Determination of the small-field output factor for 6 MV photon beam using EGSnrc Monte Carlo

被引:0
|
作者
Chuah, K. W. [1 ]
Aziz, Abdul M. Z. [2 ]
Jayamani, J. [1 ]
机构
[1] Univ Sains Malaysia, Sch Hlth Sci, Med Radiat Programme, Hlth Campus, Kubang Kerian 16150, Kelantan, Malaysia
[2] Univ Sains Malaysia, Adv Med & Dent Inst, Dept Biomed Imaging, Kepala Batas 13200, Pulau Pinang, Malaysia
关键词
Collimator scatter factor; phantom scatter factor; linear accelerator; SIMULATION; SIZE;
D O I
10.4103/jmp.jmp_40_22
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Accuracy of ionization chamber (IC) to measure the scatter output factor (Scp) of a linear accelerator (linac) is crucial, especially in small field (<4 cm x 4 cm). The common IC volume of 0.6 cc is not adequate for small-field measurement and not all radiotherapy centers can afford to purchase additional IC due to the additional cost. This study aimed to determine the efficiency of the EGSnrc Monte Carlo (MC) to calculate the Scp for various field sizes including small field in Elekta Synergy (Agility multileaf collimator) linac. The BEAMnrc and DOSXYZnrc user codes were used to simulate a 6 MV linac model for various field sizes and calculate the radiation dose output in water phantom. The modeled linac treatment head was validated by comparing the percentage depth dose (PDD), beam profile, and beam quality (Tissue Phantom Ratio (TPR)20,10) with the IC measurement. The validated linac model was simulated to calculate the Scp consisting of collimator scatter factor (Sc) and phantom scatter factor (Sp). The PDD and beam profile of the simulated field sizes were within a good agreement of +/- 2% compared with the measured data. The TPR20,10 value was 0.675 for field size 10 cm x 10 cm. The Scp, Sc, and Sp simulated values were close to the IC measurement within +/- 2% difference. The simulation for Sc and Sp in 3 cm x 3 cm field size was calculated to be 0.955 and 0.884, respectively. In conclusion, this study validated the efficiency of the MC simulation as a promising tool for the Scp calculation including small-field size for linac.
引用
收藏
页码:301 / 308
页数:8
相关论文
共 50 条
  • [31] Experimental and Monte Carlo Based Determination of Magnetic Field Correction Factors KB in 6 MV Photon Fields
    Alissa, M.
    Czarnecki, D.
    Schoenfeld, A.
    Kapsch, R.
    Frick, S.
    Zink, K.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E317 - E318
  • [32] Measured and Monte Carlo calculated output ratios in air for 6 and 18 MV photon beams
    Ding, G
    Yu, C
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1900 - 1900
  • [33] Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams
    Abdullah, R.
    Sapeei, M. S.
    Shukor, N. S. A.
    Yusof, M. F. M.
    Mohamed, M.
    Idris, N. R. N.
    Yusoff, A. L.
    Zakaria, A.
    Aziz, M. Z. A.
    [J]. INTERNATIONAL NUCLEAR SCIENCE, TECHNOLOGY AND ENGINEERING CONFERENCE 2017 (INUSTEC2017), 2018, 298
  • [34] Comparison of small-field output factor measurements
    Oliver, C.
    Takau, V.
    Butler, D.
    Williams, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S765 - S765
  • [35] Determination of initial electron parameters by means of Monte Carlo simulations for the Siemens Artiste Linac 6 MV photon beam
    Tugrul, Taylan
    Erogul, Osman
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (04) : 331 - 337
  • [36] EGSNRC Monte Carlo study of the effect of photon energy and field margin in phantoms simulating small lung lesions
    Osei, EK
    Darko, J
    Mosseri, A
    Jezioranski, J
    [J]. MEDICAL PHYSICS, 2003, 30 (10) : 2706 - 2714
  • [37] Properties of IBA Razor Nano Chamber in small-field radiation therapy using 6MV FF, 6 MV FFF, and 10 MV FFF photon beams
    Partanen, Mari
    Niemela, Jarkko
    Ojala, Jarkko
    Keyrilainen, Jani
    Kapanen, Mika
    [J]. ACTA ONCOLOGICA, 2021, 60 (11) : 1419 - 1424
  • [38] Small-field dosimetry with multiple detectors and Monte Carlo calculations
    Parker, B
    Shin, A
    Liu, H
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1372 - 1372
  • [39] Validation of PRIMO Monte Carlo Model of Clinac®iX 6MV Photon Beam
    Sarin, B.
    Bindhu, B.
    Saju, B.
    Nair, Raguram K.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2020, 45 (01) : 24 - 35
  • [40] Monte Carlo Commissioning of 6 MV Photon Beam of Varian Linear Accelerator Installed at INOR
    Bajwa, Saqib
    Gul, Attia
    Ahmed, Shahbaz
    Kakakhel, Muhammad Basim
    [J]. 2016 INTERNATIONAL CONFERENCE ON EMERGING TECHNOLOGIES (ICET), 2016,