Assessment of matrix effects associated with Fe isotope analysis using 266 nm femtosecond and 193 nm nanosecond laser ablation multi-collector inductively coupled plasma mass spectrometry

被引:48
|
作者
Zheng, Xin-Yuan [1 ,2 ]
Beard, Brian L. [1 ,2 ]
Johnson, Clark M. [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Geosci, 1215 W Dayton St, Madison, WI 53706 USA
[2] NASA, Astrobiol Inst, Washington, DC 20546 USA
关键词
MC-ICP-MS; LANGMUIR PROBE MEASUREMENTS; HIGH-PRECISION FE; IN-SITU; ELECTRON-DENSITY; SPACE-CHARGE; NONTRADITIONAL ISOTOPES; ELEMENTAL FRACTIONATION; ANALYTICAL-CHEMISTRY; IRON-METEORITES;
D O I
10.1039/c7ja00272f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study evaluated matrix effects during Fe isotope analysis using a 266 nm femtosecond (fs) laser and a 193 nm nanosecond (ns) laser coupled to a multi-collector ICP-MS. During 150 second spot analysis on pyrite, Fe isotope fractionation was not observed for fs-laser ablation (fs-LA), but was evident for ns-LA. The observed downhole Fe isotope fractionation during ns-LA is caused by multiple processes comprising ablation, transport, and ionization in the ICP. Contrary to the common perception of "matrix-free" analysis, matrix effects clearly exist during fs-LA analysis; small deviations of up to similar to 0.2 parts per thousand in the measured Fe-56/Fe-54 ratios from the true value of magnetite grains with >=similar to 8 wt% impurities were resolved using a nearly pure magnetite as the bracketing standard. Moreover, inaccurate and imprecise Fe-56/Fe-54 results were obtained when magnetite and pyrrhotite was measured against a non-matrix-matchedl standard (pyrite or Fe metal). The observed matrix effects during fs-LA cannot be explained by formation of a large heat-affected zone during ablation, but result from the influence of different chemical compositions of samples and standards on space-charge effects in the ICP-MS. Such matrix effects can be largely suppressed by water addition during analysis at a price of reduced sensitivity, so that precise and accurate Fe isotope analysis to a similar to 0.1 parts per thousand level can be routinely achieved under "wet" conditions without matrix-matching between sample and standard. These results may reconcile dramatically different precisions previously reported for Fe isotope analysis by fs-lasers, and also highlight fs-LA-MC-ICP-MS as an appealing option for in situ Fe isotope analysis on samples with complex matrices and high-symmetry minerals, both of which encounter significant analytical difficulties using secondary ion mass spectrometry (SIMS). For ns-LA, in addition to similar composition-related matrix effects experienced by fs-LA in the ICP, matrix effects also originate from ablation-related processes that produce sample particles with matrix-dependent size distributions and, often, larger aerodynamic sizes, resulting in highly inaccurate Fe-56/Fe-54 results during non-matrix-matched analysis under "dry" conditions. The collective matrix effects during Fe isotope analysis by ns-LA cannot be fully suppressed by water addition, therefore, matrix matching is required for accurate Fe isotope analysis by ns-LA.
引用
收藏
页码:68 / 83
页数:16
相关论文
共 50 条
  • [21] High precision in-situ Pb isotopic analysis of sulfide minerals by femtosecond laser ablation multi-collector inductively coupled plasma mass spectrometry
    YUAN HongLin
    YIN Cong
    LIU Xu
    CHEN KaiYun
    BAO ZhiAn
    ZONG ChunLei
    DAI MenNing
    LAI ShaoCong
    WANG Rong
    JIANG ShaoYong
    Science China(Earth Sciences), 2015, 58 (10) : 1713 - 1721
  • [22] High precision in-situ Pb isotopic analysis of sulfide minerals by femtosecond laser ablation multi-collector inductively coupled plasma mass spectrometry
    HongLin Yuan
    Cong Yin
    Xu Liu
    KaiYun Chen
    ZhiAn Bao
    ChunLei Zong
    MenNing Dai
    ShaoCong Lai
    Rong Wang
    ShaoYong Jiang
    Science China Earth Sciences, 2015, 58 : 1713 - 1721
  • [23] Isotope Ratio Analysis of Lead in Blood and Environmental Samples by Multi-collector Inductively Coupled Plasma Mass Spectrometry
    Takagi, Mai
    Yoshinaga, Jun
    Tanaka, Atsushi
    Seyama, Haruhiko
    ANALYTICAL SCIENCES, 2011, 27 (01) : 29 - 35
  • [24] Isotope Ratio Analysis of Lead in Blood and Environmental Samples by Multi-collector Inductively Coupled Plasma Mass Spectrometry
    Mai Takagi
    Jun Yoshinaga
    Atsushi Tanaka
    Haruhiko Seyama
    Analytical Sciences, 2011, 27 : 29 - 35
  • [25] Pb/U fractionation during Nd:YAG 213 nm and 266 nm laser ablation sampling with inductively coupled plasma mass spectrometry
    Liu, HC
    Borisov, OV
    Mao, XL
    Shuttleworth, S
    Russo, RE
    APPLIED SPECTROSCOPY, 2000, 54 (10) : 1435 - 1442
  • [26] Simultaneous measurement of sulfur and lead isotopes in sulfides using nanosecond laser ablation coupled with two multi-collector inductively coupled plasma mass spectrometers
    Yuan, Honglin
    Liu, Xu
    Chen, Lu
    Bao, Zhian
    Chen, Kaiyun
    Zong, Chunlei
    Li, Xiao-Chun
    Qiu, Johnson Wenhong
    JOURNAL OF ASIAN EARTH SCIENCES, 2018, 154 : 386 - 396
  • [27] Zirconium isotopic composition of zircon reference materials by laser ablation multi-collector inductively coupled plasma mass spectrometry
    Xie, Jin-Cheng
    Zhu, Di-Cheng
    Wang, Qing
    Zhang, Liang-Liang
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2023, 484
  • [28] In situ Mg isotope measurements of biogenic carbonates using laser ablation multi-collector inductively coupled plasma mass spectrometry: A new tool to understand biomineralisation
    Sadekov, Aleksey
    Lloyd, Nicholas S.
    Misra, Sambuddha
    D'Olivo, Juan Pablo
    McCulloch, Malcolm
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2020, 34 (23)
  • [29] Hf isotope ratio analysis using multi-collector inductively coupled plasma mass spectrometry: an evaluation of isobaric interference corrections
    Chu, NC
    Taylor, RN
    Chavagnac, V
    Nesbitt, RW
    Boella, RM
    Milton, JA
    German, CR
    Bayon, G
    Burton, K
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (12) : 1567 - 1574
  • [30] Influence of room temperature on magnesium isotope measurements by multi-collector inductively coupled plasma mass spectrometry
    Zhang, Xing-Chao
    Zhang, An-Yu
    Zhang, Zhao-Feng
    Huang, Fang
    Yu, Hui-Min
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2018, 32 (13) : 1026 - 1030