EPID sensitivity to delivery errors for pre-treatment verification of lung SBRT VMAT plans

被引:8
|
作者
Alharthi, Thahabah [1 ,2 ,3 ,4 ]
Arumugam, Sankar [3 ,5 ]
Vial, Phil [3 ,4 ,5 ]
Holloway, Lois [1 ,3 ,4 ,5 ,6 ]
Thwaites, David [1 ]
机构
[1] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW, Australia
[2] Taif Univ, Sch Med, At Taif, Saudi Arabia
[3] Liverpool & Macarthur Canc Therapy Ctr, Liverpool, Merseyside, England
[4] Ingham Inst Appl Med Res, Sydney, NSW, Australia
[5] Univ New South Wales, South Western Sydney Clin Sch, Sydney, NSW, Australia
[6] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
关键词
Lung SBRT VMAT; Delivery error; EPID; MODULATED ARC THERAPY; BODY RADIATION-THERAPY; ION-CHAMBER ARRAY; QUALITY-ASSURANCE; DOSE-RESPONSE; RADIOTHERAPY; IMRT; DOSIMETRY; SINGLE; DEVICE;
D O I
10.1016/j.ejmp.2019.02.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To study the sensitivity of an Electronic Portal Imaging Device (EPID) in detecting delivery errors for VMAT lung stereotactic body radiotherapy (SBRT) using the Collapsed Arc method. Methods: Baseline VMAT plans and plans with errors intentionally introduced were generated for 15 lung SBRT patients. Three types of errors were introduced by modifying collimator angles and multi-leaf collimator (MLC) field sizes (MLCFS) and MLC shifts by +/- 5, +/- 2, and +/- 1 degrees or millimeters. A total of 103 plans were measured with EPID on an Elekta Synergy Linear Accelerator (Agility MLC) and compared to both the original treatment planning system (TPS) Collapsed Arc dose matrix and the no-error plan baseline EPID measurements. Gamma analysis was performed using the OmniPro-I'mRT (IBA Dosimetry) software and gamma criteria of 1%/1 mm, 2%/1 mm, 2%/2 mm, and 3%/3. Results: When the error-introduced EPID measured dose matrices were compared to the TPS matrices, the majority of simulated errors were detected with gamma tolerance of 2%/1 mm and 1%/1 mm. When the error-introduced EPID measured dose matrices were compared to the baseline EPID measurements, all the MLCFS and MLC shift errors, and +/- 5 degrees collimator errors were detected using 2%/1 mm and 1%/1 mm gamma criteria. Conclusion: This work demonstrates the feasibility and effectiveness of the collapsed arc technique and EPID for pre-treatment verification of lung SBRT VMAT plans. The EPID was able to detect the majority of MLC and the larger collimator errors with sensitivity to errors depending on the gamma tolerances.
引用
收藏
页码:37 / 46
页数:10
相关论文
共 50 条
  • [1] Sensitivity of ASi EPID to Delivery Errors for Pre Treatment Verification of Lung SBRT VMAT Plans
    Alharthi, T.
    Arumugam, S.
    Vial, P.
    Holloway, L.
    Thwaites, D.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E324 - E324
  • [2] Pre-treatment verification of lung SBRT VMAT plans with delivery errors: Toward a better understanding of the gamma index analysis
    Alharthi, Thahabah
    Pogson, Elise M.
    Arumugam, Sankar
    Holloway, Lois
    Thwaites, David
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 49 : 119 - 128
  • [3] Intrinsic detector sensitivity analysis as a tool to characterize ArcCHECK and EPID sensitivity to variations in delivery for lung SBRT VMAT plans
    Alharthi, Thahabah
    Vial, Phil
    Holloway, Lois
    Thwaites, David
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (06): : 229 - 240
  • [4] Sensitivity of an Elekta iView GT a-Si EPID model to delivery errors for pre-treatment verification of IMRT fields
    Sri Herwiningsih
    Peta Hanlon
    Andrew Fielding
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 763 - 770
  • [5] Sensitivity of an Elekta iView GT a-Si EPID model to delivery errors for pre-treatment verification of IMRT fields
    Herwiningsih, Sri
    Hanlon, Peta
    Fielding, Andrew
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (04) : 763 - 770
  • [6] Comparison of multi-institutional pre-treatment verification for VMAT of nasopharynx with delivery errors
    Pogson, E. M.
    Arumugam, S.
    Hansen, C. R.
    Currie, M.
    Blake, S. J.
    Roberts, N.
    Carolan, M.
    Vial, P.
    Alharthi, T.
    Holloway, L.
    Thwaites, D. I.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 53 : 25 - 31
  • [7] Independent verification of gantry angle for pre-treatment VMAT QA using EPID
    Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27707, United States
    [J]. Phys. Med. Biol, 20 (6587-6600):
  • [8] Independent verification of gantry angle for pre-treatment VMAT QA using EPID
    Adamson, Justus
    Wu, Qiuwen
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20): : 6587 - 6600
  • [9] Confidence in VMAT EPID pre-treatment QA
    Monti, A. F.
    Trombetta, L.
    Brambilla, M. G.
    Carbonini, C.
    Torresin, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S170 - S170
  • [10] An in vivo dose verification method for SBRT-VMAT delivery using the EPID
    McCowan, P. M.
    Van Uytven, E.
    Van Beek, T.
    Asuni, G.
    McCurdy, B. M. C.
    [J]. MEDICAL PHYSICS, 2015, 42 (12) : 6955 - 6963