Image quality, threshold contrast and mean glandular dose in CR mammography

被引:3
|
作者
Jakubiak, R. R. [1 ]
Gamba, H. R. [2 ]
Neves, E. B. [3 ]
Peixoto, J. E. [4 ]
机构
[1] Univ Tecnol Fed Parana, Acad Dept Phys, BR-80230010 Curitiba, PR, Brazil
[2] Univ Tecnol Fed Parana, Grad Program Biomed Engn, BR-80230010 Curitiba, PR, Brazil
[3] Univ Tecnol Fed Parana, Grad Sch Elect Engn & Comp Sci CPGEI, BR-80230010 Curitiba, PR, Brazil
[4] Natl Canc Inst, BR-20231092 Rio De Janeiro, RJ, Brazil
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 18期
关键词
FIELD DIGITAL MAMMOGRAPHY; EUROPEAN PROTOCOL; MONTE-CARLO; OPTIMIZATION; PERFORMANCE; SYSTEMS;
D O I
10.1088/0031-9155/58/18/6565
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In many countries, computed radiography (CR) systems represent the majority of equipment used in digital mammography. This study presents a method for optimizing image quality and dose in CR mammography of patients with breast thicknesses between 45 and 75 mm. Initially, clinical images of 67 patients (group 1) were analyzed by three experienced radiologists, reporting about anatomical structures, noise and contrast in low and high pixel value areas, and image sharpness and contrast. Exposure parameters (kV, mAs and target/filter combination) used in the examinations of these patients were reproduced to determine the contrast-to-noise ratio (CNR) and mean glandular dose (MGD). The parameters were also used to radiograph a CDMAM(version 3.4) phantom (Artinis Medical Systems, The Netherlands) for image threshold contrast evaluation. After that, different breast thicknesses were simulated with polymethylmethacrylate layers and various sets of exposure parameters were used in order to determine optimal radiographic parameters. For each simulated breast thickness, optimal beam quality was defined as giving a target CNR to reach the threshold contrast of CDMAM images for acceptable MGD. These results were used for adjustments in the automatic exposure control (AEC) by the maintenance team. Using optimized exposure parameters, clinical images of 63 patients (group 2) were evaluated as described above. Threshold contrast, CNR and MGD for such exposure parameters were also determined. Results showed that the proposed optimization method was effective for all breast thicknesses studied in phantoms. The best result was found for breasts of 75 mm. While in group 1 there was no detection of the 0.1 mm critical diameter detail with threshold contrast below 23%, after the optimization, detection occurred in 47.6% of the images. There was also an average MGD reduction of 7.5%. The clinical image quality criteria were attended in 91.7% for all breast thicknesses evaluated in both patient groups. Finally, this study also concluded that the use of the AEC of the x-ray unit based on the constant dose to the detector may bring some difficulties to CR systems to operate under optimal conditions. More studies must be performed, so that the compatibility between systems and optimization methodologies can be evaluated, as well as this optimization method. Most methods are developed for phantoms, so comparative studies including clinical images must be developed.
引用
收藏
页码:6565 / 6583
页数:19
相关论文
共 50 条
  • [1] A Mammography Phantom to Measure Mean Glandular Dose and Image Quality
    Lopez-Pineda, E.
    Ruiz-Trejo, C.
    Brandan, M. E.
    [J]. MEDICAL PHYSICS, 2014, 41 (06)
  • [2] Optimization of image quality and average glandular dose in CR mammography
    Satoh, Kimihiko
    Kuwabara, Takao
    Yasuda, Hiroaki
    Arakawa, Satoshi
    [J]. MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [3] Historical trends in image quality and mean glandular dose in digital mammography
    Mackenzie, Alistair
    Khan, Riyad
    Young, Kenneth C.
    [J]. 15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
  • [4] Mammography equipment performance, image quality and mean glandular dose in Malta
    Borg, M.
    Badr, I.
    Royle, G. J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 156 (02) : 168 - 183
  • [5] Clinical image quality assessment and mean glandular dose for full field digital mammography
    Nassar, Joyce
    Rizk, Chadia
    Fares, Georges
    Tohme, Carla
    Braidy, Chady
    Farah, Jad
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (01)
  • [6] Average glandular dose and phantom image quality in mammography
    Oliveira, M.
    Noguelra, M. S.
    Guedes, E.
    Andrade, M. C.
    Peixoto, J. E.
    Joana, G. S.
    Castro, J. G.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 574 - 577
  • [7] ASSESSMENT OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY
    FAULKNER, K
    LAW, J
    ROBSON, KJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (812): : 877 - 881
  • [8] Image Quality and Mean Glandular Dose in Mexican Mammography Services Measured with an Object-Insert/TL Mammography Kit
    Lopez-Pineda, Eduardo
    Ruiz-Trejo, Cesar
    Brandan, Maria-Ester
    [J]. MEDICAL PHYSICS: Fourteenth Mexican Symposium on Medical Physics, 2016, 1747
  • [9] Optimal beam quality selection based on contrast-to-noise ratio and mean glandular dose in digital mammography
    Aminah, M.
    Ng, K. H.
    Abdullah, B. J. J.
    Jamal, N.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2010, 33 (04) : 329 - 334
  • [10] Optimal beam quality selection based on contrast-to-noise ratio and mean glandular dose in digital mammography
    M. Aminah
    K. H. Ng
    B. J. J. Abdullah
    N. Jamal
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2010, 33 : 329 - 334