Biomechanical Design Optimization of Distal Humerus Fracture Plates: A Review

被引:0
|
作者
Zdero, Radovan [1 ]
Brzozowski, Pawel [1 ]
Schemitsch, Emil H. [1 ,2 ]
机构
[1] Victoria Hosp, Orthopaed Biomech Lab, London, ON, Canada
[2] Western Univ, Div Orthopaed Surg, London, ON, England
关键词
biomechanics; distal humerus; fracture; optimization; plates; review; INTERNAL-FIXATION; PLATING SYSTEMS; LOCKING PLATES; PROPHYLACTIC FIXATION; RECONSTRUCTION PLATES; PROXIMAL HUMERUS; IMPLANT FAILURE; BONE; PARALLEL; STABILITY;
D O I
10.1155/2024/6015794
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The goal of this article was to review studies on distal humerus fracture plates (DHFPs) to understand the biomechanical influence of systematically changing the plate or screw variables. The problem is that DHFPs are commonly used surgically, although complications can still occur, and it is unclear if implant configurations are always optimized using biomechanical criteria. A systematic search of the PubMed database was conducted to identify English-language biomechanical optimization studies of DHFPs that parametrically altered plate and/or screw variables to analyze their influence on engineering performance. Intraarticular and extraarticular fracture (EAF) data were separated and organized under commonly used biomechanical outcome metrics. The results identified 52 eligible DHFP studies, which evaluated various plate and screw variables. The most common plate variables evaluated were geometry, hole type, number, and position. Fewer studies assessed screw variables, with number and angle being the most common. However, no studies examined nonmetallic materials for plates or screws, which may be of interest in future research. Also, articles used various combinations of biomechanical outcome metrics, such as interfragmentary fracture motion, bone, plate, or screw stress, number of loading cycles to failure, and overall stiffness (Os) or failure strength (Fs). However, no study evaluated the bone stress under the plate to examine bone "stress shielding," which may impact bone health clinically. Surgeons treating intraarticular and extraarticular distal humerus fractures should seriously consider two precontoured, long, thick, locked, and parallel plates that are secured by long, thick, and plate-to-plate screws that are located at staggered levels along the proximal parts of the plates, as well as an extra transfracture plate screw. Also, research engineers could improve new studies by perusing recommendations in future work (e.g., studying alternative nonmetallic materials or "stress shielding"), clinical ramifications (e.g., benefits of locked plates), and study quality (e.g., experimental validation of computational studies).
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Biomechanical design optimization of proximal humerus locked plates: A review
    Zdero, Radovan
    Brzozowski, Pawel
    Schemitsch, Emil H.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2024, 55 (02):
  • [2] Biomechanical design optimization of distal femur locked plates: A review
    Zdero, Radovan
    Gide, Kunal
    Brzozowski, Pawel
    Schemitsch, Emil H.
    Bagheri, Z. Shaghayegh
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2023, 237 (07) : 791 - 805
  • [3] Biomechanical Design Optimization of Clavicle Midshaft Fracture Plates: A Review
    Zdero, Radovan
    Brzozowski, Pawel
    Schemitsch, Emil H.
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2024, 44 (03) : 307 - 316
  • [4] A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates
    Korner, J
    Diederichs, G
    Arzdorf, M
    Lill, H
    Josten, C
    Schneider, E
    Linke, B
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (05) : 286 - 293
  • [5] Comparison of biomechanical characteristics in treatment of distal humerus fracture using different fixation plates: A finite element study
    Zhong, Hua
    Ma, Limin
    Wang, Ming
    Huang, Yan
    Zhou, Ye
    Chen, Lingling
    Chen, Guangmou
    Li, Jianwei
    Chen, Jin
    Zhang, Yu
    MICROSCOPY RESEARCH AND TECHNIQUE, 2019, 82 (03) : 199 - 205
  • [6] Biomechanical Testing of Additive Manufactured Proximal Humerus Fracture Fixation Plates
    Tilton, Maryam
    Armstrong, April
    Sanville, Jennifer
    Chin, Matthew
    Hast, Michael W.
    Lewis, Gregory S.
    Manogharan, Guha P.
    ANNALS OF BIOMEDICAL ENGINEERING, 2020, 48 (01) : 463 - 476
  • [7] Biomechanical Testing of Additive Manufactured Proximal Humerus Fracture Fixation Plates
    Maryam Tilton
    April Armstrong
    Jennifer Sanville
    Matthew Chin
    Michael W. Hast
    Gregory S. Lewis
    Guha P. Manogharan
    Annals of Biomedical Engineering, 2020, 48 : 463 - 476
  • [8] Distal fracture of the humerus
    Rueger, J. M.
    Ruecker, A.
    Briem, D.
    CHIRURG, 2007, 78 (10): : 959 - 971
  • [9] Standardized fracture creation in the distal humerus and the olecranon for surgical training and biomechanical testing
    Schmoelz, Werner
    Zierleyn, Jan Philipp
    Hoermann, Romed
    Arora, Rohit
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (12) : 3853 - 3861
  • [10] Standardized fracture creation in the distal humerus and the olecranon for surgical training and biomechanical testing
    Werner Schmoelz
    Jan Philipp Zierleyn
    Romed Hoermann
    Rohit Arora
    Archives of Orthopaedic and Trauma Surgery, 2022, 142 : 3853 - 3861