Isotope Ratio Analysis of Lead in Blood and Environmental Samples by Multi-collector Inductively Coupled Plasma Mass Spectrometry
被引:0
|作者:
Mai Takagi
Jun Yoshinaga
Atsushi Tanaka
Haruhiko Seyama
机构:
[1] The University of Tokyo,Department of Environmental Studies
[2] National Institute for Environmental Studies,undefined
来源:
关键词:
D O I:
暂无
中图分类号:
学科分类号:
摘要:
It is widely recognized that lead (Pb) affects children’s cognitive function, even at relatively low blood lead levels (<10 μg dL–1). The determination of the source of Pb in children is essential for effective risk management. The use of multi-collector ICPMS (MC-ICPMS) for isotope ratio measurements of Pb in environmental and biological samples was examined for this purpose. MC-ICPMS with an instrumental mass fractionation correction by Tl allowed accurate isotope ratio measurements of the Pb isotopic reference material NIST SRM 981. However, the presence of matrix elements (Al, Ca, Fe and Na) at more than 10 mg kg–1 in the sample solution significantly deteriorated the accuracy. The separation of Pb from the matrix is necessary for accurate measurements of the isotope ratio of Pb in environmental and biological samples. Bromide-complexation, followed by anion exchange was found to be satisfactory in terms of the recovery of Pb (90 to 104%) and the efficiency of matrix separation. The procedure was applied to a preliminary source analysis of Pb in the blood of Japanese children, and a significant contribution of indoor dust was demonstrated.