Increased Femoral Head Offset is Associated With Elevated Metal Ions in Asymptomatic Patients With Metal-on-Polyethylene Total Hip Arthroplasty

被引:19
|
作者
Martin, John R. [1 ]
Camp, Christopher L. [1 ]
Wyles, Cody C. [1 ]
Taunton, Michael J. [1 ]
Trousdale, Robert T. [1 ]
Lewallen, David G. [1 ]
机构
[1] Mayo Clin, Dept Orthoped Surg, 200 First St SW, Rochester, MN 55905 USA
来源
JOURNAL OF ARTHROPLASTY | 2016年 / 31卷 / 12期
关键词
total hip arthroplasty; metal ions; cobalt; chrome; head size; offset; NECK TAPER; CORROSION; TRUNNIONS;
D O I
10.1016/j.arth.2016.05.047
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Predisposing factors for trunnionosis and elevated metal ion levels in metal-on-polyethylene (MOP) total hip arthroplasty (THA) are currently unknown. Methods: This retrospective cohort study enrolled 80 consecutive patients (43 males) with an asymptomatic MOP THA at 2- to 5-year follow-up and no other metal implants. Serum cobalt (Co) and chromium (Cr) levels were collected at the time of enrollment, and retrospective review was performed regarding demographic, implant, and surgical characteristics. Mean age at the time of surgery was 65.7 years (range 35.6-85.9 years), and mean postoperative follow-up was 28.7 months (range 24.4-58.9 months). Results: Femoral head offset was the only evaluated factor shown to increase serum Co ion levels above baseline within the cohort. Mean difference in Co level for high and low offset implants was 0.58 ppb (95% confidence interval [CI] = 0.05-1.11 ppb; P = .03). Mean difference in Cr level for high and low offset implants was 0.19 ppb (95% CI = -0.23 to 0.60 ppb; P = .37). Mean difference in Co level for small and large femoral heads was 0.20 ppb (95% CI = -0.41 to 0.81 ppb; P = .59). Mean difference in Cr level for small and large femoral heads was 0.28 ppb (95% CI = -0.18 to 0.74 ppb; P = .06). Age, gender, Harris Hip Score, and implant duration were not associated with changes in metal ion levels. Conclusion: Femoral head offset appears to be an important source of elevated metal ion levels in MOP THA. Further studies will be needed to understand if increasing femoral head offset is associated with subsequent adverse local tissue reactions. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:2814 / 2818
页数:5
相关论文
共 50 条
  • [41] Sensitivity and specificity for detecting pseudotumors in patients with hip resurfacing arthroplasty, metal-on-metal or metal-on-polyethylene total hip arthroplasty-MRI versus ultrasonography performed by an orthopedic surgery resident
    Mikkelsen, Rasmus T.
    Schou, Martin
    Torfing, Trine
    Graumann, Ole
    Overgaard, Soren
    Varnum, Claus
    ACTA RADIOLOGICA OPEN, 2023, 12 (02)
  • [42] Elevated Intra-Articular Cobalt and Chromium Levels in Mechanically Assisted Crevice Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty
    McGrory, Brian J.
    Payson, Abigail M.
    MacKenzie, Johanna A.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (05): : 1654 - 1658
  • [43] Atypical pseudotumour after metal-on-polyethylene total hip arthroplasty causing deep venous thrombosis
    Picardo, Natasha E.
    Al-Khateeb, Hesham
    Pollock, Rob C.
    HIP INTERNATIONAL, 2011, 21 (06) : 762 - 765
  • [44] Effect of femoral component offset on polyethylene wear in total hip arthroplasty
    Sakalkale, DP
    Sharkey, PF
    Eng, K
    Hozack, WJ
    Rothman, RH
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (388) : 125 - 134
  • [45] MRI of asymptomatic patients with metal-on-metal and polyethylene-on-metal total hip arthroplasties
    Mistry, A.
    Cahir, J.
    Donell, S. T.
    Nolan, J.
    Toms, A. P.
    CLINICAL RADIOLOGY, 2011, 66 (06) : 540 - 545
  • [46] Serum Metal Levels for Diagnosis of Adverse Local Tissue Reactions Secondary to Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty
    Fillingham, Yale A.
    Della Valle, Craig J.
    Bohl, Daniel D.
    Kelly, Mick P.
    Hall, Deborah J.
    Pourzal, Robin
    Jacobs, Joshua J.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (09): : S272 - S277
  • [47] Trunnion corrosion in metal-on-polyethylene hip arthroplasty A SIMPLE DIAGNOSIS AND TREATMENT?
    Lachiewicz, P. F.
    O'Dell, J. A.
    BONE & JOINT JOURNAL, 2018, 100B (07): : 898 - 902
  • [48] Pseudotumor and deep venous thrombosis due to crevice corrosion of the head-neck junction in metal-on-polyethylene total hip arthroplasty
    Watanabe, Hiroshi
    Takahashi, Kenji
    Takenouchi, Kenji
    Sato, Akiko
    Kawaji, Hidemi
    Nakamura, Hiroshi
    Takai, Shinro
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2015, 20 (06) : 1142 - 1147
  • [49] Outcome of Re-Revision Surgery for Adverse Local Tissue Reaction in Metal-on-Polyethylene and Metal-on-Metal Total Hip Arthroplasty
    Bonner, Bryant
    Arauz, Paul
    Klemt, Christian
    Kwon, Young-Min
    JOURNAL OF ARTHROPLASTY, 2020, 35 (06): : S284 - S288
  • [50] Ion release in patients with metal-on-metal hip bearings in total joint replacement: A comparison with metal-on-polyethylene bearings
    Savarino, L
    Granchi, D
    Ciapetti, G
    Cenni, E
    Pantoli, AN
    Rotini, R
    Veronesi, CA
    Baldini, N
    Giunti, A
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 63 (05): : 467 - 474