Independent verification of gantry angle for pre-treatment VMAT QA using EPID

被引:17
|
作者
Adamson, Justus [1 ]
Wu, Qiuwen [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27707 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 20期
关键词
MODULATED ARC THERAPY; DOSIMETRIC VERIFICATION; QUALITY-ASSURANCE; IMRT; RADIATION; DELIVERY; IMAT; TOOL;
D O I
10.1088/0031-9155/57/20/6587
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We propose a method to incorporate independent verification of gantry angle for electronic portal imaging device (EPID)-based pre-treatment quality assurance (QA) of clinical volumetric modulated arc therapy (VMAT) plans. Gantry angle is measured using projections in the EPID of a custom phantom placed on the couch and the treatment plan is modified so as to be incident on the phantom with a portion of the beam that is collimated in the clinical plan. For our implementation, collimator and couch angles were set to zero and the inferior jaw and two most inferior multi-leaf collimator pairs were opened for the entire QA delivery. A phantom containing five gold coils was used to measure the gantry rotation through which each portal image was acquired. We performed the EPID QA for ten clinical plans and evaluated accuracy of gantry angle measurement, scatter incident on the imager due to the phantom, inter-image pixel linearity and inter-and intra-image noise. The gantry angle could be measured to within 0.0 +/- 0.3 degrees for static gantry and 0.2 +/- 0.2 degrees for arc acquisitions. Scatter due to the presence of the phantom was negligible. The procedure was shown to be feasible and adds gantry angle to the treatment planning parameters that can be verified by EPID-based pre-treatment VMAT QA.
引用
下载
收藏
页码:6587 / 6600
页数:14
相关论文
共 50 条
  • [1] Independent verification of gantry angle for pre-treatment VMAT QA using EPID
    Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27707, United States
    Phys. Med. Biol., 20 (6587-6600):
  • [2] Confidence in VMAT EPID pre-treatment QA
    Monti, A. F.
    Trombetta, L.
    Brambilla, M. G.
    Carbonini, C.
    Torresin, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S170 - S170
  • [3] Linking VMAT pre-treatment QA and patient treatment QA by EPID dosimetry
    Sassowsky, M.
    Manser, P.
    Fix, M. K.
    Frauchiger, D.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (08) : 754 - 754
  • [4] Patient Specific Pre-treatment QA Verification using an EPID Approach
    Quino, Luis A. Vazquez
    Chen, Xiaoming
    Fitzpatrick, Mathew
    Shi, Chengyu
    Stathakis, Sotirios
    Gutierrez, Alonso
    Esquivel, Carlos
    Mavroidis, Panayiotis
    Alkhatib, Hassan
    Papanikolaou, Niko
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2014, 13 (01) : 1 - 10
  • [5] Use of a commercial EPID dosimetry system in VMAT pre-treatment QA
    Parisotto, M.
    Carbonini, C.
    Brambilla, M. G.
    Monti, A. F.
    Ferrari, M. B.
    Zanni, D.
    Torresin, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S164 - S165
  • [6] Gantry-angle resolved VMAT pretreatment verification using EPID image prediction
    Woodruff, Henry C.
    Fuangrod, Todsaporn
    Rowshanfarzad, Pejman
    McCurdy, Boyd M. C.
    Greer, Peter B.
    MEDICAL PHYSICS, 2013, 40 (08)
  • [7] Can An EPID Be Commissioned for Multi-Met SRS Pre-Treatment VMAT QA?
    Raxter, D.
    Litzenberg, D.
    Moran, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E606 - E606
  • [8] EPID for QA and pre-treatment verification of electron beams in absolute dose using EpiDream method
    Sors, A.
    Dudouet, P.
    Franck, D.
    Mace, M.
    Boucarut, P.
    Boutry, C.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S472 - S472
  • [9] Is Portal Dosimetry a Truly Absolute QA Tool for VMAT Pre-Treatment Verification?
    Ai, H.
    Huang, K.
    Lu, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E280 - E280
  • [10] EPID sensitivity to delivery errors for pre-treatment verification of lung SBRT VMAT plans
    Alharthi, Thahabah
    Arumugam, Sankar
    Vial, Phil
    Holloway, Lois
    Thwaites, David
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 59 : 37 - 46