Image quality, threshold contrast and mean glandular dose in CR mammography

被引:3
|
作者
Jakubiak, R. R. [1 ]
Gamba, H. R. [2 ]
Neves, E. B. [3 ]
Peixoto, J. E. [4 ]
机构
[1] Univ Tecnol Fed Parana, Acad Dept Phys, BR-80230010 Curitiba, PR, Brazil
[2] Univ Tecnol Fed Parana, Grad Program Biomed Engn, BR-80230010 Curitiba, PR, Brazil
[3] Univ Tecnol Fed Parana, Grad Sch Elect Engn & Comp Sci CPGEI, BR-80230010 Curitiba, PR, Brazil
[4] Natl Canc Inst, BR-20231092 Rio De Janeiro, RJ, Brazil
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 18期
关键词
FIELD DIGITAL MAMMOGRAPHY; EUROPEAN PROTOCOL; MONTE-CARLO; OPTIMIZATION; PERFORMANCE; SYSTEMS;
D O I
10.1088/0031-9155/58/18/6565
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In many countries, computed radiography (CR) systems represent the majority of equipment used in digital mammography. This study presents a method for optimizing image quality and dose in CR mammography of patients with breast thicknesses between 45 and 75 mm. Initially, clinical images of 67 patients (group 1) were analyzed by three experienced radiologists, reporting about anatomical structures, noise and contrast in low and high pixel value areas, and image sharpness and contrast. Exposure parameters (kV, mAs and target/filter combination) used in the examinations of these patients were reproduced to determine the contrast-to-noise ratio (CNR) and mean glandular dose (MGD). The parameters were also used to radiograph a CDMAM(version 3.4) phantom (Artinis Medical Systems, The Netherlands) for image threshold contrast evaluation. After that, different breast thicknesses were simulated with polymethylmethacrylate layers and various sets of exposure parameters were used in order to determine optimal radiographic parameters. For each simulated breast thickness, optimal beam quality was defined as giving a target CNR to reach the threshold contrast of CDMAM images for acceptable MGD. These results were used for adjustments in the automatic exposure control (AEC) by the maintenance team. Using optimized exposure parameters, clinical images of 63 patients (group 2) were evaluated as described above. Threshold contrast, CNR and MGD for such exposure parameters were also determined. Results showed that the proposed optimization method was effective for all breast thicknesses studied in phantoms. The best result was found for breasts of 75 mm. While in group 1 there was no detection of the 0.1 mm critical diameter detail with threshold contrast below 23%, after the optimization, detection occurred in 47.6% of the images. There was also an average MGD reduction of 7.5%. The clinical image quality criteria were attended in 91.7% for all breast thicknesses evaluated in both patient groups. Finally, this study also concluded that the use of the AEC of the x-ray unit based on the constant dose to the detector may bring some difficulties to CR systems to operate under optimal conditions. More studies must be performed, so that the compatibility between systems and optimization methodologies can be evaluated, as well as this optimization method. Most methods are developed for phantoms, so comparative studies including clinical images must be developed.
引用
收藏
页码:6565 / 6583
页数:19
相关论文
共 50 条
  • [21] ASSESSING IMAGE QUALITY AND ESTIMATING THE GLANDULAR AVERAGE DOSE IN SOME MAMMOGRAPHY LABS
    Mossang, D.
    Dadulescu, E.
    [J]. ROMANIAN REPORTS IN PHYSICS, 2011, 63 (01) : 95 - 106
  • [22] Mean glandular dose in digital mammography: A dose calculation method comparison
    Suleiman M.E.
    Brennan P.C.
    McEntee M.F.
    [J]. Suleiman, Moayyad E. (msul1801@uni.sydney.edu.au), 1600, SPIE (04):
  • [23] Estimating mean glandular dose using proprietary mammography phantoms
    Radiation Health Section, Hlth. Dept. W. Australia, Locked B., Nedlands, WA 6009, Australia
    [J]. Appl. Radiat. Isot., 1 (205-213):
  • [24] Estimating mean glandular dose using proprietary mammography phantoms
    Hartley, LD
    Cobb, BJ
    Hutchinson, DE
    [J]. APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) : 205 - 213
  • [25] Evaluation of Image Quality and Average Glandular Dose Using the Resolution Testing Phantom in Mammography
    Chao, Y. T.
    Chen, C. Y.
    Dong, S. L.
    [J]. 1ST GLOBAL CONFERENCE ON BIOMEDICAL ENGINEERING & 9TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 47 : 144 - 146
  • [26] INFLUENCE OF ANODE AND FILTER MATERIAL ON IMAGE QUALITY AND GLANDULAR DOSE FOR SCREEN FILM MAMMOGRAPHY
    DESPONDS, L
    DEPEURSINGE, C
    GRECESCU, M
    HESSLER, C
    SAMIRI, A
    VALLEY, JF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (09): : 1165 - 1182
  • [27] Effect of target/filter combination on the mean glandular dose and contrast-detail threshold: A phantom study
    Nunes, R. S.
    Batista, W. O.
    [J]. RADIOGRAPHY, 2021, 27 (02) : 272 - 278
  • [28] Determination of Mean Glandular Dose in Mammography Procedure at Two Hospitals in Kumasi
    Nsiah-Akoto, Irene
    Andam, Aba Bentil
    Addison, Eric C. K.
    [J]. WOMEN IN PHYSICS, 2013, 1517 : 201 - 201
  • [29] CONFORMANCE OF MEAN GLANDULAR DOSE FROM PHANTOM AND PATIENT DATA IN MAMMOGRAPHY
    Kelaranta, A.
    Toroi, P.
    Timonen, M.
    Komssi, S.
    Kortesniemi, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 164 (03) : 342 - 353
  • [30] PORTUGUESE STUDY OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY AND COMPARISON WITH EUROPEAN REFERENCES
    dos Reis, Claudia Sa
    Fartaria, Mario Joao
    Garcia Alves, Joao H.
    Pascoal, Ana
    [J]. RADIATION PROTECTION DOSIMETRY, 2018, 179 (04) : 391 - 399