Clinical image quality assessment and mean glandular dose for full field digital mammography

被引:0
|
作者
Nassar, Joyce [1 ]
Rizk, Chadia [1 ,2 ]
Fares, Georges [1 ]
Tohme, Carla [3 ]
Braidy, Chady [3 ]
Farah, Jad [4 ]
机构
[1] St Joseph Univ, Fac Sci, POB 11-514, Beirut 11072050, Lebanon
[2] Natl Council Sci Res, Lebanese Atom Energy Commiss, Beirut 118281, Lebanon
[3] Hotel Dieu France Hosp, Radiol Dept, POB 166830, Beirut, Lebanon
[4] Vis RT Ltd, Dove House,Arcadia Ave, London N3 2JU, England
关键词
mammography; image quality; system performance; positioning performance; average glandular dose; DIAGNOSTIC REFERENCE LEVELS; EQUIPMENT PERFORMANCE; CRITERIA; RELIABILITY; VALIDITY; FAILURE;
D O I
10.1088/1361-6498/ad1cd4
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study aims to assess the image quality (IQ) of 12 mammographic units and to identify units with potential optimisation needs. Data for 350 mammography examinations meeting inclusion criteria were collected retrospectively from April 2021 to April 2022. They were categorised based on the medical reports into 10 normal cases, 10 cases displaying calcifications and 10 cases presenting lesions. Two radiologists assessed the IQ of 1400 mammograms, evaluating system performance per Boita et al's study and positioning performance following European guidelines. To measure agreement between the two radiologists, the Cohen's Kappa coefficient (kappa) was computed, quantifying the excess of agreement beyond chance. The visual grading analysis score (VGAS) was computed to compare system and positioning performance assessments across different categories and facilities. Median average glandular dose (AGD) values for cranio caudal and medio lateral oblique views were calculated for each category and facility and compared to the national diagnostic reference levels. The health facilities were categorised by considering both IQ VGAS and AGD levels. Inter-rater agreement between radiologists ranged from poor (kappa < 0.20) to moderate (0.41 < kappa < 0.60), likely influenced by inherent biases and distinct IQ expectations. 50% of the facilities were classified as needing corrective actions for their system performance as they had IQ or high AGD that could increase recall rate and radiation risk and 50% of the health facilities exhibited insufficient positioning performance that could mask tumour masses and microcalcifications. The study's findings emphasise the importance of implementing quality assurance programs to ensure optimal IQ for accurate diagnoses while adhering to radiation exposure guidelines. Additionally, comprehensive training for technologists is essential to address positioning challenges. These initiatives collectively aim to enhance the overall quality of breast imaging services, contributing to improved patient care.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Historical trends in image quality and mean glandular dose in digital mammography
    Mackenzie, Alistair
    Khan, Riyad
    Young, Kenneth C.
    [J]. 15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
  • [2] Average glandular dose with full-field digital mammography - Clinical results
    Hermann, KP
    Kehbel, S
    Obenauer, S
    Fischer, U
    Grabbe, E
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2003, : 502 - 504
  • [3] Comparison of Mean Glandular Dose between Full-Field Digital Mammography and Digital Breast Tomosynthesis
    Teoh, Kar Choon
    Manan, Hanani Abdul
    Norsuddin, Norhashimah Mohd
    Rizuana, Iqbal Hussain
    [J]. HEALTHCARE, 2021, 9 (12)
  • [4] A Mammography Phantom to Measure Mean Glandular Dose and Image Quality
    Lopez-Pineda, E.
    Ruiz-Trejo, C.
    Brandan, M. E.
    [J]. MEDICAL PHYSICS, 2014, 41 (06)
  • [5] EVALUATION OF MEAN GLANDULAR DOSE IN A FULL-FIELD DIGITAL MAMMOGRAPHY UNIT IN TABRIZ, IRAN
    Riabi, Hamed Alizadeh
    Mehnati, Parinaz
    Mesbahi, Asghar
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 142 (2-4) : 222 - 227
  • [6] ASSESSMENT OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY
    FAULKNER, K
    LAW, J
    ROBSON, KJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (812): : 877 - 881
  • [7] Image quality, threshold contrast and mean glandular dose in CR mammography
    Jakubiak, R. R.
    Gamba, H. R.
    Neves, E. B.
    Peixoto, J. E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (18): : 6565 - 6583
  • [8] Mammography equipment performance, image quality and mean glandular dose in Malta
    Borg, M.
    Badr, I.
    Royle, G. J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 156 (02) : 168 - 183
  • [9] Average glandular dose with amorphous silicon full-field digital mammography - Clinical results
    Hermann, KP
    Obenauer, S
    Marten, K
    Kehbel, S
    Fischer, U
    Grabbe, E
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (06): : 696 - 699
  • [10] Determination of dose delivery accuracy and image quality in full - Field digital mammography
    Sosu, E. K.
    Boadu, M.
    Mensah, S. Y.
    [J]. JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2018, 11 (03) : 232 - 236