Characterization of 3-Dimensional Printing and Casting Materials for use in Magnetic Resonance Imaging Phantoms at 3 T

被引:6
|
作者
Yunker, B. E. [1 ,2 ]
Stupic, K. F. [1 ]
Wagner, J. L. [2 ]
Huddle, S. [2 ]
Shandas, R. [2 ]
Weir, R. F. [2 ]
Russek, S. E. [1 ]
Keenan, K. E. [1 ]
机构
[1] NIST, Phys Measurement Lab, Boulder, CO 80305 USA
[2] Univ Colorado Denver Anschutz, Aurora, CO 80045 USA
关键词
3D printing; medical imaging; MRI; phantom; polymer; RELAXATION-TIMES; BREAST-TISSUE; IN-VIVO; MRI; T-1; MODELS; SEQUENCES; RESECTION;
D O I
10.6028/jres.125.028
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Imaging phantoms are used to calibrate and validate the performance of magnetic resonance imaging (MRI) systems. Many new materials have been developed for additive manufacturing (three-dimensional [3D] printing) processes that may be useful in the direct printing or casting of dimensionally accurate, anatomically accurate, patient-specific, and/or biomimetic MRI phantoms. The T-1, T-2, and T-2* spin relaxation times of polymer samples were tested to discover materials for use as tissue mimics and structures in MRI phantoms. This study included a cohort of polymer compounds that was tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed at 3 T using inversion recovery, spin echo, and gradient echo sequences for T-1, T-2, and T-2*, respectively. T-1, T-2, and T-2* values were plotted with error bars to allow the reader to assess how well a polymer matches a tissue for a specific application. A correlation was performed between T-1, T-2, T-2* values and material density, elongation, tensile strength, and hardness. Two silicones, SI_XP-643 and SI_P-45, may be usable mimics for reported liver values; one silicone, SI_XP-643, may be a useful mimic for muscle; one silicone, SI_XP-738, may be a useful mimic for white matter; and four silicones, SI_P-15, SI_GI-1000, SI_GI-1040, and SI_GI-1110, may be usable mimics for spinal cord. Elongation correlated to T-2 (p = 0.0007), tensile strength correlated to T-1 (p = 0.002), T-2 (p = 0.0003), and T-2* (p = 0.003). The 80 samples not providing measurable signal with T-1, T-2, T-2* relaxation values too short to measure with the standard sequences, may be useful for MRI-invisible fixturing and medical devices at 3 T.
引用
收藏
页数:24
相关论文
共 50 条
  • [41] Quantification of transplanted iron oxide-labelled islet cells by 3-dimensional 3T magnetic resonance imaging (MRI)
    Ris, Frederic
    Crowe, Lindsey
    Morel, Philippe
    Armanet, Mathieu
    Masson, Solange
    Nielles-Vallespin, Sonia
    Bosco, Domenico
    Speier, Peter
    Vallee, Jean-Paul
    Berney, Thierry
    XENOTRANSPLANTATION, 2009, 16 (05) : 299 - 300
  • [42] QUANTIFICATION OF TRANSPLANTED IRON OXIDE-LABELLED ISLET CELLS BY 3-DIMENSIONAL 3T MAGNETIC RESONANCE IMAGING (MRI)
    Ris, Frederic
    Crowe, Lindsey
    Morel, Phillipe
    Armanet, Mathieu
    Masson, Solange
    Nielles-Vallespin, Sonia
    Bosco, Domenico
    Speier, Peter
    Vallee, Jean-Paul
    Berney, Thierry
    TRANSPLANT INTERNATIONAL, 2009, 22 : 45 - 45
  • [43] 3-DIMENSIONAL ELECTRON-SPIN RESONANCE IMAGING
    WOODS, RK
    BACIC, GG
    LAUTERBUR, PC
    SWARTZ, HM
    JOURNAL OF MAGNETIC RESONANCE, 1989, 84 (02): : 247 - 254
  • [44] 3-DIMENSIONAL DISTORTION IN MAGNETIC-RESONANCE IMAGES
    DESOTO, LA
    HAYNOR, DR
    CHOI, HS
    KIM, Y
    ROBERTS, TS
    IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 : 494 - 495
  • [45] A Fast 3-Dimensional Magnetic Resonance Imaging Reconstruction for Surgical Planning of Uterine Myomectomy
    Lee, Sa Ra
    Kim, Young Jae
    Kim, Kwang Gi
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2018, 33 (02)
  • [46] TRUE 3-DIMENSIONAL NUCLEAR MAGNETIC-RESONANCE IMAGING BY FOURIER RECONSTRUCTION ZEUGMATOGRAPHY
    LAI, CM
    JOURNAL OF APPLIED PHYSICS, 1981, 52 (03) : 1141 - 1145
  • [47] Definition of a Stereotactic 3-Dimensional Magnetic Resonance Imaging Template of the Human Insula COMMENTS
    Kasper, Ekkehard Matthias
    Chen, Clark
    NEUROSURGERY, 2013, 72 : 46 - 46
  • [48] 3-DIMENSIONAL DISPLAY OF CARDIAC STRUCTURES USING RECONSTRUCTED MAGNETIC-RESONANCE-IMAGING
    HOSONO, M
    NAKANO, Y
    URAYAMA, S
    KONISHI, J
    UOKAWA, K
    TANAKA, Y
    JOURNAL OF DIGITAL IMAGING, 1995, 8 (03) : 105 - 115
  • [49] A 3-dimensional magnetic resonance imaging model for the assessment of lumbosacral cerebrospinal fluid volume
    Grouper, S
    Sullivan, JT
    Parrish, TB
    Walker, MT
    McCarthy, RJ
    Wong, CA
    ANESTHESIOLOGY, 2004, 100 (05) : B44 - B44
  • [50] Diagnosis of herniated intervertebral disc assisted by 3-dimensional, multiaxial, magnetic resonance imaging
    Hsieh, MS
    Tsai, MD
    JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 1999, 98 (05) : 347 - 355