A method to derive appropriate exposure parameters from target exposure index and patient thickness in pediatric digital radiography

被引:11
|
作者
Zhang, Menglong [1 ]
Liu, Kai [1 ]
Niu, Xuecai [2 ]
Liu, Xinli [2 ]
机构
[1] Shandong Med Imaging Res Inst, Dept Radiog, Jinan 250021, Shangdong, Peoples R China
[2] Fourth Hosp Jinan, Dept Radiat Oncol, Jinan, Shandong, Peoples R China
关键词
Radiation dose; Pediatrics; Digital radiography; Exposure index; Radiation protection; COMPUTED RADIOGRAPHY; IMAGE QUALITY; X-RAY; OPTIMIZATION; CHILDREN; CHEST;
D O I
10.1007/s00247-012-2555-3
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Little information exists concerning appropriate exposure and measuring overall patient dose in pediatric digital radiography. To establish a convenient method of appropriate exposure from target exposure index (EI) and thickness in pediatric digital radiography and estimate patient entrance-surface dose (ESD) and dose-area product (DAP) associated with chest, abdomen and pelvis radiography. A formula was deduced to calculate appropriate mAs changed with children's weight and height. EI was used to control image quality. With this formula, dose-optimized procedures were carried out. Data were collected from 180 pediatric examinations, including chest, abdomen and pelvis anterior-posterior (AP) projections. The children were divided into the following age bands: newborns (0-28 days), infants (28 days-2 years) and older children (2-7 years). In each age band, ten children were exposed with the calculated appropriate mAs and EI values were kept steady in appropriate range (referred as target group) and ten children were exposed to the factors routinely used in practice (referred to as the routine group). DAP to children was measured with a DAP meter, and ESD was calculated using measured DAP and data from the National Radiological Protection Board. Data were compared between groups. ESD ranges in the target group were 32-202 mu Gy (chest AP), 57-333 mu Gy (abdomen AP) and 52-372 mu Gy (pelvis AP). For every radiographic procedure, chi-square Student's t tests showed a significant difference in average ESD and DAP between the two groups (P < 0.005). Most ESD values from the routine group were two times higher than those from the target group. The study established a convenient method to set appropriate exposure parameters (mAs) to reach a target EI using the child's weight and height in pediatric radiography. By this method, ESD and DAP can be significantly reduced in children.
引用
收藏
页码:568 / 574
页数:7
相关论文
共 50 条
  • [1] A method to derive appropriate exposure parameters from target exposure index and patient thickness in pediatric digital radiography
    Menglong Zhang
    Kai Liu
    Xuecai Niu
    Xinli Liu
    Pediatric Radiology, 2013, 43 : 568 - 574
  • [2] Evaluation of the effects of subject thickness on the exposure index in digital radiography
    Takaki T.
    Takeda K.
    Murakami S.
    Ogawa H.
    Ogawa M.
    Sakamoto M.
    Radiological Physics and Technology, 2016, 9 (1) : 116 - 120
  • [3] A Standardized Exposure Index for Digital Radiography
    Brake, Dean Ann
    RADIOLOGIC TECHNOLOGY, 2016, 87 (05) : 581 - 585
  • [4] Automated detection of changes in patient exposure in digital projection radiography using exposure index from DICOM header metadata
    Kallman, Hans-Erik
    Halsius, Erik
    Folkesson, Mikael
    Larsson, Ylva
    Stenstrom, Mats
    Bath, Magnus
    ACTA ONCOLOGICA, 2011, 50 (06) : 960 - 965
  • [5] Exposure index in digital radiography: initial results of awareness and knowledge from Nigerian digital radiography practices
    Esien-umo, Emmanuel O.
    Erim, Akwa E.
    Chiaghanam, Ndubuisi O.
    Ogbu, Treasure
    Ijever, Andrew W.
    Archibong, Bassey E.
    Osakwe, Chidinma A.
    Ekpo, Ernest U.
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2023, 54 (01) : 58 - 65
  • [6] Establishing Target Exposure Index Values for Computed Radiography
    Murphy, N.
    Tchou, P.
    Belcher, K.
    Scott, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 136 - +
  • [7] The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric population
    Seibert, J. Anthony
    Morin, Richard L.
    PEDIATRIC RADIOLOGY, 2011, 41 (05) : 573 - 581
  • [8] The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric population
    J. Anthony Seibert
    Richard L. Morin
    Pediatric Radiology, 2011, 41 : 573 - 581
  • [9] Method for Determining Partition and Exposure Parameters of Gas Turbine Blades by Digital Radiography
    Li B.
    Li Z.
    Chen L.
    Hsi-An Chiao Tung Ta Hsueh/Journal of Xi'an Jiaotong University, 2019, 53 (10): : 31 - 39
  • [10] Reduction of patient exposure using digital luminescence radiography
    Seifert, H
    Chapot, C
    JOURNAL DE RADIOLOGIE, 1999, 80 (11): : 1555 - 1560