Automated detection of changes in patient exposure in digital projection radiography using exposure index from DICOM header metadata

被引:11
|
作者
Kallman, Hans-Erik [1 ,2 ]
Halsius, Erik [3 ]
Folkesson, Mikael
Larsson, Ylva
Stenstrom, Mats
Bath, Magnus [4 ,5 ]
机构
[1] Uppsala Univ, Dept Radiol Oncol & Radiat Sci, Uppsala, Sweden
[2] Ctr Clin Res Dalarna, Falun, Sweden
[3] Agfa Healthcare Sweden, Kista, Sweden
[4] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, Gothenburg, Sweden
[5] Univ Gothenburg, Inst Clin Sci, Dept Radiat Phys, Gothenburg, Sweden
关键词
TIME-SERIES; COMPUTED RADIOGRAPHY; TESTS;
D O I
10.3109/0284186X.2011.579622
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose. Automated collection of image data from DICOM headers enables monitoring of patient dose and image quality parameters. Manual monitoring is time consuming, owing to the large number of exposure scenarios, thus automated methods for monitoring needs to be investigated. The aim of the present work was to develop and optimise such a method. Material and methods. Exposure index values from digital systems in projection radiography were collected over a period of five years, representing data from 1.2 million projection images. The exposure index values were converted to detector dose and an automated method for detection of sustained level shifts in the resulting detector dose time series was applied using the statistical analysis tool R. The method combined handling of outliers, filtering and estimation of variation in combination with two different statistical rank tests for level shift detection. A set of 304 time series representing central body parts was selected and the level shift detection method was optimised using level shifts identified by ocular evaluation as the gold standard. Results. Two hundred and eighty-one level changes were identified that were deemed in need of further investigation. The majority of these changes were abrupt. The sensitivity and specificity of the optimised and automated detection method concerning the ocular evaluation were 0.870 and 0.997, respectively, for detected abrupt changes. Conclusions. An automated analysis of exposure index values, with the purpose of detecting changes in exposure, can be performed using the R software in combination with a DICOM header metadata repository containing the exposure index values from the images. The routine described has good sensitivity and acceptable specificity for a wide range of central body part projections and can be optimised for more specialised purposes.
引用
收藏
页码:960 / 965
页数:6
相关论文
共 14 条
  • [1] Reduction of patient exposure using digital luminescence radiography
    Seifert, H
    Chapot, C
    JOURNAL DE RADIOLOGIE, 1999, 80 (11): : 1555 - 1560
  • [2] A method to derive appropriate exposure parameters from target exposure index and patient thickness in pediatric digital radiography
    Zhang, Menglong
    Liu, Kai
    Niu, Xuecai
    Liu, Xinli
    PEDIATRIC RADIOLOGY, 2013, 43 (05) : 568 - 574
  • [3] A method to derive appropriate exposure parameters from target exposure index and patient thickness in pediatric digital radiography
    Menglong Zhang
    Kai Liu
    Xuecai Niu
    Xinli Liu
    Pediatric Radiology, 2013, 43 : 568 - 574
  • [4] Evaluation of digital radiography practice using exposure index tracking
    Scott, Alexander W.
    Zhou, Yifang
    Allahverdian, Janet
    Nute, Jessica L.
    Lee, Christina
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (06): : 343 - 355
  • [5] A Landmark Detection Approach Applied to Robust Estimation of the Exposure Index in Digital Radiography
    Irrera, P.
    Bloch, I.
    Delplanque, M.
    IRBM, 2017, 38 (01) : 42 - 55
  • [6] Exposure index in digital radiography: initial results of awareness and knowledge from Nigerian digital radiography practices
    Esien-umo, Emmanuel O.
    Erim, Akwa E.
    Chiaghanam, Ndubuisi O.
    Ogbu, Treasure
    Ijever, Andrew W.
    Archibong, Bassey E.
    Osakwe, Chidinma A.
    Ekpo, Ernest U.
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2023, 54 (01) : 58 - 65
  • [7] Estimating pediatric entrance skin dose from digital radiography examination using DICOM metadata: A quality assurance tool
    Brady, S. L.
    Kaufman, R. A.
    MEDICAL PHYSICS, 2015, 42 (05) : 2489 - 2497
  • [8] Estimating Pediatric Entrance Skin Dose From Digital Radiography Examination Using DICOM Metadata: A Quality Assurance Tool
    Brady, S.
    Kaufman, R.
    MEDICAL PHYSICS, 2015, 42 (06) : 3718 - 3718
  • [9] DOSE OPTIMIZATION FOR DIFFERENT MEDICAL IMAGING TASKS FROM EXPOSURE INDEX, EXPOSURE CONTROL FACTOR, AND mAs IN DIGITAL RADIOGRAPHY
    Zhang, Menglong
    Zhao, Bin
    Wang, Yaying
    Chen, Weixia
    Hou, Lixia
    HEALTH PHYSICS, 2012, 103 (03): : 235 - 240
  • [10] Feasibility of new patient dose management tool in digital radiography: Using clinical exposure index data of mobile chest radiography in a large university hospital
    Park, Hyemin
    Kim, Jungsu
    Kang, Eun-Ju
    Kim, Yeji
    Jo, Hyejin
    Heo, Jin-Haeng
    Yang, Wonseok
    Yoon, Yongsu
    HELIYON, 2023, 9 (10)