Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations

被引:45
|
作者
Cranmer-Sargison, Gavin [1 ,3 ]
Weston, Steve [2 ,3 ]
Sidhu, Narinder P. [1 ,4 ]
Thwaites, David I. [3 ,5 ]
机构
[1] Saskatchewan Canc Agcy, Dept Med Phys, Saskatoon, SK, Canada
[2] Leeds Teaching Hosp NHS Trust, St Jamess Inst Oncol, Leeds, W Yorkshire, England
[3] Univ Leeds, Acad Unit Med Phys, Leeds LS2 9JT, W Yorkshire, England
[4] Univ Saskatchewan, Dept Phys & Engn Phys, Saskatoon, SK S7N 0W0, Canada
[5] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
关键词
Small field; Diode detector; Dosimetry; PLASTIC SCINTILLATION DETECTORS; HEAD SCATTER FACTORS; PHOTON FIELDS; DOSIMETRY; BEAMS; RADIOTHERAPY; SIZE;
D O I
10.1016/j.radonc.2011.09.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The goal of this work was to measure 6 MV small field, detector specific, output ratios (ORdet) using the IBA stereotactic field diode (SFD) and the PTW T60008, T60012, T60016 and T60017 field diodes on both Varian iX and Elekta Synergy accelerators, to establish estimates for the experimental uncertainty and characterize the measurement precision under various conditions. Materials and methods: Data were acquired at depths of 1.5, 5.0 and 10.0 cm for square field sizes of 3.0, 1.0, 0.9, 0.8, 0.7, 0.6 and 0.5 cm. Three isocentric measurements comprised of five readings were made to calculate an experimental output ratio ORdet with respect to a field size of 5.0 cm. The coefficient of variation (CV) was calculated to characterize the precision associated with each detector-linac combination. Another measurement set was made to investigate the influence of jaw position accuracy. Results: As expected for field sizes smaller than 3.0 cm, the measured ORdet were not consistent across all detectors. The standard percent uncertainty in measured ORdet was found to be nearly consistent across all detector-linac combinations: less than +/- 0.25% for the 3.0 cm field size, increasing to approximately +/- 1.25% for the smallest field sizes. As the field size was reduced to 0.5 cm the CV increased to 0.10% and 0.15% on the Varian and Elekta linacs, respectively. Conclusion: Experimental small field ORdet measured with the diode detectors used in this study are reproducible to within +/- 1.25% (standard uncertainty), with the precision of any one set of measurements can be characterized with a CV between 0.10% and 0.15%. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 100 (2011) 429-435
引用
收藏
页码:429 / 435
页数:7
相关论文
共 38 条
  • [1] EXPERIMENTAL SMALL FIELD OUTPUT FACTOR ANALYSIS FOR VARIOUS DIODE DETECTOR AND ACCELERATOR COMBINATIONS
    Cranmer-Sargison, G.
    Weston, S.
    Sidhu, N.
    Thwaites, D. I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S148 - S148
  • [2] Small field output factor measurements: a detector comparison for Versa HD 6 MV and 6 MV FFF beams.
    Monasor Denia, P.
    Castellet Garcia, M. C.
    Manjon Garcia, C.
    Lopez Tarjuelo, J.
    Juan Senabre, X. J.
    Quiros Higueras, J. D.
    De Marco Blancas, N.
    Bonaque Alandi, J.
    Montenegro Iglesias, N.
    Santos Serra, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S920 - S921
  • [3] Comparison of detector performance in small 6 MV and 6 MV FFF beams using a Versa HD accelerator
    Denia, Paula Monasor
    Castellet Garcia, Maria del Carmen
    Manjon Garcia, Carla
    Quiros Higueras, Juan David
    Blancas, Noelia de Marco
    Alandi, Jorge Bonaque
    Juan Senabre, Xavier Jordi
    Serra, Agustin Santos
    Lopez-Tarjuelo, Juan
    [J]. PLOS ONE, 2019, 14 (03):
  • [4] The characteristics of small field beam quality and output factor of 6 MV FFF
    Azzi, A.
    Ryangga, D.
    Pawiro, S. A.
    [J]. 18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [5] Evaluating the Performance of Various Detectors in the Small Field Size by 6 MV Linac
    Al Atawneh S.J.
    [J]. Iranian Journal of Medical Physics, 2023, 20 (01) : 1 - 6
  • [6] Determination of small field output factors in 6 and 10 MV flattening filter free photon beams using various detectors
    Masanga, W.
    Tangboonduangjit, P.
    Khamfongkhruea, C.
    Tannanonta, C.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [7] Evaluation of PTW MicroSilicon Diode Detector for Small Field Dosimetry with the Elekta Versa HD Linear Accelerator
    Kim, J.
    Hong, C.
    Kang, H.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E773 - E773
  • [8] A Comparison of Field Output Corrections Factors for a Diode Detector Among Various Measurement Setups and Among Various Energies for Stereotactic Cones
    Lee, Y.
    Kim, Y.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [9] Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams
    Gul, Attia
    Fukuda, Shigekazu
    Mizuno, Hideyuki
    Taku, Nakaji
    Kakakhel, M. Basim
    Mirza, Sikander M.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 23 - 36
  • [10] Comparison of synthetic Diamond Detector TM 60019 with other PTW Small-field Detector with 6 MV Photon Energy
    Scherf, C.
    Henrich, N.
    Koehn, J.
    Bluemer, N.
    Heine, B.
    Loutfi-Krauss, B.
    Mielke, M.
    Zink, K.
    Ramm, U.
    Licher, J.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 : 93 - 93