Are conventional reconstruction plates equivalent to precontoured locking plates for distal humerus fracture fixation? A biomechanics cadaver study

被引:30
|
作者
Koonce, Ryan C. [1 ]
Baldini, Todd H. [2 ]
Morgan, Steven J. [3 ]
机构
[1] Skagit Reg Clin, Mt Vernon, WA 98274 USA
[2] Univ Colorado, Sch Med, Dept Orthopaed Surg, Aurora, CO 80045 USA
[3] Swedish Med Ctr, MOTUS Orthopaed, Englewood, CO 80113 USA
关键词
Humerus fracture; Fixation; Locking plate; Biomechanics; Cadaver; PLATING SYSTEMS; PARALLEL;
D O I
10.1016/j.clinbiomech.2012.03.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The optimal plate type and configuration for distal humerus fracture fixation has yet to be defined. Available biomechanical studies show conflicting results. No existing studies compare conventional reconstruction plates to newer precontoured distal humerus locking plates in both parallel and perpendicular configurations. Methods: Three groups of humerus specimens were compared via biomechanical testing in a cadaver model simulating metaphyseal comminution. Group 1 consisted of conventional reconstruction plates in a perpendicular configuration. Group 2 used precontoured locking plates in a perpendicular configuration. Group 3 used precontoured locking plates in a parallel configuration. Each group was tested for stiffness in anterior bending, posterior bending, axial compression, and torsion. The specimens then underwent cyclic loading followed by single load to failure in posterior bending. Findings: There was no significant difference between the three groups for anterior bending, posterior bending, axial compression, or torsional stiffness. There was no significant difference in load to failure for any of the three groups. Screw loosening was significantly higher in Group 1 when compared to Groups 2 and 3 after cyclic loading. Interpretation: In the early postoperative period, less expensive perpendicular conventional reconstruction plate constructs provide similar stiffness and load to failure properties to newer precontoured locking plate systems regardless of plate configuration. (c) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:697 / 701
页数:5
相关论文
共 50 条
  • [21] Percutaneous Fixation of Distal Tibial Fractures Using Locking Plates
    Shanmugam, Chezhiyan
    Rahmatalla, Aziz
    Maffulli, Nicola
    TECHNIQUES IN ORTHOPAEDICS, 2007, 22 (03) : 162 - 166
  • [22] Management of comminuted fractures of the distal humerus: clinical outcome after primary external fixation versus immediate fixation with locking plates
    Clemens Kösters
    Simon Lenschow
    Esther Schulte-Zurhausen
    Steffen Roßlenbroich
    Michael J. Raschke
    Benedikt Schliemann
    Archives of Orthopaedic and Trauma Surgery, 2017, 137 : 1693 - 1698
  • [23] Management of comminuted fractures of the distal humerus: clinical outcome after primary external fixation versus immediate fixation with locking plates
    Koesters, Clemens
    Lenschow, Simon
    Schulte-Zurhausen, Esther
    Rolenbroich, Steffen
    Raschke, Michael J.
    Schliemann, Benedikt
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2017, 137 (12) : 1693 - 1698
  • [24] Biomechanical Design Optimization of Distal Humerus Fracture Plates: A Review
    Zdero, Radovan
    Brzozowski, Pawel
    Schemitsch, Emil H.
    BIOMED RESEARCH INTERNATIONAL, 2024, 2024
  • [25] Locking Plates for Distal Femur Fractures: Is There a Problem With Fracture Healing?
    Henderson, Christopher E.
    Kuhl, Lori L.
    Fitzpatrick, Daniel C.
    Marsh, J. L.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 : S8 - S14
  • [26] Biomechanical Evaluation of the Fixation Methods for Transcondylar Fracture of the Humerus: ONI Plate Versus Conventional Plates and Screws
    Shimamura, Yasunori
    Nishida, Keiichiro
    Imatani, Junya
    Noda, Tomoyuki
    Hashizume, Hiroyuki
    Ohtsuka, Aiji
    Ozaki, Toshifumi
    ACTA MEDICA OKAYAMA, 2010, 64 (02) : 115 - 120
  • [27] Analysis of 101 Mechanical Failures in Distal Femur Fractures Treated with 3 Generations of Precontoured Locking Plates
    Collinge, Cory Alan
    Reeb, Alexander Francis
    Rodriguez-Buitrago, Andres Felipe
    Archdeacon, Michael T.
    Beltran, Michael J.
    Gardner, Michael J.
    Jeray, Kyle James
    Miller, Anna N.
    Crist, Brett D.
    Sems, Stephen A.
    Shah, Nihar Samir
    Fogel, Nathaniel
    Tibbo, Meagan
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2023, 37 (01) : 8 - 13
  • [28] Fixation of Distal Femur Fractures With the Use of Periarticular Tibial Locking Plates
    Friedman, Lisa G. M.
    Maniar, Hemil
    Horwitz, Daniel S.
    TECHNIQUES IN ORTHOPAEDICS, 2024, 39 (02) : 55 - 58
  • [29] Biomechanical Testing of Additive Manufactured Proximal Humerus Fracture Fixation Plates
    Maryam Tilton
    April Armstrong
    Jennifer Sanville
    Matthew Chin
    Michael W. Hast
    Gregory S. Lewis
    Guha P. Manogharan
    Annals of Biomedical Engineering, 2020, 48 : 463 - 476
  • [30] Biomechanical Testing of Additive Manufactured Proximal Humerus Fracture Fixation Plates
    Tilton, Maryam
    Armstrong, April
    Sanville, Jennifer
    Chin, Matthew
    Hast, Michael W.
    Lewis, Gregory S.
    Manogharan, Guha P.
    ANNALS OF BIOMEDICAL ENGINEERING, 2020, 48 (01) : 463 - 476