Increased Femoral Head Offset is Associated With Elevated Metal Ions in Asymptomatic Patients With Metal-on-Polyethylene Total Hip Arthroplasty

被引:19
|
作者
Martin, John R. [1 ]
Camp, Christopher L. [1 ]
Wyles, Cody C. [1 ]
Taunton, Michael J. [1 ]
Trousdale, Robert T. [1 ]
Lewallen, David G. [1 ]
机构
[1] Mayo Clin, Dept Orthoped Surg, 200 First St SW, Rochester, MN 55905 USA
来源
JOURNAL OF ARTHROPLASTY | 2016年 / 31卷 / 12期
关键词
total hip arthroplasty; metal ions; cobalt; chrome; head size; offset; NECK TAPER; CORROSION; TRUNNIONS;
D O I
10.1016/j.arth.2016.05.047
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Predisposing factors for trunnionosis and elevated metal ion levels in metal-on-polyethylene (MOP) total hip arthroplasty (THA) are currently unknown. Methods: This retrospective cohort study enrolled 80 consecutive patients (43 males) with an asymptomatic MOP THA at 2- to 5-year follow-up and no other metal implants. Serum cobalt (Co) and chromium (Cr) levels were collected at the time of enrollment, and retrospective review was performed regarding demographic, implant, and surgical characteristics. Mean age at the time of surgery was 65.7 years (range 35.6-85.9 years), and mean postoperative follow-up was 28.7 months (range 24.4-58.9 months). Results: Femoral head offset was the only evaluated factor shown to increase serum Co ion levels above baseline within the cohort. Mean difference in Co level for high and low offset implants was 0.58 ppb (95% confidence interval [CI] = 0.05-1.11 ppb; P = .03). Mean difference in Cr level for high and low offset implants was 0.19 ppb (95% CI = -0.23 to 0.60 ppb; P = .37). Mean difference in Co level for small and large femoral heads was 0.20 ppb (95% CI = -0.41 to 0.81 ppb; P = .59). Mean difference in Cr level for small and large femoral heads was 0.28 ppb (95% CI = -0.18 to 0.74 ppb; P = .06). Age, gender, Harris Hip Score, and implant duration were not associated with changes in metal ion levels. Conclusion: Femoral head offset appears to be an important source of elevated metal ion levels in MOP THA. Further studies will be needed to understand if increasing femoral head offset is associated with subsequent adverse local tissue reactions. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:2814 / 2818
页数:5
相关论文
共 50 条
  • [31] Risk Factors Associated With Early Complications of Revision Surgery for Head-Neck Taper Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty
    Kwon, Young-Min
    Rossi, Daniel
    MacAuliffe, John
    Peng, Yun
    Arauz, Paul
    JOURNAL OF ARTHROPLASTY, 2018, 33 (10): : 3231 - 3237
  • [32] Systemic lupus erythematosus induced by adjuvants after metal-on-polyethylene total hip arthroplasty
    Wang Chang-Chao
    Huang Yan
    Huang Ya-Dong
    中华医学杂志(英文版), 2020, 133 (12) : 1499 - 1500
  • [33] Systemic lupus erythematosus induced by adjuvants after metal-on-polyethylene total hip arthroplasty
    Wang, Chang-Chao
    Huang, Yan
    Huang, Ya-Dong
    CHINESE MEDICAL JOURNAL, 2020, 133 (12) : 1499 - 1500
  • [34] Gait analysis of patients with metal-on-metal resurfacing hip arthroplasty compared with big-femoral-head total hip arthroplasty
    Chen, Yun-Su
    Zhao, Song
    Cao, Le
    Zhang, Xian-Long
    SAUDI MEDICAL JOURNAL, 2011, 32 (04) : 394 - 399
  • [35] Ultrasound findings in asymptomatic patients with modular metal on metal total hip arthroplasty
    Nicholas B. Frisch
    Nolan M. Wessell
    Kevin Taliaferro
    Marnix Van Holsbeeck
    Craig D. Silverton
    Skeletal Radiology, 2017, 46 : 641 - 649
  • [36] Ultrasound findings in asymptomatic patients with modular metal on metal total hip arthroplasty
    Frisch, Nicholas B.
    Wessell, Nolan M.
    Taliaferro, Kevin
    Van Holsbeeck, Marnix
    Silverton, Craig D.
    SKELETAL RADIOLOGY, 2017, 46 (05) : 641 - 649
  • [37] Metal-on-metal vs. metal-on-polyethylene total hip arthroplasty tribological evaluation of retrieved components and periprosthetic tissue
    Topolovec, Matev
    Coer, Andrej
    Milosev, Ingrid
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 34 : 243 - 252
  • [38] Massive pseudotumor of unknown etiology in a cemented metal-on-polyethylene total hip arthroplasty: a case report
    Gazendam, Aaron
    Masrouha, Karim
    Popovic, Snezana
    Ghert, Michelle
    Wilson, David
    ANNALS OF JOINT, 2022,
  • [39] Bearing Change to Metal-On-Polyethylene for Ceramic Bearing Fracture in Total Hip Arthroplasty; Does It Work?
    Lee, Soong Joon
    Kwak, Hong Suk
    Yoo, Jeong Joon
    Kim, Hee Joong
    JOURNAL OF ARTHROPLASTY, 2016, 31 (01): : 204 - 208
  • [40] Evidence-Based Management of Trunnionosis in Metal-on-Polyethylene Total Hip Arthroplasty: A Systematic Review
    Sultan, Assem A.
    Cantrell, William A.
    Khlopas, Anton
    Berger, Ryan J.
    Sodhi, Nipun
    Molloy, Robert M.
    Krebs, Viktor E.
    Mont, Michael A.
    JOURNAL OF ARTHROPLASTY, 2018, 33 (10): : 3343 - 3353