Optimization of image quality and average glandular dose in CR mammography

被引:0
|
作者
Satoh, Kimihiko [1 ]
Kuwabara, Takao [1 ]
Yasuda, Hiroaki [1 ]
Arakawa, Satoshi [1 ]
机构
[1] Fuji Film Corp, Med Syst Dev Ctr, 798,Miyanodai,Kaisei-Machi, Kanagawa 258, Japan
关键词
digital mammography; computed radiography; quality control; image quality; automatic exposure control; contrast-to-noise ratio; European reference organization for quality assured breast screening and diagnostic services (EUREF);
D O I
10.1117/12.706781
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The concept proposed by EUREF which determines the AGD between upper limit corresponding to the acceptable level of AGD and lower limit corresponding to lower limit of image quality was applied to CR Mammography, and the resulting object thickness tracking and tube voltage tracking were determined. EUREF specifies threshold contrast visibility for a 5 cm of PMMA. In accordance with this definition, the lower limit of CNR for a 5 cm of PMMA was determined by measuring the CNR at the lower limit of AGD where threshold contrast visibility was just acceptable. The obtained lower limit of CNR was then multiplied by the object thickness compensation factor to estimate the lower limits of CNR for all object thicknesses. AGDs are now determined for each lower limit of CNR to obtain thickness tracking and tube voltage tracking characteristics. Although the limited range of examined target/filter combinations. and tube voltages should be taken into consideration, these tracking characteristics for constant CNR differ in their profiles from those of a screen-film system with AEC for providing constant optical density. Among our findings, we found that a lower AGD is achieved while maintaining CNR for a thick object when using a combination of target/filter and tube voltage that generates higher X-ray energy compared to the combination given in the EUREF's typical spectra per PMMA thickness (Mo/Rh 32 kV at PMMA 4cm and Rh/Rh 28 kV at PMMA 6 cm, for example). We also found that the thickness tracking characteristics for constant S value behaves similarly to constant CNR under certain conditions of the target/filter combination and tube voltage.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Average glandular dose and phantom image quality in mammography
    Oliveira, M.
    Noguelra, M. S.
    Guedes, E.
    Andrade, M. C.
    Peixoto, J. E.
    Joana, G. S.
    Castro, J. G.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 574 - 577
  • [2] Effect of filter on average glandular dose and image quality in digital mammography
    Songsaeng, C.
    Krisanachinda, A.
    Theerakul, K.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [3] Image quality, threshold contrast and mean glandular dose in CR mammography
    Jakubiak, R. R.
    Gamba, H. R.
    Neves, E. B.
    Peixoto, J. E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (18): : 6565 - 6583
  • [4] ASSESSING IMAGE QUALITY AND ESTIMATING THE GLANDULAR AVERAGE DOSE IN SOME MAMMOGRAPHY LABS
    Mossang, D.
    Dadulescu, E.
    [J]. ROMANIAN REPORTS IN PHYSICS, 2011, 63 (01) : 95 - 106
  • [5] Evaluation of Image Quality and Average Glandular Dose Using the Resolution Testing Phantom in Mammography
    Chao, Y. T.
    Chen, C. Y.
    Dong, S. L.
    [J]. 1ST GLOBAL CONFERENCE ON BIOMEDICAL ENGINEERING & 9TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 47 : 144 - 146
  • [6] Optimizing the anode-filter combination in the sense of image quality and average glandular dose in digital mammography
    Varjonen, Mari
    Stroemmer, Pekka
    [J]. MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [7] Optimizing the target-filter combination in digital mammography in the sense of image quality and average glandular dose
    Varjonen, Mari
    Strommer, Pekka
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 570 - 576
  • [8] Patient Specific Average Glandular Dose in Mammography
    Patel, V.
    Highnam, R.
    Tromans, C.
    Pizzutiello, R. J.
    Destounis, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [9] NORMALIZED AVERAGE GLANDULAR DOSE IN MAGNIFICATION MAMMOGRAPHY
    LIU, B
    GOODSITT, M
    CHAN, HP
    [J]. RADIOLOGY, 1995, 197 (01) : 27 - 32
  • [10] Method for the evaluation of a average glandular dose in mammography
    Okunade, AA
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1153 - 1164