and PDDs were clearly consistent with the previously published literature regarding the Versa HD linac. Correcting our first OFs, taken as ratio of detector charges, with the IAEA's TRS 483 corrections to obtain the final OFs, did not make the former significantly different.
引用
收藏
页数:19
相关论文
共 50 条
  • [21] Scatter Factors Comparison of 6MV Flattened and 7MV Unflattened Beams
    Ashokkumar, S.
    Sinha, S. N.
    Yadav, Girigesh
    Raman, Kothanda
    Thiyagarajan, Rajesh
    Nambiraj, N. Arunai
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3737 - 3737
  • [22] COMPARISON OF A MONTE CARLO 6MV LINAC MODEL WITH MEASURED 6MV BEAMS FOR PERIPHERAL DOSIMETRY
    Joosten, A.
    Levi, F.
    Baechier, S.
    Mirimanoff, R. O.
    Bochud, F.
    Moeckli, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S483 - S483
  • [23] Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations
    Cranmer-Sargison, Gavin
    Weston, Steve
    Sidhu, Narinder P.
    Thwaites, David I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) : 429 - 435
  • [24] Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator
    Moran Fitch, F.
    Garcia-Garduno, O. A.
    Martinez-Davalos, A.
    [J]. ELEVENTH MEXICAN SYMPOSIUM ON MEDICAL PHYSICS, 2010, 1310 : 110 - +
  • [25] Dosimetric comparison of mDCAT and VMAT techniques according to 6MV-FFF and 10MV-FFF energies in patients with single adrenal metastasis
    Durmus, Ismail Faruk
    Okumus, Ayse
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2022, 30 (05) : 1023 - 1032
  • [26] Monte Carlo Simulation of Surface Dose and Dose Rate in 6MV-FFF Beams
    Assalmi, Mustapha
    Diaf, El Yamani
    [J]. MOSCOW UNIVERSITY PHYSICS BULLETIN, 2021, 76 (SUPPL 1) : S36 - S51
  • [27] Monte Carlo Simulation of Surface Dose and Dose Rate in 6MV-FFF Beams
    Mustapha Assalmi
    El Yamani Diaf
    [J]. Moscow University Physics Bulletin, 2021, 76 : S36 - S51
  • [28] Suitability of Athena Monolithic Active Pixel Sensor for upstream monitoring of 6 MV, 6 FFF and 10 MV treatments
    Beck, Lana
    Ballisat, Laura
    Fletcher, Sally
    Velthuis, Jaap
    [J]. APPLIED RADIATION AND ISOTOPES, 2023, 198
  • [29] Spectral reconstruction for a 6 MV linear accelerator
    Hernández-Bojórquez, M
    Martínez-Dávalos, A
    Lárraga, JM
    [J]. MEDICAL PHYSICS, 2004, 724 : 254 - 257
  • [30] Evaluation of photoneutron dose equivalent in 10 MV and 15 MV beams for wedge and open fields in the Elekta Versa HD linac
    Khilafath, Hajee Reyaz Ali Sahib
    Ganesan, Bharanidharan
    Sekar, Nandakumar
    Mohapatra, Dinakrushna
    Mahadevan, Pramod
    Vellingiri, Jayaprakash
    Prakasarao, Aruna
    Singaravelu, Ganesan
    [J]. APPLIED RADIATION AND ISOTOPES, 2022, 188