Depth-profile analysis of thermoelectric layers on Si wafers by pulsed r.f. glow discharge time-of-flight mass spectrometry

被引:14
|
作者
Reinsberg, K. -G. [1 ]
Schumacher, C. [2 ]
Tempez, A. [3 ]
Nielsch, K. [2 ]
Broekaert, J. A. C. [1 ]
机构
[1] Univ Hamburg, Inst Inorgan & Appl Chem, D-20146 Hamburg, Germany
[2] Univ Hamburg, Inst Appl Phys, D-20355 Hamburg, Germany
[3] HORIBA Jobin Yvon, F-91160 Longjumeau, France
关键词
Thermoelectrics; Pulsed r.f. glow discharge; Time-of-flight mass spectrometry; Depthprofiling; BISMUTH TELLURIDE; ELECTRODEPOSITION; OPTIMIZATION; FILMS;
D O I
10.1016/j.sab.2012.06.005
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this work the depth-profile analysis of thermoelectric layers deposited on Au and Cr covered Si wafers with the aid of pulsed radiofrequency glow discharge time-of-flight mass spectrometry (pulsed RF-GD-TOFMS also called plasma profiling TOFMS (PP-TOFMS (TM))) is described. For thermoelectric materials the depth resolutions obtained with both PP-TOFMS and secondary ion mass spectrometry (SIMS) are shown to be well comparable and in the order of the roughness of the corresponding layers (between 20 and 3700 nm). With both methods a direct solid analysis without any preparation steps is possible. In addition, the analysis of the samples with PP-TOFMS proved to be faster by a factor of 26 compared to SIMS, as sputtering rates were found to be 80 nm s(-1) and 3 nm s(-1), respectively. For the analyzed samples the results of PP-TOFMS and SIMS show that a homogeneous deposition was obtained. Quantitative results for all samples could also be obtained directly by PP-TOFMS when the stoichiometry of one sample was determined beforehand for instance by inductively coupled plasma optical emission spectrometry (ICP-OES) and scanning electron microscopy energy dispersive X-ray fluorescence spectrometry (SEM-EDX). For Bi2Te3 the standard deviation for the main component concentrations within one sample then is found to be between 1.1% and 1.9% and it is 3.6% from sample to sample. For Sb2Te3 the values within one sample are from 1.7% to 42% and from sample to sample 5.3%, respectively. (C) 2012 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:175 / 180
页数:6
相关论文
共 50 条
  • [21] Improving pulsed radiofrequency glow discharge for time-of-flight mass spectrometry simultaneous elemental and molecular analysis
    Auristela Solà-Vázquez
    Beatriz Fernández
    José M. Costa-Fernández
    Rosario Pereiro
    Alfredo Sanz-Medel
    Analytical and Bioanalytical Chemistry, 2014, 406 : 7431 - 7443
  • [22] Direct isotope analysis of Chernobyl microparticles using time-of-flight mass spectrometry with pulsed glow discharge
    Ganeev, Aleksander
    Gubal, Anna
    Korotetski, Boris
    Bogdanova, Oksana
    Burakov, Boris
    Titova, Anna
    Solovyev, Nikolay
    Ivanenko, Natalya
    Drobyshev, Evgenii
    Iakovleva, Evgenia
    Sillanpaa, Mika
    MICROCHEMICAL JOURNAL, 2017, 132 : 286 - 292
  • [23] Gas chromatography coupled to tunable pulsed glow discharge time-of-flight mass spectrometry for environmental analysis
    Sola-Vazquez, Auristela
    Lara-Gonzalo, Azucena
    Costa-Fernandez, Jose M.
    Pereiro, Rosario
    Sanz-Medel, Alfredo
    ANALYST, 2010, 135 (05) : 987 - 993
  • [24] Pulsed radiofrequency glow discharge time-of-flight mass spectrometry for molecular depth profiling of polymer-based films
    Tuccitto, Nunzio
    Lobo, Lara
    Tempez, Agnes
    Delfanti, Ivan
    Chapon, Patrick
    Canulescu, Stela
    Bordel, Nerea
    Michler, Johann
    Licciardello, Antonino
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (05) : 549 - 556
  • [25] Pulsed r.f.-glow-discharge time-of-flight mass spectrometry for fast surface and interface analysis of conductive and non-conductive materials
    Hohl, M
    Kanzari, A
    Michler, J
    Nelis, T
    Fuhrer, K
    Gonin, M
    SURFACE AND INTERFACE ANALYSIS, 2006, 38 (04) : 292 - 295
  • [26] Depth Profile Analysis of Amorphous Silicon Thin Film Solar Cells by Pulsed Radiofrequency Glow Discharge Time of Flight Mass Spectrometry
    Alvarez-Toral, Aitor
    Sanchez, Pascal
    Menendez, Armando
    Pereiro, Rosario
    Sanz-Medel, Alfredo
    Fernandez, Beatriz
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2015, 26 (02) : 305 - 314
  • [27] Direct determination of uranium and thorium in minerals by time-of-flight mass spectrometry with pulsed glow discharge
    Ganeev, Aleksandr
    Bogdanova, Oksana
    Ivanov, Ilia
    Burakov, Boris
    Agafonova, Natalia
    Korotetski, Boris
    Gubal, Anna
    Solovyev, Nikolay
    Iakovleva, Evgenia
    Sillanpaa, Mika
    RSC ADVANCES, 2015, 5 (99) : 80901 - 80910
  • [28] Pulsed radiofrequency glow discharge time of flight mass spectrometry for coated glass analysis
    20151900828891
    Fernández, Beatriz (fernandezbeatriz@uniovi.es), 1600, Royal Society of Chemistry (30):
  • [29] Polymer screening by radiofrequency glow discharge time-of-flight mass spectrometry
    Lobo, L.
    Tuccitto, N.
    Bordel, N.
    Pereiro, R.
    Pisonero, J.
    Licciardello, A.
    Tempez, A.
    Chapon, P.
    Sanz-Medel, A.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 396 (08) : 2863 - 2869
  • [30] Polymer screening by radiofrequency glow discharge time-of-flight mass spectrometry
    L. Lobo
    N. Tuccitto
    N. Bordel
    R. Pereiro
    J. Pisonero
    A. Licciardello
    A. Tempez
    P. Chapon
    A. Sanz-Medel
    Analytical and Bioanalytical Chemistry, 2010, 396 : 2863 - 2869