Influence of room temperature on magnesium isotope measurements by multi-collector inductively coupled plasma mass spectrometry

被引:11
|
作者
Zhang, Xing-Chao [1 ]
Zhang, An-Yu [2 ]
Zhang, Zhao-Feng [3 ]
Huang, Fang [1 ]
Yu, Hui-Min [1 ]
机构
[1] Univ Sci & Technol China, Dept Earth & Space Sci, CAS Key Lab Crust Mantle Mat & Environm, Hefei 230026, Anhui, Peoples R China
[2] Thermo Fisher Sci China, Shanghai 201206, Peoples R China
[3] Chinese Acad Sci, Guangzhou Inst Geochem, State Key Lab Isotope Geochem, Guangzhou 510640, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
NONTRADITIONAL STABLE-ISOTOPES; MC-ICP-MS; RATIO MEASUREMENTS; PLENARY LECTURE; SPACE-CHARGE; BIAS; GEOCHEMISTRY; PRECISE; MATRIX; ROCKS;
D O I
10.1002/rcm.8122
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
RationaleWe observed that the accuracy and precision of magnesium (Mg) isotope analyses could be affected if the room temperature oscillated during measurements. To achieve high-quality Mg isotopic data, it is critical to evaluate how the unstable room temperature affects Mg isotope measurements by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS). MethodsWe measured the Mg isotopes for the reference material DSM-3 using MC-ICP-MS under oscillating room temperatures in spring. For a comparison, we also measured the Mg isotopes under stable room temperatures, which were achieved by the installation of an improved temperature control system in the laboratory. ResultsThe Mg-26 values measured under oscillating room temperatures have a larger deviation (Mg-26 from -0.09 to 0.08, with average Mg-26=0.00 +/- 0.08 parts per thousand) than those measured under a stable room temperature (Mg-26 from -0.03 to 0.03 parts per thousand, with average Mg-26=0.00 +/- 0.02 parts per thousand) using the same MC-ICP-MS system. ConclusionsThe room temperature variation can influence the stability of MC-ICP-MS. Therefore, it is critical to keep the room temperature stable to acquire high-precision and accurate isotopic data when using MC-ICP-MS, especially when using the sample-standard bracketing (SSB) correction method.
引用
收藏
页码:1026 / 1030
页数:5
相关论文
共 50 条
  • [1] Polynomial mass bias functions for the internal standardization of isotope ratio measurements by multi-collector inductively coupled plasma mass spectrometry
    Doherty, William
    Gregoire, D. Conrad
    Bertrand, Nicole
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2008, 63 (03) : 407 - 414
  • [2] Precise and accurate Ga isotope ratio measurements of geological samples by multi-collector inductively coupled plasma mass spectrometry
    Zhang, Yuxu
    Qiao, Pan
    Zhu, Chuanwei
    Fan, Haifeng
    Wen, Hanjie
    [J]. ACTA GEOCHIMICA, 2024,
  • [3] Application of a "Continuous-Acquisition-Method" to potassium isotope measurements by multi-collector inductively coupled plasma mass spectrometry
    Gu, Hai-Ou
    Sun, He
    Huang, Chao
    Wang, Fangyue
    Ge, Can
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2021, 35 (13)
  • [4] Precise and accurate Ga isotope ratio measurements of geological samples by multi-collector inductively coupled plasma mass spectrometry
    Yuxu Zhang
    Pan Qiao
    Chuanwei Zhu
    Haifeng Fan
    Hanjie Wen
    [J]. Acta Geochimica., 2024, 43 (06) - 1064
  • [5] Isotope fractionation during ion beam formation in multi-collector inductively coupled plasma mass spectrometry
    Kivel, Niko
    Guenther-Leopold, Ines
    Vanhaecke, Frank
    Guenther, Detlef
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2012, 76 : 126 - 132
  • [6] Precise determination of rhenium isotope composition by multi-collector inductively-coupled plasma mass spectrometry
    Miller, Christian A.
    Peucker-Ehrenbrink, Bernhard
    Ball, Lary
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (08) : 1069 - 1078
  • [7] Adapting automated instrumentation for high-throughput calcium isotope measurements by multi-collector inductively coupled plasma mass spectrometry
    Razionale, Dan
    Bryant, Roger N.
    Blattler, Clara L.
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2022, 36 (07)
  • [8] Comparison of factors affecting the accuracy of high-precision magnesium isotope analysis by multi-collector inductively coupled plasma mass spectrometry
    Teng, Fang-Zhen
    Yang, Wei
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2014, 28 (01) : 19 - 24
  • [9] Isotope Ratio Analysis of Lead in Blood and Environmental Samples by Multi-collector Inductively Coupled Plasma Mass Spectrometry
    Takagi, Mai
    Yoshinaga, Jun
    Tanaka, Atsushi
    Seyama, Haruhiko
    [J]. ANALYTICAL SCIENCES, 2011, 27 (01) : 29 - 35
  • [10] Isotope Ratio Analysis of Lead in Blood and Environmental Samples by Multi-collector Inductively Coupled Plasma Mass Spectrometry
    Mai Takagi
    Jun Yoshinaga
    Atsushi Tanaka
    Haruhiko Seyama
    [J]. Analytical Sciences, 2011, 27 : 29 - 35