Depth-profile analysis of thermoelectric layers on Si wafers by pulsed r.f. glow discharge time-of-flight mass spectrometry

被引:14
|
作者
Reinsberg, K. -G. [1 ]
Schumacher, C. [2 ]
Tempez, A. [3 ]
Nielsch, K. [2 ]
Broekaert, J. A. C. [1 ]
机构
[1] Univ Hamburg, Inst Inorgan & Appl Chem, D-20146 Hamburg, Germany
[2] Univ Hamburg, Inst Appl Phys, D-20355 Hamburg, Germany
[3] HORIBA Jobin Yvon, F-91160 Longjumeau, France
关键词
Thermoelectrics; Pulsed r.f. glow discharge; Time-of-flight mass spectrometry; Depthprofiling; BISMUTH TELLURIDE; ELECTRODEPOSITION; OPTIMIZATION; FILMS;
D O I
10.1016/j.sab.2012.06.005
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this work the depth-profile analysis of thermoelectric layers deposited on Au and Cr covered Si wafers with the aid of pulsed radiofrequency glow discharge time-of-flight mass spectrometry (pulsed RF-GD-TOFMS also called plasma profiling TOFMS (PP-TOFMS (TM))) is described. For thermoelectric materials the depth resolutions obtained with both PP-TOFMS and secondary ion mass spectrometry (SIMS) are shown to be well comparable and in the order of the roughness of the corresponding layers (between 20 and 3700 nm). With both methods a direct solid analysis without any preparation steps is possible. In addition, the analysis of the samples with PP-TOFMS proved to be faster by a factor of 26 compared to SIMS, as sputtering rates were found to be 80 nm s(-1) and 3 nm s(-1), respectively. For the analyzed samples the results of PP-TOFMS and SIMS show that a homogeneous deposition was obtained. Quantitative results for all samples could also be obtained directly by PP-TOFMS when the stoichiometry of one sample was determined beforehand for instance by inductively coupled plasma optical emission spectrometry (ICP-OES) and scanning electron microscopy energy dispersive X-ray fluorescence spectrometry (SEM-EDX). For Bi2Te3 the standard deviation for the main component concentrations within one sample then is found to be between 1.1% and 1.9% and it is 3.6% from sample to sample. For Sb2Te3 the values within one sample are from 1.7% to 42% and from sample to sample 5.3%, respectively. (C) 2012 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:175 / 180
页数:6
相关论文
共 50 条
  • [1] Critical evaluation of the potential of radiofrequency pulsed glow discharge–time-of-flight mass spectrometry for depth-profile analysis of innovative materials
    Jorge Pisonero
    Nerea Bordel
    Claudia Gonzalez de Vega
    Beatriz Fernández
    Rosario Pereiro
    Alfredo Sanz-Medel
    Analytical and Bioanalytical Chemistry, 2013, 405 : 5655 - 5662
  • [2] Pulsed glow discharge time-of-flight mass spectrometry
    Harrison, WW
    Hang, W
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) : 835 - 840
  • [3] Critical evaluation of the potential of radiofrequency pulsed glow discharge-time-of-flight mass spectrometry for depth-profile analysis of innovative materials
    Pisonero, Jorge
    Bordel, Nerea
    Gonzalez de Vega, Claudia
    Fernandez, Beatriz
    Pereiro, Rosario
    Sanz-Medel, Alfredo
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2013, 405 (17) : 5655 - 5662
  • [4] Quantitative depth profile analysis of boron implanted silicon by pulsed radiofrequency glow discharge time-of-flight mass spectrometry
    Pisonero, J.
    Lobo, L.
    Bordel, N.
    Tempez, A.
    Bensaoula, A.
    Badi, N.
    Sanz-Medel, A.
    SOLAR ENERGY MATERIALS AND SOLAR CELLS, 2010, 94 (08) : 1352 - 1357
  • [5] Time-of-flight mass spectrometry with a pulsed glow discharge ionization source
    Steiner, RE
    Lewis, CL
    King, FL
    ANALYTICAL CHEMISTRY, 1997, 69 (09) : 1715 - 1721
  • [6] Pulsed radiofrequency glow discharge time-of-flight mass spectrometry: Depth profile analysis of multilayers on conductive and non-conductive substrates
    Lobo, Lara
    Fernandez, Beatriz
    Aranaz, Marta
    Fernandez Lorenzo, Andres
    Ignacio Martin-Carbajo, Jose
    Pereiro, Rosario
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2020, 168
  • [7] Depth profiling by pulsed glow discharge time-of-flight mass spectrometry with a combined hollow cathode cell
    Gubal, Anna
    Chuchina, Victoria
    Lyalkin, Yegor
    Mikhailovskii, Vladimir
    Yakobson, Viktor
    Solovyev, Nikolay
    Ganeev, Alexander
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (08) : 1587 - 1596
  • [8] Elimination of water interference in pulsed glow discharge time-of-flight mass spectrometry
    A. A. Ganeev
    A. R. Gubal
    S. V. Potapov
    S. E. Pogarev
    S. E. Sholupov
    K. N. Uskov
    I. S. Ivanov
    Journal of Analytical Chemistry, 2013, 68 : 1205 - 1211
  • [9] Elimination of water interference in pulsed glow discharge time-of-flight mass spectrometry
    Ganeev, A. A.
    Gubal, A. R.
    Potapov, S. V.
    Pogarev, S. E.
    Sholupov, S. E.
    Uskov, K. N.
    Ivanov, I. S.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2013, 68 (14) : 1205 - 1211
  • [10] Study of a pulsed glow discharge ion source for time-of-flight mass spectrometry
    Su, YX
    Zhou, Z
    Yang, PY
    Wang, XR
    Huang, BL
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (05) : 633 - 641