Characterization of 3-Dimensional Printing and Casting Materials for use in Magnetic Resonance Imaging Phantoms at 3 T

被引:6
|
作者
Yunker, B. E. [1 ,2 ]
Stupic, K. F. [1 ]
Wagner, J. L. [2 ]
Huddle, S. [2 ]
Shandas, R. [2 ]
Weir, R. F. [2 ]
Russek, S. E. [1 ]
Keenan, K. E. [1 ]
机构
[1] NIST, Phys Measurement Lab, Boulder, CO 80305 USA
[2] Univ Colorado Denver Anschutz, Aurora, CO 80045 USA
关键词
3D printing; medical imaging; MRI; phantom; polymer; RELAXATION-TIMES; BREAST-TISSUE; IN-VIVO; MRI; T-1; MODELS; SEQUENCES; RESECTION;
D O I
10.6028/jres.125.028
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Imaging phantoms are used to calibrate and validate the performance of magnetic resonance imaging (MRI) systems. Many new materials have been developed for additive manufacturing (three-dimensional [3D] printing) processes that may be useful in the direct printing or casting of dimensionally accurate, anatomically accurate, patient-specific, and/or biomimetic MRI phantoms. The T-1, T-2, and T-2* spin relaxation times of polymer samples were tested to discover materials for use as tissue mimics and structures in MRI phantoms. This study included a cohort of polymer compounds that was tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed at 3 T using inversion recovery, spin echo, and gradient echo sequences for T-1, T-2, and T-2*, respectively. T-1, T-2, and T-2* values were plotted with error bars to allow the reader to assess how well a polymer matches a tissue for a specific application. A correlation was performed between T-1, T-2, T-2* values and material density, elongation, tensile strength, and hardness. Two silicones, SI_XP-643 and SI_P-45, may be usable mimics for reported liver values; one silicone, SI_XP-643, may be a useful mimic for muscle; one silicone, SI_XP-738, may be a useful mimic for white matter; and four silicones, SI_P-15, SI_GI-1000, SI_GI-1040, and SI_GI-1110, may be usable mimics for spinal cord. Elongation correlated to T-2 (p = 0.0007), tensile strength correlated to T-1 (p = 0.002), T-2 (p = 0.0003), and T-2* (p = 0.003). The 80 samples not providing measurable signal with T-1, T-2, T-2* relaxation values too short to measure with the standard sequences, may be useful for MRI-invisible fixturing and medical devices at 3 T.
引用
收藏
页数:24
相关论文
共 50 条
  • [21] 3-DIMENSIONAL MAGNETIC-RESONANCE-IMAGING OF THE PROLACTIN-SECRETIN MICROADENOMAS
    GIRARD, N
    CORTESI, L
    CHABERTORSINI, V
    MAMAN, P
    BRUE, T
    JAQUET, P
    RAYBAUD, C
    ANNALES D ENDOCRINOLOGIE, 1992, 53 (01) : 8 - 15
  • [22] 3-DIMENSIONAL MAGNETIC-RESONANCE CARDIAC IMAGING SHOWS INITIAL PROMISE
    KIRN, TF
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 259 (15): : 2194 - 2195
  • [23] Diagnosis of Dental Abnormalities in Children Using 3-Dimensional Magnetic Resonance Imaging
    Tymofiyeva, Olga
    Proff, Peter C.
    Rottner, Kurt
    Duering, Markus
    Jakob, Peter M.
    Richter, Ernst-Juergen
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2013, 71 (07) : 1159 - 1169
  • [24] 3-DIMENSIONAL MAGNETIC-RESONANCE-IMAGING EVALUATION OF PEDIATRIC TRACHEOBRONCHIAL TREE
    DONNELLY, KJ
    BANK, ER
    PARKS, J
    GUSSACK, GS
    DAVENPORT, P
    TODD, NW
    SHEPHERD, S
    LARYNGOSCOPE, 1994, 104 (12): : 1425 - 1430
  • [25] OPTIMAL IMAGING STRATEGIES FOR 3-DIMENSIONAL NUCLEAR-MAGNETIC-RESONANCE MICROSCOPY
    AHN, CB
    CHU, WC
    JOURNAL OF MAGNETIC RESONANCE, 1991, 94 (03): : 455 - 470
  • [26] Definition of a Stereotactic 3-Dimensional Magnetic Resonance Imaging Template of the Human Insula
    Afif, Afif
    Becq, Guillaume
    Mertens, Patrick
    NEUROSURGERY, 2013, 72 : 35 - 46
  • [27] Prenatal Diagnosis of Seckel Syndrome on 3-Dimensional Sonography and Magnetic Resonance Imaging
    Napolitano, Raffaele
    Maruotti, Giuseppe M.
    Quarantelli, Mario
    Martinelli, Pasquale
    Paladini, Dario
    JOURNAL OF ULTRASOUND IN MEDICINE, 2009, 28 (03) : 369 - 374
  • [28] Magnetic resonance imaging and 3-dimensional analysis of external anal sphincter anatomy
    Hsu, Y
    Fenner, DE
    Weadock, WJ
    DeLancey, JOL
    OBSTETRICS AND GYNECOLOGY, 2005, 106 (06): : 1259 - 1265
  • [29] ARTIFACTS DUE TO RESIDUAL MAGNETIZATION IN 3-DIMENSIONAL MAGNETIC-RESONANCE IMAGING
    WOOD, ML
    RUNGE, VM
    MEDICAL PHYSICS, 1988, 15 (06) : 825 - 831
  • [30] New process and materials developments in 3-dimensional printing, 3DP™
    Uhland, SA
    Holman, RK
    Cima, MJ
    Sachs, E
    Enokido, Y
    SOLID FREEFORM AND ADDITIVE FABRICATION, 1999, 542 : 153 - 158